Daniel J. West1, Lucilio Cordero-Grande1,2,3, Rui P. A. G. Teixeira1,2, Giulio Ferrazzi4, Joseph V. Hajnal1,2, and Shaihan J. Malik1,2
1Biomedical Engineering and Imaging Sciences, King's College London, London, United Kingdom, 2Centre for the Developing Brain, King's College London, London, United Kingdom, 3Biomedical Image Technologies, ETSI Telecomunicación, Universidad Politécnica de Madrid & CIBER-BNN, Madrid, Spain, 4IRCCS San Camilo Hospital, Venice, Italy
Synopsis
ihMT is a promising approach for myelin
imaging due to its specificity to substances with non-zero dipolar order.
However, tissue model quantification requires high resolution acquisitions in
excess of twenty minutes in length and so necessitates the use of motion
correction methods to prevent artefacts. In this work we combine our recent MT-mediated MRF sequence with the
DISORDER retrospective motion correction method. This new framework can acquire
and reconstruct high resolution motion-compensated 3D time-resolved data from
fingerprinting sequences. Semi‑quantitative MT and ihMT ratio maps as well as
quantitative maps of tissue parameters can be obtained from the resulting
images.
Introduction
Inhomogeneous
magnetization transfer (ihMT) contrast is sensitive to myelin1,2 and, as we have previously
shown, can be generated efficiently using multiband RF pulses either in a
steady-state3 or cyclic-state4 framework. In the example
shown in Figure 1, the sequence alternates between RF pulses with 1, 2 and 3
bands, constructed such that the on‑resonance flip angle never changes5. Tissues with no
significant MT or ihMT effect give constant signal throughout but
MT-mediated time-varying signal fluctuations are generated in most brain
tissues. When combined with randomized encoding, these time‑varying signals are
akin to an MRF sequence, and resulting data can be reconstructed using the
low-rank inversion (LRI) approach proposed by Assländer et al.6
However, 3D brain
volume acquisitions using these methods can exceed a duration of twenty
minutes, so motion correction techniques are required to prevent imaging
artefacts. Cordero‑Grande et al. proposed a motion estimation method
whereby partial k-space information provided by receiver coils is used to
estimate the position of an imaged object during multi-shot acquisitions7. This was
subsequently combined with randomized‑chequered Cartesian trajectories to boost
motion resolvability in the so-called DISORDER method8. In this work we show
that LRI MRF reconstruction and the Cartesian DISORDER motion
estimation/compensation framework can be combined for motion-tolerant and
well-conditioned reconstructions for our MT-MRF sequence.Methods
Although joint motion estimation and LRI is
conceptually possible, we have observed that for our application it is more efficient to split these subproblems. Motion
estimation can be accomplished as in conventional DISORDER. Then, motion
corrected MT-MRF datasets can be reconstructed by incorporating a low-rank
representation of a dictionary of signal evolutions UR (computed via SVD and shuffled to match k‑space
collection ordering) and motion operators Tθ (from estimated parameters θ) into a conjugate‑gradient sensitivity encoding
reconstruction:
$$[1] \ \hat{\textbf{x}}=\underset{\textbf{x}}{\arg\min}||\textbf{U}_{\textbf{R}}\textbf{FS}\textbf{T}_{θ}\textbf{x}-\textbf{y}||^{2}_{2}$$
x are singular values of a low-rank approximation
to the temporal signal evolution in each voxel, S are coil sensitivities, F represents a discrete Fourier transform and y is measured k-space data.
Experiments were conducted on a 1.5T Philips Ingenia
MR system using a bSSFP sequence with parameters: repetition time = 5.3ms, flip
angle = 29.5˚, pulse duration =
2ms, off-resonance frequency = 8.1kHz and root-mean-square B1 = 4μT.
Three
phantoms were scanned with FOV = 182×364×240mm (Cartesian sampling): MnCl2-doped
water (0.05mM concentration; no MT effect), bovine serum albumin (BSA; MT but
no ihMT effect) and prolipid 161 (PL161; strong ihMT effect), resulting in a scan time of 13.5 minutes (5 fully‑encoded volumes). Figure 2 illustrates sampling order of
the phase-encoding plane.
A healthy
adult male volunteer (aged 24) was scanned using the same acquisition
scheme as for the phantom experiment but FOV = 200×400×364mm and scan time was ~40 minutes
(16 fully‑encoded volumes) for a 2mm isotropic resolution to improve SNR. 1200 time-points were reconstructed (Figure 1),
making our method highly undersampled. One motion estimate was obtained approximately every
twenty seconds - sufficient for the involuntary motion considered here. Equation 1 was used to reconstruct all datasets before
time domain volumes at t1, t2 and t3 were combined to produce MT ratio
(MTR) and ihMT ratio (ihMTR) maps. Since the signals display temporal evolution,
quantification of tissue parameters can be achieved using dictionary fitting. However,
the ihMT signal model has nine free parameters that cannot all be estimated so,
similar to other studies9,10, we use a constrained fit but
for estimation of free pool T1f, semisolid fraction f and dipolar T1Ds.Results
Figure 3 summarizes
results from a phantom experiment. MT effects only exist in BSA and PL161, while
only PL161 provides a non-negligible ihMTR of ~20%.
