Bernhard Gruber1,2, Jason P. Stockmann1, Azma Mareyam1, Boris Keil3, Anpreet Ghotra3, David A. Feinberg4, and Lawrence L. Wald1
1Department of Radiology, A.A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Harvard Medical School, Charlestown, MA, United States, 2High Field MR Center, Center for Medical Physics and Biomedical Engineering, Medical University Vienna, Vienna, Austria, 3Department of Life Science Engineering, Institute of Medical Physics and Radiation Protection, Mittelhessen University of Applied Science, Giessen, Germany, 4Helen Wills Neuroscience Institute, University of California, Berkeley, Berkeley, CA, United States
Synopsis
A 128-channel receive array
for cortical brain imaging at 7T was simulated and constructed. The tight-fitting
brain-only coil was designed for use with a recently constructed head gradient
system (Gmax = 200 mT/m and Smax = 900mT/m/s) for use with either a single
channel birdcage Tx or a 24-channel pTx coil. The goal was to maximize cortical
SNR and achieve the high acceleration needed for single-shot EPI based fMRI at high sub-millimeter isotropic resolution.
Introduction
Human fMRI studies with
single shot acquisitions have recently broken the 1 mm
isotropic resolution barrier, but resolutions surpassing 0.5 mm are needed to
approach the biological limits of fMRI resolution [1] [2] and delineate laminar and
columnar functional organization. Although zoomed imaging techniques [3] can help with the spatial encoding [4] [5] [6],
larger regions are needed for a more complete view of cortical architecture,
requiring higher SNR and parallel imaging performance than current receive
coils (Rx) can achieve. As part of the MR
Corticography (MRCoG) NIH Brain Initiative project (U01EB025162), which is
exploring the limits of cortical imaging, a customized tight fitting 128-channel
receive head-only coil array is designed, simulated and constructed.Methods
Figure 1 shows the tight space constraints within the 37
cm space provided on the table top of the 44 cm dia. head gradient (with 40 cm
dia. clear-bore) of the upgraded Siemens MAGNETOM Terra system (Siemens
Healthineers, Erlangen, Germany). Space constraints made it necessary to
minimize electronics and their interaction with each other to not only fit the
receive and transmit components but also house future integration of B0
shim components. The configuration is shown
in Fig.1. Figure 1C shows the 3D printed (Fortus F360mc, Stratasys, MN, USA) non-splitable
helmet former (ABC-PC), with 128 attached loops of 44 mm diameter, which fits
75% of the male population and almost 100% of the female population.
Each
loop (Fig. 3B), made of 16 AWG tin-plated wire, is tuned to 297.2 MHz and impedance
matched
under loaded condition, to minimize the noise figure of the Siemens 7T
preamplifiers. All bench measured values were acquired without an RF fuse in
the loop, using a loosely coupled double probe (S21: -75dB) in 20 mm
distance to the loop and with a 5 mm foam spacer between the sample (loading
phantom: 0.2 S/m DC, with 0.2 ml/l Gd) and the loop. Adjacent elements were
inductively decoupled using critical overlap [7] [8].
Coupling between next-nearest neighbours is addressed utilizing preamplifier
(PA) decoupling, by transforming the low impedance of the PA to an open circuit
in the loop using a 50 mm long coaxial cable and a series capacitor of 8 pF.
The
Tx coil consisted of a 1-ch detune-able band-pass Tx birdcage coil with
16-rungs of length 280 mm length, mounted on the inside of a 320 mm fiberglass
tube (330mm length) within a slotted shield with a 269 mm diameter (Fig. 1B and
2A).
Coil
simulations were carried out in MARIE (Magnetic Resonance Integral Equation
Suite) [9] with a unit
driving coil in each loop laid out to accurately match the constructed loop
configuration (Fig. 3A). The simulated performance was compared to a 32-ch (loop
dia. 88 mm) and 64-ch (loop dia. 63 mm) head coil [10] [11], with each loop layout
positioned on the same helmet former. The simulation in MARIE used a
homogeneous head phantom (average brain: ε=52, σ=0.55 S/m). SNR maps are
computed according to Roemer et al. [8] and included coil copper
losses, estimated from measured Q-ratio data.Results
The 44mm
dia. loops show a Qloaded of 43.2 and Qunloaded of 189 (Q-ratio=4.38),
and with six surrounding elements the Qratio drops to 2.73 (Fig. 3C).
Inductive decoupling between two loops is -17 dB, and preamplifier decoupling
is about -30 dB. In average for the whole array, the values go down to -12 dB
for the inductive decoupling and -15 dB for the preamplifier decoupling.
