Jiying Dai1,2, Tijl A. van der Velden1, Johannes M. Hoogduin1, Fabian Bartel2, Ettore F Meliadò1,2, Mark van Uden2, Catalina S. Arteaga de Castro2, Evita C. Wiegers1, Martijn Froeling1, Mark Gosselink1, Alexander J. E. Raaijmakers1,3, and Dennis W. J. Klomp1,4
1Radiology, UMC Utrecht, Utrecht, Netherlands, 2Tesla Dynamic Coils, Zaltbommel, Netherlands, 3Biomedical Engineering, Eindhoven University of Technology, Eindhoven, Netherlands, 4Tesla Engineering Ltd, West Sussex, United Kingdom
Synopsis
Multi-nuclei magnetic resonance imaging and
spectroscopy are interesting techniques to study metabolism, treatment efficacy
tracking, and early diagnoses of many diseases. However, there are challenges
to acquire high-quality MR images or spectra of nuclei other than proton, for
example, much lower signal level, as well as the inconvenience of switching RF
hardware in between scans, which leads to repositioning of the subject and
extra preparation time. To overcome the above issues, we developed a quintuple-tuned
RF coil for sensitive whole-brain scans, targeting five nuclei: 1H, 19F,
31P, 23Na and 13C. Bench tests and in-vivo
scans have shown promising SNR.
Introduction
Metabolic imaging is powerful in the detection
of treatment effects or for early diagnoses of many diseases. However,
metabolic imaging via 1H MRSI is challenging primarily due to the dominant
signals of water and lipids that can obscure the metabolite signals. When targeting
other nuclei than 1H, metabolic imaging can be simpler, and
sufficient sensitivity can be obtained at high fields. Yet due to the
requirement of differently tuned coils, metabolic scans are normally not
applied to patient studies that rely on conventional MRI as well. This abstract
presents the design and performance of a quintuple-tuned head setup for an MRI
system focused on merging anatomic with metabolic MRI (META-scan). Our setup
combines a broadband eight-channel RF transceiver for 1H and 19F,
as well as a triple-tuned local-receiver array for 31P, 23Na
and 13C. Bench-top or scan measurements for all 5 nuclei are presented
to indicate the efficiency. For sodium, phosphorus and proton, in vivo results
from a human brain are shown. Methods
The META head
coil (Figure 1a, 1d) contains an eight-channel
dipole array (Figure 1b) as the RF transceiver for 1H
and 19F, a fifteen-channel surface-loop array (Figure 1c) as RF receiver for 31P,
23Na and 13C, and a digital interface platform (Figure 1e) towards the scanner.
The dipoles[1]
are tuned at 290MHz, which is between the Larmor frequencies of 1H
and 19F, taking advantage of the wide bandwidth of the dipole
design. Each surface loop of the receiver array is dual-tuned for 31P
and 23Na, and the third frequency is realized by switching an extra
capacitance with a PIN diode to decrease the lower resonance from 79MHz(23Na)
to 75MHz (13C). Each surface loop is directed to a low-loss diplexer
to split signals of different frequencies. The split signals are fed to narrow-band
preamplifiers and digital receivers, for 31P on one hand, 13C
and 23Na on the other hand.
The coil
was interfaced to a 7T whole body MR system (Philips Healthcare, Best, NL). The
system was equipped with an 8-channel Tx/Rx setup for 1H and 19F
(8*2kW), a quadrature birdcage bore-coil[3] tuned for 31P,
connected to a 25kW amplifier, and a Helmholtz clamp for 23Na
excitation, connected to a 4 kW amplifier. In addition, the system was modified
to accommodate multi-channel receiver array for all nuclei.
