Ke Dai1, Hao Chen1, Hongda Shao2, Jianjun Liu2, and Zhiyong Zhang1
1School of Biomedical Engineering, Shanghai Jiao Tong University, Shanghai, China, 2Departments of Nuclear Medicine, Renji Hospital, School of Medicine, Shanghai Jiao Tong University, Shanghai, China
Synopsis
xSPEN is a single-shot echo-planar imaging-based MRI approach with exceptional resilience to chemical
shifts and field inhomogeneities. We introduce a time
increasing (t1) evolution as chemical shift encoding to fast obtain multiple-voxel
spectroscopy. The new method endows a 2D self-navigated motion
correction and a unique J-decoupled spectrum by splitting
the t1 evolution into τ-t1/2 and t1/2 on two
sides of π pulse, which results a constant-τ J coupling evolution. We present in vitro results, demonstrating
an alternative fast MRSI and increasing chemical shift
separation and detection with the robustness to the in-plane motion and the
unique J-decoupled spectrum capability.
Introduction
Echo Planar Spectroscopic
Imaging (EPSI) significantly accelerates MRSI by simultaneous encoding of one
spatial and the temporal dimension with echo-planar fast gradient oscillation
during signal read-out [1]. However, only half spectral bandwidth is used in
conventional EPSI, as the even and odd readout echoes are utilized separately
during reconstruction. Limited spectral bandwidth is a problem particularly at
higher magnetic field strengths (>3 T), as chemical field dispersion effects
increase, leading to spectral aliasing [2]. Besides, the splittings because of
J couplings result a broader peak along the chemical shift axis, which make
coupled compounds such as GABA and glutamate harder to detect and quantify [3,
4]. Recently, cross-term spatiotemporal encoding (xSPEN) [5] is an introduced
imaging approach delivering single-scan 2D images with exceptional resilience to
chemical shifts and field inhomogeneities. The shot-to-shot distortion-free
xSPEN images endow robustness to in-plane motion. Here, we introduced a time
increasing (t1) evolution as chemical shift encoding to fast obtain
multiple-voxel spectroscopy. This xSPEN spectroscopy has no limitation of
spectral bandwidth while the echo-planar imaging acquisition grants a high
sampling efficiency of MRSI. Furthermore, we demonstrate in vitro results with
an increasing chemical shift separation and detection by taking advantage of the unique J-decoupled spectrum capability of this
technique.Methods
The xSPEN spectroscopy
sequence is shown in Fig. 1c by performing multiple scans with increasing time
(t1) evolution while the echo-planar imaging-based acquisition of
xSPEN grants a high sampling efficiency of MRSI. The chemical shift spectrum of
each voxel is obtained by the Fourier transform along the time-increasing
dimension. The resolution and bandwidth of the chemical shift spectrum will be
determined by the usual Nyquist criteria of the maximum t1 and the t1
step, which can be flexibly adapted. By
splitting the t1 evolution into τ-t1/2
and t1/2 on two sides of π pulse, it results a constant-τ
J coupling evolution bringing
a J-decoupled xSPEN spectroscopy sequence as shown in Fig.1d. This unique
J-decoupled xSPEN spectroscopy increases the chemical shift separation and
detection of coupled compounds. Besides, the single-shot
distortion-free xSPEN images allow a self-navigated in-plane motion correction.
Studies were conducted
with a 3.0T United Imaging uMR790 system (Shanghai, China). To demonstrate the
concept, a phantom of ~200 mM GABA aqueous solution with three different
multiplets correspond to the three methylene (CH2) groups in the GABA
molecule. Parameters for the EPSI include TR =2 s, 200×200 mm2 FOV,
10 mm slice thickness, 80×80 image matrix size was cropped into 32×32 for the comparison,
the echo spacing is 0.53 ms. Parameters for the xSPEN spectroscopy include TR =2
s, 200×200 mm2 FOV, 10 mm slice thickness, 32×32 image matrix size,
a total of 160 t1 encodings with two bandwidths of 1000 Hz and 2000
Hz. The constant J coupling evolution τ=120 ms is used for J-decoupled xSPEN
spectroscopy. For the robustness to motion study, factitious phantom motions
were introduced during the xSPEN spectroscopy scanning with/without water
suppression.Results and Discussion
Figure 2 shows the
comparison among EPSI, xSPEN spectroscopy, and J-decoupled xSPEN spectroscopy.
The splittings because of J couplings in the EPSI and xSPEN spectroscopy
result broader peaks along the chemical
shift axis, while high-resolution singlets of
J-decoupled xSPEN spectroscopy enhance the chemical shift separation and
detection. A larger spectral bandwidth of 2000 Hz can be easily achieved by
xSPEN spectroscopy sequences as shown in Fig.2b.
Both the field
inhomogeneities and J couplings can result spectral broadening, which may lead
to inaccurate chemical shift maps as shown in Fig. 3 while the J-decoupled
xSPEN spectroscopy can benefit the quantification by simplifying the coupled compounds.
Figure 4 demonstrates the
robustness of xSPEN spectroscopy to the in-plane motion, shot-to-shot chemical
shift encoded xSPEN images allow a self-navigated motion tracking and
correction to greatly reduce the spectral distributions. Conclusions
We propose an alternative
fast magnetic spectroscopy imaging termed as xSPEN spectroscopy endowed with
flexible spectral bandwidths, a unique J-decoupled spectrum capability and the
robustness to in-plane motion. Acknowledgements
This work is supported by
National Science Foundation of China (No. 62001290) and sponsored by Shanghai
Sailing Program (20YF1420900).References
[1] Posse S, Otazo R,
Dager SR, Alger J. MR spectroscopic imaging: principles and recent advances. J
Magn Reson Imaging 2013;37(6):1301-1325.
[2] Al-Iedani O,
Lechner-Scott J, Ribbons K, Ramadan S. Fast magnetic resonance spectroscopic
imaging techniques in human brain- applications in multiple sclerosis. J Biomed
Sci 2017;24(1):17.
[3] Puts NA, Edden RA. In
vivo magnetic resonance spectroscopy of GABA: a methodological review. Prog
Nucl Magn Reson Spectrosc 2012;60:29-41.
[4] Ramadan S, Lin A,
Stanwell P. Glutamate and glutamine: a review of in vivo MRS in the human
brain. NMR Biomed 2013;26(12):1630-1646.
[5] Zhang Z, Seginer A,
Frydman L. Single-scan MRI with exceptional resilience to field
heterogeneities. Magn Reson Med 2017;77(2):623-634.