Zhehao Hu1,2, Jiayu Xiao1, Xianglun Mao1, Yibin Xie1, Alan Kwan1,3, Xiaoming Bi4, Shlee Song5, Alison Wilcox6, Debiao Li1,2, Anthony Christodoulou1,2, and Zhaoyang Fan1,6,7
1Biomedical Imaging Research Institute, Cedars-Sinai Medical Center, Los Angeles, CA, United States, 2Department of Bioengineering, University of California, Los Angeles, Los Angeles, CA, United States, 3Smidt Heart Institute, Cedars-Sinai Medical Center, Los Angeles, CA, United States, 4Siemens Medical Solutions USA, Inc., Los Angeles, CA, United States, 5Department of Neurology, Cedars-Sinai Medical Center, Los Angeles, CA, United States, 6Department of Radiology, Keck School of Medicine, University of Southern California, Los Angeles, CA, United States, 7Department of Radiation Oncology, Keck School of Medicine, University of Southern California, Los Angeles, CA, United States
Synopsis
Non-invasive imaging
of cardiac anatomy plays an important role in diagnosis, risk
stratification, and planning of procedures in patients with cardiovascular
disease. MR imaging has the potential to provide a comprehensive evaluation of
cardiac chambers and thoracic vessels. However, the clinical workflow for the acquisition of conventional cardiac MR imaging is complex and time-consuming.
MR MultiTasking based 3D Multi-dimensional Assessment
of Cardiovascular System (MT-MACS) technique has recently been
demonstrated in thoracic aortic diseases without need for ECG- and navigator-gating.
In this work, we further extend the application of MT-MACS to the assessment of the
whole heart and great thoracic vessels.
Introduction
Imaging
assessment of cardiac and vascular anatomy and function is a key component for
diagnosis, risk stratification, and procedural planning in patients with
cardiovascular diseases1-3. MR imaging has the potential to provide a comprehensive
assessment of the entire heart through multi-contrast, motion-resolved and
water/fat imaging. However, the clinical workflow for conventional cardiac MR
imaging is complex and time-consuming, due to (a) the complexity of
cardiovascular anatomy, (b) cardiac and respiratory motion of the heart, and (c) the resultant long imaging time4. Hence, current cardiac MR images are
acquired in separate scans with several non-standard planes, and the acquisition
requires ECG-triggering with multiple breath-holds. An MR MultiTasking
based 3D Multi-dimensional Assessment of Cardiovascular System
(MT-MACS) technique has recently been shown to provide multi-contrast and
cardiac phase-resolved imaging of the thoracic aorta in a single scan without
the need for ECG triggering, navigator gating or breath-holds5. In this work, we further extend the application
of MT-MACS to the assessment of the whole cardiac structures and great thoracic
vessels with water/fat separation. Methods
Sequence
Design: In this
work, we modified the continuous 3D acquisition scheme of our recently
developed MT-MACS technique5 from single-echo Cartesian FLASH into
dual-echo stack-of-stars FLASH readouts with tiny-golden-angle ($$$\Psi$$$=32.039⁰) in-plane k-space sampling. T2-prepared inversion
recovery (T2IR) magnetization preparations are applied at constant intervals to
maximize the contrast between myocardium/vessel wall and blood and create flexible
contrast weightings during T1 recovery. Following each T2IR, RF pulse flip
angles are 3⁰ for the first 300 segments, and 1⁰ for the next 200 segments
(allowing for greater magnetization recovery). Auxiliary data are interleaved
with imaging data every 6 readouts and are collected at the 0⁰ radial spoke of
the center partition (Figure 1). Partition-encoding ordering for imaging
data is randomized with a variable-density Gaussian distribution.
Imaging
Framework: MT-MACS adopts
a low-rank tensor imaging model6 with a cardiac time dimension for
phase-resolved cine imaging, a respiratory time dimension for free-breathing
imaging, a T1 recovery dimension for multi-contrast assessment and a T2* decay
dimension for dual-echo imaging. The model represents the multidimensional
cardiac image as $$$\textbf{U}\mathbf{\Phi }$$$, where temporal basis functions $$$\mathbf{\Phi }$$$ are extracted from
high-temporal-resolution auxiliary data and spatial coefficients $$$\textbf{U}$$$ are determined by
least-squares fitting of $$$\mathbf{\Phi }$$$ to the undersampled
imaging data $$$\textbf{d}$$$:
$$\hat{\textbf{U}}=\underset{\textbf{U}}{\arg \min}\left \| \textbf{d}-\Omega (\textbf{EU}\mathbf{\Phi })) \right \|_{2}^{2}+\lambda R(\textbf{U})$$
where $$$\Omega $$$ is the undersampling
pattern, $$$\textbf{E}$$$ is the signal model
including Fourier transform and coil sensitivities. $$$R$$$ was chosen as a
wavelet sparsity penalty. Finally, the water/fat images are generated based on
a two-point Dixon method7.
In vivo
Study: Nine healthy
subjects were scanned on a 3T system (Vida, Siemens Healthcare). The prototype MT-MACS sequence was prescribed
based on an axial scout scan to cover the whole heart. Major imaging parameters
included: coronal orientation, FOV=224×224×162.4 mm3, spatial
resolution=1.4×1.4×2.8 mm3 (interpolated to 1.4 mm3
isotropic), T2-prep duration=60 ms, TR/TE1/TE2=3.94/1.23/2.46 ms, 500 readouts after
each T2IR, total scan time=10 min. 2D Cine SSFP and 2D T2-TSE were acquired as
references.
Image
Analysis: Image
reconstruction was performed offline to generate water-only images with
multiple contrast weightings (i.e. bright-blood [BB], dark-blood [DB],
gray-blood [GB]) and corresponding cine series as well as fat-only images.
