Ria Forner1, Kyung Min Nam1, Tijl van der Velden1, and Dennis Klomp1
1UMC Utrecht, Utrecht, Netherlands
Synopsis
Here
we present the very first 31P MRS data in the human tongue at 7T. Three
different receive coils were designed and evaluated to compare their
performance: two each inside the mouth: a loop and also a saddle coil and a 3-channel
loop array positioned externally. The transmit setup consists of an integrated
31P body birdcage coil for 31P transmit and an array of 8 1H dipoles for image
registration. The loop coil seems to outperform the other two coils albeit that
signals may originate from jaw. For voxels inside the tongue, relatively high
phosphomonoesters were observed in all volunteers.
Introduction
Tongue
cancer has a 5 year- survival rate of 50-60%. Current clinical practice is to
use ultrasound image guided surgery to only remove the tumour without removing
substantial parts of healthy tissue. Unfortunately, in many cases, the boundary
of the excised specimen shows residual tumour tissue indicating extra surgery. An
alternative to indicate tumour extent may be to reveal elevated cell
proliferation by observing phospholipid metabolism1 with 31P MRSI. Here
we present two intra-oral lingual – and one 3-channel external 31P Rx (receive)
coils, combined with a distal 8 channel TxRx (transceiver) dipole array for 1H
excitation placed inside a 31P full-body Tx birdcage coil integrated in a 7T MR
system to investigate feasibility of non-invasive metabolic imaging of the
tongue. Methods
Three
versions of the 31P receive coils were constructed (Fig1). Version a) is a
saddle coil formed around a mouth guard. Version b) is a simple loop coil to be
inserted into the mouth such that it lies above the tongue with the head bent to
maximize orthogonal orientation to the main field. Version c) is an array of
three external loop coils which cover both cheeks as well as the mouth area.
All
coils had diode detuning implemented for detuning the receive coils during 31P
transmit. This detuning circuitry was tested on the bench via a comparison
between the transmission S-parameter measurements between the coil-under-test
and a sniffer-loop. Additionally, the malfunctioning circuitry was verified
before each RF pulse. For the phantom tests of
coils b) and c) a spherical phantom containing two smaller spheres was placed
inside the dipole array and the coil under test was positioned so as to be
similarly located as the in vivo measurements. There are 4 different metabolites
in the sphere: 200mM Pi in one smaller sphere; and 50mM each of PC, PE and GPC
in the other smaller sphere.
For coil a), it was necessary to additionally
submerge the coil in a 0.9% saline solution so as to provide sufficient load.
The subjects were positioned head-first and supine with each of the coils used successively
to perform the measurements.
In
each case, the possibility of coupling associated local SAR hotspots between
the 1H TxRx array and the 31P receive coils was verified by a comparison of B1
maps – with and without the presence of 31P receive coils inside the 1H dipole
array. A deviation of less than 10% was seen in all 3 cases, which was
acceptable.
The
protocol started with a localiser survey scan using the 1H array, followed by
B0 shimming of the tongue area. An FID is acquired to determine the frequency
offset in Hertz to place the metabolite of interest (PCr in vivo) at 0ppm. A 3D
CSI in the transverse plane is then performed, with slightly varying parameters
in vivo (FOV 224*224*225mm3 ; resolution 15*15*15mm3) than in the phantom (FOV
220*220*105mm3; resolution 22*22*7mm3) measurements. The CSI scans are
performed at a low flip angle (12 degrees as Ernst angle) to accommodate a high
bandwidth excitation at short TR; with a correspondingly short TE of 0.61ms and
a TR of 60ms. RF block pulses are used giving a B1 of 4.5uT and a total scan
time of 17 mins.
Post-processing of data was
done in two main parts. First, the raw data were read into the Compass
Framework (Tesla Dynamic Coils, Zaltbommel, the Netherlands), based on the WSVD
channel combination method (for version c)). The
PCr peak in vivo (Pi in the phantom) of this best voxel in the tongue area is
then further processed to determine the area under the peak to account for B0
inhomogeneities. Similarly, the SNR for the best voxel is determined by the
ratio of this area and the standard deviation of the spectral noise.Results
Good background images
could be obtained with the dipole array driven with a fixed amplitude and phase
setting between the channels as optimized in a separate scan session (Fig 2). For
all three receiver setups, good 31P MRSI could be obtained from the tongue,
albeit that B0 shimming is severely compromised due to substantial air- tissue
boundaries in the mouth (Fig 3). Apart from the data acquired with the loop, relatively
high phosphomonesters (PME) are observed when compared to typically reported
31P MR spectra of muscle. In fact, the spectrum obtained from the loop may
originate from the jaw muscles rather than the tongue, as judged from the
background images and as it resembles the typically reported muscle spectra. The
highest SNR of the voxel in the tongue area for volunteer 1 was 62 for the saddle,
174 for the loop and 96 for the external array. Discussion and Conclusion
Successful
31P MRSI can be obtained from the human tongue in vivo at 7T. Surprisingly, the
external array outperformed the saddle coil, which may be due to B0
inhomogeneities and destructive phase coherence considering the relatively
large voxels. Owing to the large chemical shift of 31P, despite suboptimal B0
shims, PME signals that are indicative of tumour grade can be well distinguished. Acknowledgements
H2020-FETopen:NICIReferences
[1] K. Glunde, C. Jie, and Z. M. Bhujwalla,
“Molecular Causes of the Aberrant Choline Phospholipid Metablism in Breast
Cancer,” Cancer Res., vol. 64, pp. 4270–4276, 2004.