Yibo Zhao1,2, T. Kevin Hitchens3,4, Michele Herneisey5, Jelena M. Janjic5, Rong Guo1,2, Yudu Li1,2, and Zhi-Pei Liang1,2
1Department of Electrical and Computer Engineering, University of Illinois, Urbana-Champaign, Urbana, IL, United States, 2Beckman Institute for Advanced Science and Technology, University of Illinois, Urbana-Champaign, Urbana, IL, United States, 3Animal Imaging Center, University of Pittsburgh, Pittsburgh, PA, United States, 4Department of Neurobiology, University of Pittsburgh, Pittsburgh, PA, United States, 5Graduate School of Pharmaceutical Sciences, Duquesne University, Pittsburgh, PA, United States
Synopsis
19F-MRSI has the potential to track
multiple perfluorocarbon nanoemulsions simultaneously, but existing 19F-MRSI schemes have been limited to CSI, which provides a poor
tradeoff between resolution and scan time. In this work, a novel method is
proposed to enable fast high-resolution 19F-MRSI. In the proposed method, (k,t)-space is sampled rapidly in
EPSI trajectories; data processing is accomplished using a union-of-subspaces
model with pre-learned spectral basis. The proposed method has been evaluated
using simulation and experimental data, producing encouraging results. The
proposed method may open up new opportunities for simultaneous tracking of
different labeled cell populations in vivo.
Introduction
Perfluorocarbon (PFC) nanoemulsions (NEs)
can be used to label and track cells in vivo.1,2 19F-MRSI has been recognized as a potentially powerful tool for multi-color imaging of PFC-labeled cells because
of negligible background signals, high biocompatibility and unique spectral
patterns of different PFCs.3-5
However, unlike 1H chemical shifts, 19F chemical shift dispersion is large, stretching out over 300 ppm.6 Such a large chemical shift range makes it difficult to implement
spatiotemporal sampling trajectories (e.g., EPSI and spiral trajectories)
without spectral aliasing, especially in high magnetic field scanners. Besides,
it can also introduce chemical shift artifacts along the readout direction.7 Therefore, existing 19F-MRSI methods have been limited to CSI-type data acquisition
schemes, which provide a poor tradeoff between spatial resolution and scan time.
Although some methods have been proposed to accelerate 19F CSI, like compressed sensing,8,9
the current capability is still rather limited.
This work proposes a new method to enable
fast high-resolution 19F-MRSI
using SPICE. With a union-of-subspaces model and pre-learned spectral prior
information,10,11 the proposed method is able to
reconstruct high-quality spatiospectral distributions from highly undersampled
EPSI data. The proposed method has been validated
using both simulation and experimental data, producing very encouraging
results.Methods
Imaging model
We represent the desired 19F-MRSI signal using the union-of-subspaces model12:
$$\rho(\boldsymbol{x},f)=\sum_{m=1}^M\sum_{\ell=1}^{L_m}c_{m,\ell}(\boldsymbol{x})v_{m,\ell}(f),$$
where $$$\{v_{m,\ell}(f)\}_{\ell=1}^{L_m}$$$ is a set of spectral basis functions for the $$$m$$$-th molecule and $$$\{c_{m,\ell}(\boldsymbol{x})\}_{\ell=1}^{L_m}$$$ are the corresponding spatial coefficients.
The spectral basis functions can be pre-learned from high-quality training data
as in previous works.10,11 In 19F-MRSI studies, $$$\rho(\boldsymbol{x},f)$$$ is composed of signals from the PFC molecules
used in the specific study, for example, perfluoro-15-crown-5-ether (PCE) and
perfluorooctyl bromide (PFOB). The union-of-subspaces model effectively reduces
the number of degrees-of-freedom for representing the desired spatiospectral
distributions, relaxing the sampling requirements for accurate recovery of the
underlying spatiospectral function.
Data
acquisition
Enabled by the union-of-subspaces model, a 3D EPSI acquisition scheme was proposed. The data acquisition scheme, illustrated in Fig. 1, has
the two key features: (a) extended k-space coverage in EPSI trajectories to
achieve high spatial resolution, and (b) highly sparse temporal sampling to
reduce data acquisition time. This scheme offers high spatiospectral encoding
efficiency, yielding a factor of ~40 acceleration over the conventional CSI
scheme, but the data thus acquired have a 6-fold spectral undersampling and
cover (k,t)-space in tilted trajectories, which require special data processing
scheme to avoid aliasing and chemical shift artifacts.
Data processing
Reconstruction of the desired spatiospectral
functions was accomplished using a model-based method. The proposed method has
two key features: (a) explicit incorporation of sparse and non-uniform
(k,t)-space sampling patterns in the imaging model for effective correction of
chemical shift artifacts, and (b) use of spectral prior information in the form
of pre-learned spectral basis for separation of different molecules from
undersampled data. Note that after processing, artifact-free spatiospectral
function of each molecule can be obtained. In other words, the proposed method performs
reconstruction and spectral quantification directly from the (k,t)-space data.