B1+
and SAR simulations were performed using DUKE[4] (Sim4Life, Swiss)
to assess safety and efficiency for 19F and 1H. Bench-top
measurement of Q-factors of the receiver-loops in loaded and unloaded
conditions were obtained and was compared to single-tuned loops to ensure high
efficiency. 31P channel was also verified with a phantom scan
comparison to a quadrature 31P birdcage[2]. An in vivo
scan session was obtained with B1+ phase shimming for 1H,
automated power adjustments, and B0 shimming, followed by a 3D TFE
sequence of 6.5 minutes. In addition, a 3D sodium scan was obtained in 10
minutes with a TE of 1.35ms and TR of 90ms. Finally a 3D 31P CSI FID
was obtained in 10 minutes using a flip angle of 13deg and TR of 71.1ms.Results
Proton and Fluorine: Figure 2 shows the simulation model, the simulated
B1+ field and the local SAR at both 19F and 1H
Larmor frequencies at 7T, with 1 Watt accepted power per channel, in quadrature
mode. It indicates that, with the same amount of accepted power, the transmit
fields for 1H and 19F are highly similar.
Figure 3 shows the TFE images. 18µT was achieved at the center of the ROI, with 922 Watt forward peak
power per channel.
Phosphorus: The SNR in the center of META head
coil is 11% higher compared to a highly efficient dual-tuned (31P
and 1H) head-birdcage[2] under identical scan conditions
(i.e. same flip angle, TR, and load). Figure 4c-d show the signal-level comparison.
Figure 4a-b show a slice of the 3D CSI,
indicating good coverage over the entire brain and high SNR.
Sodium: Figure 5a-b show 23Na images of
an in-vivo scan. Considering the very short T2* of 23Na,
the used echo time (1.35ms) is far from optimal.
Carbon: Bench measurements show that the receiving
performance of the 13C elements is similar to that of 23Na
(at max 1dB loss): S21 between two pick-up probes next to an unloaded
surface-loop shows that the Q-factor of the unloaded surface-loop does not
change when shifting from the resonance of 23Na to that of 13C
(Figure 5c).Discussion and Conclusion
We have
demonstrated a successful integration of an RF setup that could detect 23Na,
13C, 31P, 19F and 1H in a single
scan session. Since tracers are generally required for 13C and 19F,
we show natural-abundance 3D images of only 31P, 23Na and
1H, in a healthy human brain. Since the frequency of 13C
is close to 23Na and 19F close to 1H, we do
not expect the sensitivity to be different for these nuclei. Excluding the preparation
time, the acquisition time for the 1H, 31P and 23Na
scans can be less than 25 minutes in total, paving the way to incorporate
metabolic MRI of all five nuclei in one scan session of acceptable scan time.Acknowledgements
I would like to thank Dutch Research Council (NWO) for funding the META-scan project.
References
[1] Raaijmakers,
A.J., Italiaander, M., Voogt, I.J., Luijten, P.R., Hoogduin, J.M., Klomp, D.W.
and van den Berg, C.A. (2016), The fractionated dipole antenna: A new antenna
for body imaging at 7 Tesla. Magn. Reson. Med., 75: 1366-1374. https://doi.org/10.1002/mrm.25596
[2] Hendriks, AD, van der Kemp, WJM, Luijten, PR, Petridou,
N, Klomp, DWJ. SNR optimized
31P functional MRS to detect mitochondrial and extracellular pH change during
visual stimulation. NMR in Biomedicine. 2019; 32:e4137. https://doi.org/10.1002/nbm.4137
[3] van Houtum, Q, Welting, D, Gosselink, WJM, Klomp, DWJ,
Arteaga de Castro, CS, van der Kemp, WJM. Low SAR 31P (multi‐echo) spectroscopic imaging using an integrated
whole‐body transmit coil at 7T. NMR in Biomedicine. 2019; 32:e4178. https://doi.org/10.1002/nbm.4178
[4] Christ
A, Kainz W, Hahn EG, et al. The Virtual Family—development of surface‐based
anatomical models of two adults and two children for dosimetric simulations.
Phys Med Biol. 2010;55:N23–N38.
[5] Froeling
M: QMRTools: a Mathematica toolbox for quantitative MRI analysis. J Open Source
Softw 2019; 4:1204.