Christof Boehm1, Marianne Goeger-Neff2, Hendrik T. Mulder3, Benjamin Zilles2, Lars H. Lindner2, Gerard C. van Rhoon3, Dimitrios C. Karampinos4, and Mingming Wu1
1Technical University of Munich, Munich, Germany, 2Department of Medicine III, University Hospital, LMU Munich, Munich, Germany, 3Erasmus MC Cancer Institute, Rotterdam, Netherlands, 4Department of Diagnostic and Interventional Radiology, School of Medicine, Technical University of Munich, Munich, Germany
Synopsis
Motion-induced
susceptibility changes induce field variations, leading to large errors during
MR thermometry based on the linear proton resonance frequency shift. These
artefacts aggravate temperature quantification in the face of both
the long treatment duration and the mild temperature change during mild RF
hyperthermia treatments. We show with the help of simulations, a phantom
heating experiment, volunteer scans and mild hyperthermia treatment of a
patient with cervical cancer and a sarcoma patient how to correct for
this artefact source by methods known from quantitative susceptibility mapping.
The recently introduced total field inversion shows advantages over the background field removal methods.
Purpose
Mild
heating of various cancer types sensitizes tumors to radio- and chemotherapy1. This
treatment modality has thus found its way into clinical practice in recent
years, including the treatment of sarcoma patients and patients with cervical
cancer2-5. MR
temperature monitoring of mild hyperthermia (HT) treatment (40-44°C) of cancer
is done by exploiting the linear resonance frequency of water6-8. However,
susceptibility distribution changes between different time points, due to
digestive motion including gas or movement of air bubbles inside the
circulating water bolus create much stronger phase changes than the
temperature-induced phase change of -0.01ppm/°C9. Recently,
it was proposed to correct for the susceptibility artifacts by solving the
Laplacian boundary value problem (LBV)10 or by
projection onto dipole (PDF)11,12. These two
methods are well-known in the context of quantitative susceptibility mapping
for separating the foreground from the background fields. However, PDF is known to overfit especially at air-tissue interfaces and Laplacian-based method reduce the field of view. Total field inversion (TFI)13 has been proposed to alleviate the above problems and is presently applied.Theory
To estimate the phase contributions originating from susceptibility sources the following TFI cost function is solved similar to13:$$y'=\underset{y}{\arg\min}=\frac{1}{2}||w(f-d*Py)||_1+\lambda{||}\nabla{P}y||_1,$$where$$$\:{w}\:$$$the a data weighting term,$$$\:{f}\:$$$is the input phase,$$$\:{d}\:$$$is the dipole kerne in$$$\:{k}\:$$$-space,$$$\:{P}\:$$$the preconditioner and$$$\:\nabla\:$$$the gradient operation. The final susceptibility distribution is computed as$$$\:\chi=Py$$$.Methods
The
phantom heating experiment as well as the volunteer data set were acquired on a
1.5T GE system (GE Discovery MR450w/USA). The BSD2000-3D Sigma Eye
MR-compatible RF applicator (PYREXAR Medical/USA) consists of 24 dipole
antennas arranged in three rings of 8 antennas. No water was circulated during
the phantom measurement or the volunteer study. For phase unwrapping, we used
the code available at (https://gitlab.com/veronique_fortier/Quality_guided_unwrapping).
Simulations:
The
simulated matrix size was$$$\:150\times150\times150$$$, which corresponded to a
FOV of$$$\:50\times50\times50cm^3$$$. The simulated field disturbance originated
from a susceptibility change $$$\Delta\chi\:$$$within a sphere of$$$\:2\:$$$cm diameter in
the center. The susceptibility difference $$$\Delta\chi$$$ was the one between
water and air. A 3D Gaussian temperature increase profile with a peak value of
10°C and a standard deviation of$$$\:5\:$$$pixels was added to the image. A spatially variant 1st order phase was added to imitate B0 drift.
Phantom
heating:
A
double echo GRE with slice interleaved acquisition scheme was scanned for
temperature monitoring. Using the phase signal at both TEs compensated for
conductivity change-induced phase offsets (7) (TR=620$$$\:$$$ms, 25 slices, scan time=83 s, TE_1=4.8 ms, TE_2=19.1 ms, matrix size=$$$128\times128$$$,
FOV=50 cm$$$\times\:$$$50 cm, flip angle=40°, slice thickness=10 mm, bandwidth=325.5 Hz/px.)
Volunteers:
Single
echo data sets were acquired for in total 4 volunteer data sets (3 male, 1
female) (TE/TR=15ms/21ms, 20 slices matrix size=$$$128\times160$$$, FOV=50 cm$$$\times\:$$$50 cm, flip angle=14°, slice thickness=10mm). As$$$\:$$$at constant temperature, the conductivity bias did not need to be considered, a single echo$$$\:$$$acquisition scheme was$$$\:$$$sufficient. 30 min passed between
the two acquisition time points.
Patient
treatment:
Patient
treatment$$$\:$$$scans were performed with the$$$\:$$$approval of the respective local ethics
board.
A
Double-Echo Gradient Echo (DEGRE) acquisition corrected for the conductivity
bias14. We
evaluated the mild RF-HT of a cervical tumor treated$$$\:$$$with the aforementioned
BSD2000-3D Sigma Eye applicator inside a$$$\:$$$1.5T GE system. Furthermore, we
evaluated the mild RF-HT of a sarcoma in the thigh, which was treated with the
more novel BSD-2000 3D/MR applicator inside a 1.5T Philips system.Results
Simulations:
All
three methods can eliminate both the linear phase as well as the dipole, while
preserving the simulated heat distribution. While the LBV method performed
best, PDF and TFI performed similarly, as seen in$$$\:$$$the cumulative error plot(Fig.1).