More specifically, the spatial coefficients of each molecule were estimated from
the acquired data, $$$d(\boldsymbol{k},t)$$$, by solving the following optimization problem:
$$\{\hat{c}_{m,\ell}(\boldsymbol{x})\}=\arg\min_{\{c_{m,\ell}(\boldsymbol{x})\}}\left\|d(\boldsymbol{k},t)-\Omega(\boldsymbol{k},t)\mathcal{F}_{\boldsymbol{x}}\mathcal{F}_{f}B_0\left\{\sum_{m=1}^{M}\sum_{\ell=1}^{L_m}c_{m,\ell}(\boldsymbol{x})v_{m,\ell}(f)\right\}\right\|_2^2+\lambda \sum_{m=1}^{M}\sum_{\ell=1}^{L_m}\|\Phi\left\{c_{m,\ell}(\boldsymbol{x})\right\}\|_1,$$
where $$$\Omega(\boldsymbol{k},t)$$$ denotes the sparse and non-uniform sampling operator in
(k,t)-space, $$$B_0$$$ the phase term caused by the field
inhomogeneity, $$$\mathcal{F}_{\boldsymbol{x}}$$$, $$$\mathcal{F}_{f}$$$ Fourier encoding along spatial and spectral directions,
respectively, and $$$\Phi$$$ the total variation transform. After the
spatial coefficients were estimated, the desired spatiospectral distribution of each
molecule can be reconstructed as:$$\hat{\rho}_m(\boldsymbol{x},f)=\sum_{\ell=1}^{L_m}\hat{c}_{m,\ell}(\boldsymbol{x})v_{m,\ell}(f).$$
Results
Simulation
A set of simulation results is shown in
Fig. 2. Subspace-based quantification separated PCE and PFOB molecules well from
sparsely sampled data. However, non-uniform (k,t)-space sampling caused chemical
shift artifacts. By incorporating the non-uniform sampling into the model,
accurate reconstruction and quantification results were obtained.
Phantom
Phantom experiments were performed on a phantom with two vials filled
with NEs of PCE and PFOB on a 9.4T Bruker AV3HD scanner equipped with a BGA-12S HP gradient set, a dual-tuned 1H/19F 40-mm birdcage resonator and ParaVision
6.0.1. The data acquisition
parameters were: $$$\mathrm{FOV}=40\times40\times80\mathrm{~mm}^3$$$, $$$\mathrm{FA}=15^\circ$$$, $$$\mathrm{TR/TE}=800/0.2\mathrm{~ms}$$$, spatial matrix size $$$=40\times40\times40$$$ and total data acquisition time $$$=21.3$$$ min. A set of representative phantom results is shown in Fig. 3. The Fourier reconstructed PFOB signals suffered from off-resonance chemical
shift artifacts, and the spectra suffered from aliasing artifacts. The
proposed method produced high-quality PCE and PFOB maps and spectra without chemical
shift or aliasing artifacts. It would require more than 14 hours to obtain the
same results using conventional CSI schemes.
Biological phantom
PCE and PFOB NEs were injected into a fixed rat brain and EPSI data were
acquired using the parameters similar to the phantom experiment. Figure 4 shows
the results obtained using the proposed method. As can be seen, high-quality
PCE and PFOB spatiospectral functions were obtained using the proposed method. The
spatial distributions of PCE and PFOB matched our expectation.Conclusions
A new subspace-based approach has been proposed for fast high-resolution
19F-MRSI, with an unprecedented capability
of $$$1\times1\times2\mathrm{~mm^3}$$$ spatial resolution in 21.3 min. Simulation and experimental results have demonstrated
the feasibility of obtaining high-quality maps of multiple PFC NEs using the proposed
method. With further development, this method may provide a powerful tool for in vivo multi-color
labeled cell imaging.Acknowledgements
The authors gratefully
acknowledge Dr. Johannes Schneider from Bruker BioSpin for advice and
assistance implementing the 3D EPSI sequence.References
1. Partlow KC, Chen J,
Brant JA et al. 19F magnetic resonance imaging for stem/progenitor cell
tracking with multiple unique perfluorocarbon nanobeacons. FASEB J. 2007;
21(8): 1647-1654.
2. Srinvas M,
Heerschap A, Ahrens ET, Figdor CG and de Vries IJM. 19F MRI for quantitative in
vivo cell tracking. Trends Biotechnol. 2010; 28(7): 363-370.
3. Stoak CH, Hees PS,
Huang HN et al. A new perfluorocarbon for use in fluorine‐19
magnetic resonance imaging and spectroscopy. Magn Reson Med. 1993; 29(2): 188-195.
4. Ruiz-Caballo J,
Bernett BP, Bottomley PA and Bulte JW. Fluorine (19F) MRS and MRI in
biomedicine. NMR Biomed. 2011; 24(2): 114-129.
5. Jacaby C, Temme S,
Mayenfels F et al. Probing different perfluorocarbons for in vivo inflammation
imaging by 19F MRI: image reconstruction, biological half‐lives
and sensitivity. NMR Biomed. 2014; 27(3): 261-271.
6. de Graaf RA, In vivo
NMR spectroscopy: principles and techniques. Hoboken, NJ: John Wiley and Sons,
2007.
7. Dwyer AJ, Knop RH
and Hoult DI. Frequency shift artifacts in MR imaging. J Comput Assist Tomogr.
1985; 9(1); 16-18.
8. Kampf T, Fisher A,
Basse- Lüsebrink TC et al. Application
of compressed sensing to in vivo 3D 19F CSI. J. Magn. Reson. 2010; 207(2):
262-273.
9. Zhong J, Mills PH,
Hitchens TK and Ahrens ET. Accelerated fluorine‐19 MRI cell tracking
using compressed sensing. Magn Reson Med. 2013; 69(6): 1683-1690.
10. Li Y, Lam F,
Clifford B and Liang Z-P. A subspace approach to spectral quantification for MR
spectroscopic imaging. IEEE Trans Biomed Eng. 2017; 64: 2486-2489.
11. Lam F, Li Y, Guo
R, Clifford B and Liang Z-P. Ultrafast magnetic resonance spectroscopic imaging
using SPICE with learned subspaces. Magn Reson Med. 2020; 83: 377-390.
12. Ma C, Lam F,
Johnson CL and Liang Z-P. Removal of nuisance signals from limited and sparse
1H MRSI data using a union-of-subspaces model. Magn Reson Med. 2016;75(2):488‐497.