Unaccelerated
SNR (Fig. 4A) in the cortical area is predicted to have a substantial increase going from 32- to 128-channels and a 2.3-fold
increase from 64- to 128 receive channels with no gain at the center of the
head (Fig. 4C). 2D and 1D accelerated SNR with R=3x3 and R=5 for the
128-channel array demonstrate the benefits for accelerated imaging of the
cortex, compared to the 64- or 32-channel array (Fig. 4B and 5). At
acceleration rate R=4, the maximum 1/g-factor for the 128-channel array is
about 72 % of that of the 32-channel array. For R=3x3 in 2D acceleration, the
maximum g-factor was 88 % of the 32-channel
array.Discussion and Conclusion
The simulated and presented 128-ch array (Fig. 2B) is
expected to increase the sensitivity in cortical regions by providing
SNR substantially higher than with a 64-channel
array through a combination of intrinsic SNR and g-factor improvements. The 128-channel
array will
be tested in vivo following the completion of the system upgrade to 128 Rx channels
for 7T.Acknowledgements
Research reported in this publication was supported by the NIH BRAIN
Initiative, the National Institute of Biomedical Imaging and Bioengineering,
under award numbers U01EB025162.
The content
is solely the responsibility of the authors and does not necessarily represent
the official views of the National Institutes of Health.
References
[1]
A. J. Poplawsky, M. Fukuda, B. Kang, J. H. Kim, M. Suh and
S.-G. Kim, "Dominance of layer-specific microvessel dilation in
contrast-enhanced high-resolution fMRI: Comparison between hemodynamic spread
and vascular architecture with CLARITY," NeuroImage, no. 197, pp.
657-667, 2019.
[2]
J. R. Polimeni and L. L. Wald, "Magnetic Resonance
Imaging technology - bridging the gap between non-invasive human imaging and
optical microscopy," Current Opinion in Neurobiology, no. 50, pp.
250-260, 2018.
[3]
D. A. Feinberg, J. C. Hoenninger, L. E. Crooks, L. Kaufman,
J. C. Watts and M. Arakawa, "Inner volume MR imaging: technical
concepts and their application," Radiology, vol. 156, no. 3,
1985.
[4]
K. Setsompop, D. A. Feinberg and J. R. Polimeni,
"Rapid brain MRI acquisition techniques at ultra-high fields," NMR
in Biomedicine, no. 29(9), pp. 1198-1221, 2016.
[5]
L. Huber, D. Ivanov, S. N. Krieger, M. N. Streicher, T.
Mildner, B. A. Poser, H. E. Moeller and R. Turner, "Slab selective,
BOLD-corrected VASO at 7 Tesla provides measures of cerebral blood volume
reacitivity with high signal-to-noise ratio," Magn. Reson. in Med., no.
72(1), pp. 137-148, 2014.
[6]
R. M. Heidemann, D. Ivanov, R. Trampel, F. Fasano, H.
Meyer, J. Pfeuffer and Turner R., "Isotropic submilimetere fMRI in the
human brain at 7T: combining reduced field-of-view imaging and partially
parallel acquisition," Magn. Reson. in Med., no. 68(5), pp.
1506-1516, 2012.
[7]
B. Keil and L. L. Wald, "Massively parallel MRI
detector arrays," Journal of Magnetic Resonance, no. 229, pp.
75-89, April 2013.
[8]
P. B. Roemer, W. A. Edelstein, C. E. Hayes, S. P. Souza and
O. M. Mueller, "The NMR phased array," Magnetic Resonance in
Medicine, vol. 16, pp. 192-225, 1990.
[9]
J. F. Villena, A. G. Polimeridis, A. Hochman, Y. Eryaman,
E. Adalsteinsson, L. L. Wald, J. K. White and L. Daniel, "Fast
electromagnetic analysis of MRI transmit RF coilsbased on accelerated
integral equation methods," IEEE Trans. Biomed. Eng., 2016.
[10]
A. Mareyam, J. Cohen-Adad, J. N. Blau, J. R. Polimeni, B.
Keil and L. L. Wald, "A 32 Channel receive-only 3T array optimized for
brain and cervical spine imaging," in Proce. Intl. Soc. Mag. reson.
Med. 18, Stockholm, 2010.
[11]
B. Keil, J. N. Blau, S. Biber, P. Hoecht, V. Tountcheva, K.
Setsompop, C. Triantafyllou and L. L. Wald, "A 64-channel 3T array coil
for accelerated brain MRI," Magnetic Resonance in Medicine, vol.
70, pp. 248-258, 2012.