Phantom$$$\:$$$heating:
The
PDF algorithm overestimates the background field effect particularly at the
border of the phantom(black arrow in Fig.2). Furthermore, we could see that
all three$$$\:$$$methods successfully removed the$$$\:$$$B0 drift effect and matched well to
the sensor probe data.
Volunteers:
In
contrast to the simulation results, here, TFI showed the best performance in
terms of$$$\:$$$cumulative error in all 4 volunteers(Fig.3).
Patient
treatment:
The
advantage of preserving pixel layers becomes particularly apparent in case of
tumors in proximity to tissue/air interfaces, as$$$\:$$$for the case$$$\:$$$of cervical cancer treatments with MR thermometry monitoring(Fig.4). It is important to note that the severity and spatial extension of the
susceptibility artifacts would not allow for B0 drift correction, as the
silicon tubes would be superimposed by that same artifact. Results for a
sarcoma patient treatment are shown(Fig.5).Discussion
TFI
performs more robustly in the presence of noise, as seen in the phantom and
volunteer measurements. This is a big advantagein the context of RF-HT treatments, as the heating device precludes the use of other MR
receive coils than the body coil and thus suffers from low SNR. In
contrast to LBV, all pixels are preserved. This is particularly useful as the tumor is oftentimes located next to
intestinal gas, and thus at the edge of the foreground mask.Conclusion
LBV,
PDF, and TFI successfully remove susceptibility artifacts without subtracting
the phase change due to temperature change. The B0 drift correction comes for
free. The PDF algorithm has the tendency to overestimate the background field
contribution, whereas the LBV method peels valuable pixel layers away for
fitting. Thus, TFI might be a promising method for gaining accurate temperature
maps during mild RF-HT treatments.Acknowledgements
No acknowledgement found.References
1. Kampinga HH. Cell biological effects of
hyperthermia alone or combined with radiation or drugs: a short introduction to
newcomers in the field. Int J Hyperthermia. 2006;22(3):191-6.
2. Issels RD, Lindner LH, Verweij J,
Wessalowski R, Reichardt P, Wust P, et al. Effect of Neoadjuvant Chemotherapy
Plus Regional Hyperthermia on Long-term Outcomes Among Patients With Localized
High-Risk Soft Tissue Sarcoma: The EORTC 62961-ESHO 95 Randomized Clinical
Trial. JAMA Oncol. 2018;4(4):483-92.
3. Issels R, Kampmann E, Kanaar R, Lindner
LH. Hallmarks of hyperthermia in driving the future of clinical hyperthermia as
targeted therapy: translation into clinical application. Int J Hyperthermia.
2016;32(1):89-95.
4. Bucklein V, Limmroth C, Kampmann E, Schuebbe
G, Issels R, Roeder F, et al. Ifosfamide, Carboplatin, and Etoposide (ICE) in
Combination with Regional Hyperthermia as Salvage Therapy in Patients with
Locally Advanced Nonmetastatic and Metastatic Soft-Tissue Sarcoma. Sarcoma.
2020;2020:6901678.
5. Franckena M, Fatehi D, de Bruijne M,
Canters RA, van Norden Y, Mens JW, et al. Hyperthermia dose-effect relationship
in 420 patients with cervical cancer treated with combined radiotherapy and
hyperthermia. Eur J Cancer. 2009;45(11):1969-78.
6. Rieke V, Vigen KK, Sommer G, Daniel BL,
Pauly JM, Butts K. Referenceless PRF shift thermometry. Magn Reson Med.
2004;51(6):1223-31.
7. Gellermann J, Hildebrandt B, Issels R,
Ganter H, Wlodarczyk W, Budach V, et al. Noninvasive magnetic resonance
thermography of soft tissue sarcomas during regional hyperthermia: correlation
with response and direct thermometry. Cancer. 2006;107(6):1373-82.
8. Gellermann J, Wlodarczyk W, Hildebrandt
B, Ganter H, Nicolau A, Rau B, et al. Noninvasive magnetic resonance
thermography of recurrent rectal carcinoma in a 1.5 Tesla hybrid system. Cancer
Res. 2005;65(13):5872-80.
9. Winter L, Oberacker E, Paul K, Ji Y,
Oezerdem C, Ghadjar P, et al. Magnetic resonance thermometry: Methodology,
pitfalls and practical solutions. Int J Hyperthermia. 2016;32(1):63-75.
10. Zhou D, Liu T, Spincemaille P, Wang Y.
Background field removal by solving the Laplacian boundary value problem. NMR
Biomed. 2014;27(3):312-9.
11. Liu T, Khalidov I, de Rochefort L,
Spincemaille P, Liu J, Tsiouris AJ, et al. A novel background field removal
method for MRI using projection onto dipole fields (PDF). NMR Biomed.
2011;24(9):1129-36.
12. Wu M, Mulder HT, Baron P, Coello E,
Menzel MI, van Rhoon GC, et al. Correction of motion-induced susceptibility
artifacts and B0 drift during proton resonance frequency shift-based MR
thermometry in the pelvis with background field removal methods. Magn Reson
Med. 2020;84(5):2495-511.
13. Liu Z, Kee Y, Zhou D, Wang Y, Spince-maille P. Preconditioned Total Field Inver-sion (tfi) Method for Quantitative Susceptibil-ity Mapping. Magnetic Resonance in Medicine2016; 78:303–315. 10.1002/mrm.26331.
14. Peters RD, Henkelman RM. Proton-resonance
frequency shift MR thermometry is affected by changes in the electrical
conductivity of tissue. Magn Reson Med. 2000;43(1):62-71.