Siyuan Hu1, Ignacio Rozada2, Rasim Boyacioglu3, Stephen Jordan4, Sherry Huang3, Matthias Troyer4, Mark Griswold3, Debra McGivney3, and Dan Ma3
1Biomedical Engineering, Case Western Reserve University, Cleveland, OH, United States, 21Qbit, Vancouver, BC, Canada, 3Case Western Reserve University, Cleveland, OH, United States, 4Microsoft, Redmond, WA, United States
Synopsis
MR Fingerprinting is able to quantify multiple tissue
properties simultaneously. Here we propose an advanced MR Fingerprinting
optimization framework that computes and minimizes the quantitative random
errors, undersampling errors and background phase errors in MRF maps simultaneously
in the cost function. The optimization is solved by quantum-inspired
algorithms. The proposed framework could provide accelerated MRF scans that are
robust to undersampling and system imperfections, and outperform the
human-designed sequence on the tradeoff between duration and precision.
Introduction
MR Fingerprinting (MRF) is a quantitative imaging method
that simultaneously provides multiple tissue property maps1. Appropriate pulse sequences
can reduce scan time with improved measurement precision. However, such an optimization
problem is very challenging. It is nontrivial and computationally expensive to
model both spatially and temporally dependent acquisition errors and random errors
in the objective function. To address these issues, here we propose an advanced
MRF optimization framework that 1) uses a fast computing strategy to model spatially
and temporally dependent artifacts due to systematic errors including
undersampling and phase errors, and 2) directly minimizes the resulting
quantitative errors in the simulated tissue property maps in the cost function.
A quantum-inspired optimization (QIO) approach, which has been shown to be able
to escape from local minima, is adopted to handle this optimization problem
with minimal constraints.Methods
Cost
Function definition
While the proposed framework could be applied to any MR
sequence, we have started with an MRF-FISP sequence2. The cost function was constructed
as a weighted combination of T1 and T2 errors due to random noise and
systematic errors with a scan time penalty. The measure of random error was inferred
in terms of quality factors that estimates the likelihood of signal errors due
to thermal noise3. The systematic errors include
aliasing artifacts due to undersampling and phase errors due to system
imperfections such as B0 inhomogeneity. They are the dominant error sources in
fast MRF scans, and the interplay between the two results in commonly seen
shading artifacts in in vivo MRF maps (Figure 1.d). Accurately predicting these
spatially and temporally dependent artifacts is a key to evaluate sequence performance
during optimization and provide fast and robust MRF scans. The systematic errors
were estimated using a fast image series simulator4, where a three-tissue digital
brain phantom was used to simulate highly undersampled images imposed with additional
phase errors to obtain the aliased MRF maps (Figure 1.c). This simulator alone
achieved a factor of over 100 accelerations in computing MRF images as compared
to a conventional NUFFT method.
Optimization
480 flip angles and 480 TR variables were separately parameterized
as a function of 20 pulse indexes using cubic splines. Optimizations were carried
out by substochastic Monte Carlo5, a quantum inspired
optimization algorithm, and simulated annealing method6 to find a global optimum.
Validation
We first evaluated the image quality of the optimized scans in
simulations. The fully sampled images of a brain phantom were first simulated
using the ground truth signal evolutions. Spatially dependent phase maps with
four different directional variations were simulated, and multiplied to the
fully sampled images to model the phase errors. The images were then
undersampled with a single shot spiral in k-space per TR. Finally, the
undersampled images were reconstructed into T1 and T2 MRF maps via dictionary
matching (Figure 1.c).
The optimized sequences were validated by multiple in vivo
scans to evaluate image quality and reproducibility. In vivo scans were
performed on healthy volunteers in compliance with the IRB in a Siemens 3T
Skyra scanner. All scans were acquired with an image resolution of 1.2×1.2
mm2 using a spiral acquisition, with an undersampling factor of 48.
Reproducibility of the optimized MRF sequences against
random errors were evaluated via bootstrapping simulations7 using in vivo data. The
results were compared against the performance of the human-designed MRF
sequence, where the sequence was truncated to different durations to obtain a tradeoff
relation between its precision and duration.Results
Figure 1 shows the simulation results of a MRF sequence
using different simulation methods. The simulated maps obtained by
incorporating both undersampling and phase errors show shading artifacts similar
to those in in vivo scans, which supports this simulation method as replicating
in vivo conditions.
Figure 2 shows the simulation results of an example
optimized sequence. In the simulation, maps from the human-designed sequence are
subject to severe shading artifacts; whereas the optimized scan is immune to
such artifacts. In vivo performance (Figure 3) of several optimized scans show
similar results, which validates their robustness against undersampling and
system variations.
Figure 4 shows the bootstrap statistics of optimized
sequences generated from various cost functions in comparison with
human-designed sequences. The optimizer can yield sequences achieving higher
precision at given duration or shorter length for comparable precision. Note
that bootstrap simulations only estimate random errors but not undersampling errors
and phase errors, thus the optimized sequences that behaves the best in
bootstrapping exams may not be the ones exhibiting the best image quality.
Figure 5 shows an optimized sequence pattern. It illustrates
very distinct pattern features that are consistently observed in optimized
sequences, such as “saturated” flip angles and “spiked” TR values at low flip
angle regions.Conclusion
We propose an optimization framework to automatically design
MRF scans to achieve increased precision and intrinsic robustness against undersampling
artifacts and system imperfections. This optimization paradigm facilitates
great flexibility in cost function design. Therefore, it could be extended for
more complex MRF implementations, or modified for different MRF applications.Acknowledgements
The authors would like to
acknowledge funding from Siemens Healthineers, Microsoft and NIH grant
EB026764-01 and NS109439-01References
1. Ma D, Gulani V, Seiberlich N, et al.
Magnetic Resonance Fingerprinting. Nature. 2013;495(7440):187-192.
2. Jiang Y, Ma D, Seiberlich N, Gulani V,
Griswold MA. MR fingerprinting using fast imaging with steady state precession
(FISP) with spiral readout. Magn Reson Med. 2015;74(6):1621-1631.
doi:10.1002/mrm.25559
3. Kara D, Fan M, Hamilton J, Griswold M,
Seiberlich N, Brown R. Parameter map error due to normal noise and aliasing
artifacts in MR fingerprinting. Magn Reson Med. 2019;81(5):3108-3123.
doi:10.1002/mrm.27638
4. McGivney D, Boyacioglu R, Jordan S, et
al. A Fast Approximation of Undersampling Artifacts in MR Fingerprinting. In: ISMRM
Submitted. ; 2019.
5. Jarret M, Jordan SP, Lackey B.
Adiabatic optimization versus diffusion Monte Carlo methods. Phys Rev A.
2016;94(4). doi:10.1103/PhysRevA.94.042318
6. Kirkpatrick S, Gelatt CD, Vecchi MP.
Optimization by simulated annealing. Science (80- ).
1983;220(4598):671-680. doi:10.1126/science.220.4598.671
7. Riffe MJ, Blaimer M, Barkausas KJ,
Duerk JL, Griswold MA. SNR Estimation in Fast Dynamic Imaging Using
Bootstrapped Statistics. In: Proceedings of the 15th Scientific Meeting,
International Society for Magnetic Resonance in Medicine. ; 2007:1879.