Qualitative (image quality score 1-4) and quantitative (left ventricular
ejection fraction [LVEF], left atria [LA]/left ventricle [LV]/right atria [RA]/right
ventricle [RV] wall thickness at middiastole) analyses were performed. Results
Figure 2 shows example images generated by MT-MACS, including water-only
multi-contrast whole-heart anatomical images and corresponding cine series,
together with fat-only images. A total of 36 3D image sets (3 water-only
contrasts and 1 fat-only) were scored for image quality. For water-only images
with three representative contrasts, cardiac chambers and thoracic aorta were
evaluated separately (Table 1). MT-MACS provided slightly lower LVEF
measurements compared with Cine SSFP (Regression line: Y=1.060X–0.075; R2=0.855,
P<0.001), but the overall values were still within the physiological
range8 (Figure 3). MT-MACS
quantified the wall thickness of each cardiac chamber at their own optimal DB
phases, thanks to the flexible image contrasts (Figure 4A). The thicknesses were
2.52 ± 0.19 mm, 8.99 ± 0.53 mm, 2.50 ± 0.21 mm, and 4.32 ± 0.78 mm,
respectively for LA, LV, RA, and RV, which are all within the normal anatomical
range9-12. The
cardiac structures on 2D T2-TSE images from 2 subjects were substantially
blurred due to respiratory motion and were therefore excluded from analysis. Good
agreement was observed between MT-MACS and TSE in the remaining 7 subjects
(Figure 4B). Discussion
MT-MACS provides
a comprehensive ECG- and navigator-free assessment of cardiac and great vessel
anatomy within a single 10-minute scan. By adopting the MR multitasking
framework, the cardiac image is modeled as a multidimensional low-rank tensor,
which contains a cardiac motion, an inversion recovery and a multi-echo time
dimension for phase-resolved cardiac cine imaging, multi-contrast assessment,
and water/fat separation, respectively. Furthermore, the continuous data
acquisition scheme eliminates the need for ECG triggering, respiratory
navigators or breath-holds, and provides flexible image contrasts for better
cardiac chamber delineation, which can greatly simplify the acquisition
workflow and potentially improve interpretation accuracy. Conclusion
In this work,
we developed an MT-MACS technique for comprehensive assessment of the combined
cardiac and thoracic aortic system. Further clinical validation is underway. Acknowledgements
This work was supported by NIH 1R01EB028146 and 1R01HL147355. References
1. Pushparajah K, Duong P, Mathur S,
Babu-Narayan SV. EDUCATIONAL SERIES IN CONGENITAL HEART DISEASE: Cardiovascular
MRI and CT in congenital heart disease. Echo research and practice
2019;6(4):R121-R138.
2. Milotta
G, Ginami G, Cruz G, Neji R, Prieto C, Botnar RM. Simultaneous 3D whole‐heart bright‐blood and black blood imaging for
cardiovascular anatomy and wall assessment with interleaved T2prep‐IR. Magnetic resonance in medicine
2019;82(1):312-325.
3. Leiner
T, Bogaert J, Friedrich MG, Mohiaddin R, Muthurangu V, Myerson S, Powell AJ,
Raman SV, Pennell DJ. SCMR Position Paper (2020) on clinical indications for
cardiovascular magnetic resonance. Journal of Cardiovascular Magnetic Resonance
2020;22(1):1-37.
4. Feng
L, Coppo S, Piccini D, Yerly J, Lim RP, Masci PG, Stuber M, Sodickson DK, Otazo
R. 5D whole‐heart sparse MRI.
Magnetic resonance in medicine 2018;79(2):826-838.
5. Hu
Z, Christodoulou AG, Wang N, Shaw JL, Song SS, Maya MM, Ishimori ML, Forbess
LJ, Xiao J, Bi X. Magnetic resonance multitasking for multidimensional
assessment of cardiovascular system: Development and feasibility study on the
thoracic aorta. Magnetic Resonance in Medicine 2020.
6. Christodoulou
AG, Shaw JL, Nguyen C, Yang Q, Xie Y, Wang N, Li D. Magnetic resonance
multitasking for motion-resolved quantitative cardiovascular imaging. Nature
biomedical engineering 2018;2(4):215-226.
7. Coombs
BD, Szumowski J, Coshow W. Two‐point Dixon
technique for water‐fat signal
decomposition with B0 inhomogeneity correction. Magnetic resonance in medicine
1997;38(6):884-889.
8. Kosaraju
A, Makaryus AN. Left ventricular ejection fraction. 2019.
9. Takahashi
K, Okumura Y, Watanabe I, Nagashima K, Sonoda K, Sasaki N, Kogawa R, Iso K,
Ohkubo K, Nakai T. Relation between left atrial wall thickness in patients with
atrial fibrillation and intracardiac electrogram characteristics and ATP‐provoked dormant pulmonary vein
conduction. Journal of Cardiovascular Electrophysiology 2015;26(6):597-605.
10. Bishop
M, Rajani R, Plank G, Gaddum N, Carr-White G, Wright M, O'Neill M, Niederer S.
Three-dimensional atrial wall thickness maps to inform catheter ablation
procedures for atrial fibrillation. Europace 2016;18(3):376-383.
11. Yoneyama
K, Venkatesh BA, Bluemke DA, McClelland RL, Lima JA. Cardiovascular magnetic
resonance in an adult human population: serial observations from the
multi-ethnic study of atherosclerosis. Journal of Cardiovascular Magnetic
Resonance 2017;19(1):52.
12. Ho
S, Nihoyannopoulos P. Anatomy, echocardiography, and normal right ventricular
dimensions. Heart 2006;92(suppl 1):i2-i13.