Using DISORDER, estimates of translational and
rotational motion can be obtained from the data, prior to LRI and then
integrated into the reconstruction. Figure 4 shows equivalent traces alongside semi-quantitative metrics reconstructed with and without motion correction. Corresponding results from a constrained in vivo dictionary fit are in Figure 5. Though
MTR and f are less affected, ihMTR and dipolar estimates are heavily artefacted
before correction.Discussion
Figure 3
demonstrates the effectiveness of our proposed framework since reconstructed
signals show excellent agreement with results from a matched non-phase-encoded
phantom experiment.
Figure 4 illustrates that retrospective motion estimation is successful for our in vivo dataset. Resultant ratio
maps concur with the literature and previous simulations3,4: WM ihMTR~5-6%. The in vivo acquisition
length was not optimal but ensured sufficient data was acquired. We expect scan
time can be reduced to <30 minutes without significantly degrading image quality.
The
reconstruction resolves temporal modulation of the signal, so it can be used for
quantitative estimation as well as semi-quantitative MTR and ihMTR measurement.
Cramér-Rao lower bound calculations indicated that 3-4 parameters can be
simultaneously estimated using our current sequence. Since remaining parameters
are fixed, bias may exist though fitted values largely agree with the
literature: T1Ds is highest at 6.44±0.22ms in corticospinal
tracts11,12 but 2.79±0.12ms in GM,
whereas f is highest at 0.111±0.006 in
the corpus callosum13 but 0.056±0.003 in GM.Conclusion
We have presented
a general method for transient fingerprinting-style scans in a motion corrected
Cartesian framework. Further to the MT-based sequence shown here, this may
prove a useful technique for high resolution MRF experiments using 3D
acquisitions.Acknowledgements
This work
was supported by King’s College London & Imperial College London EPSRC
Centre for Doctoral Training in Medical Imaging [EP/L015226/1], by core funding
from the Wellcome/EPSRC Centre for Medical Engineering [WT203148/Z/16/Z] and by
the National Institute for Health Research (NIHR) Biomedical Research Centre
based at Guy’s and St Thomas’ NHS Foundation Trust and King’s College London
and/or the NIHR Clinical Research Facility. The views expressed are those of
the author(s) and not necessarily those of the NHS, the NIHR or the Department
of Health and Social Care.References
1. Girard, O. M. et al.
Magnetization transfer from inhomogeneously broadened lines (ihMT): experimental
optimization of saturation parameters for human brain imaging at 1.5 Tesla. Magn.
Reson. Med. 73, 2111-2121 (2015).
2. Ercan, E. et al.
Microstructural correlates of 3D steady-state inhomogeneous magnetization
transfer (ihMT) in the human brain white matter assessed by myelin water
imaging and diffusion tensor imaging. Magn. Reson. Med. 80, 2402-2414
(2018).
3. Malik, S. J., Teixeira,
R. P. A. G., West, D. J., Wood, T. C. & Hajnal, J. V. Steady‐state imaging
with inhomogeneous magnetization transfer contrast using multiband
radiofrequency pulses. Magn. Reson. Med. 83, 935-949 (2019).
4. West, D. J. et al.
Transient-State Inhomogeneous Magnetisation Transfer: Towards Magnetisation
Transfer Fingerprinting. Proceedings of the 28th Annual Meeting of ISMRM (2020).
5. West, D. J. et al. Magnetization Transfer-Mediated MR Fingerprinting. Manuscript In Preparation. (2020).
6. Assländer, J. et al. Low rank alternating direction method of multipliers reconstruction for MR fingerprinting. Magn. Reson. Med. 79, 83-96 (2018).
7. Cordero-Grande, L. et
al. Sensitivity Encoding for Aligned Multishot Magnetic Resonance
Reconstruction. IEEE Trans. Comput. Imaging 2, 266-280 (2016).
8. Cordero-Grande, L. et
al. Motion corrected MRI with DISORDER: Distributed and Incoherent Sample
Orders for Reconstruction Deblurring using Encoding Redundancy. Magn. Reson.
Med. 84, 713-726 (2020).
9. Hilbert, T. et al.
Magnetization Transfer in Magnetic Resonance Fingerprinting. Magn. Reson.
Med. 84, 128-141 (2019).
10. Cohen, O., Huang, S.,
Mcmahon, M. T., Rosen, M. S. & Farrar, C. T. Rapid and quantitative
chemical exchange saturation transfer (CEST) imaging with magnetic resonance
fingerprinting (MRF). Magn. Reson. Med. 80, 2449-2463 (2018).
11. Swanson, S. D. et al.
Molecular, dynamic and structural origin of inhomogeneous magnetization
transfer in lipid membranes. Magn. Reson. Med. 77, 1318-1328
(2017).
12. Varma, G. et al. In
vivo measurement of a new source of contrast, the dipolar relaxation time, T1D,
using a modified inhomogeneous magnetization transfer (ihMT). sequence. Magn.
Reson. Med. 78, 1362-1372 (2017).
13. Varma, G. et al.
Low duty-cycle pulsed irradiation reduces magnetization transfer and increases
the inhomogeneous magnetization transfer effect. J. Magn. Reson. 296,
60-71 (2018).
14. Mchinda, S. et al.
Whole brain inhomogeneous magnetization transfer (ihMT) imaging: Sensitivity
enhancement within a steady-state gradient echo sequence. Magn. Reson. Med.
79, 2607-2619 (2018).