Ali Caglar Özen1,2, Djaudat Idiyatullin3, Gregor Adriany3, Steve Jungst3, Naoharu Kobayashi3, Beth R. Groenke4, Michael Bock1, Michael Garwood3, and Donald R. Nixdorf4,5
1Deptartment of Radiology, Medical Physics, University Medical Center Freiburg, University of Freiburg, Freiburg, Germany, 2German Consortium for Translational Cancer Research Partner Site Freiburg, German Cancer Research Center (DKFZ), Heidelberg, Germany, 3Center for Magnetic Resonance Research and Department of Radiology, University of Minnesota, Minneapolis, MN, United States, 4Division of TMD & Orofacial Pain, School of Dentistry, University of Minnesota, Minneapolis, MN, United States, 5Department of Neurology and Radiology, Medical School, University of Minnesota, Minneapolis, MN, United States
Synopsis
Previous
studies showed that in dental MRI intraoral loop coils provide higher
signal-to-noise ratio (SNR) than extraoral coils. An intraoral dipole that fits
the dental arch can be used for reduced FOV and high transmit efficiency.
Besides, dipoles do not restrict tongue movement. The design approach is based on comparative FDTD field simulations. The best transmit
efficiency and homogeneity was achieved with a multi-wire curved dipole
antenna. Additional high-permittivity cap further improved the transmit field
inhomogeneity. When combined with extraoral flexible shielded loop resonators,
SNR was increased and the coupling between the coils was less than
-32dB.
Introduction
In dental MRI
submillimeter structures (e.g., small fractures within dental roots) must be
resolved. To achieve this high resolution within clinically acceptable
measurement times, the imaging FOV must be restricted to the target region that
can be realized by local RF excitation,
, a localized
receive sensitivity
, or a
combination of both. So far mainly extraoral surface coils and coil arrays have
been used (1–3).
An intraoral loop coil
with the coil plane orthogonal to
was developed, where the sensitive volume
included the most important dental structures (4). However, the loop obstructed the tongue
movement resulting in patient discomfort.
In
this study, we developed a novel intraoral dipole antenna, which does not
restrict tongue, and improved its performance based on the B_1 field
simulations. For the best-performing dipole, a prototype was constructed and
compared to the reference loop coil in phantom and in vivo measurements. The dipole was also combined with a
flexible shielded loop resonator (SLR) array to increase sensitivity.Methods
A new intraoral coil
design was realized with the following target features: providing a uniform
sensitivity over the teeth and surrounding tissues, having a penetration depth that
allows sensitive detection of the signal from dental roots, and minimizing coil
coupling with exterior Rx coil elements. A dipole design was chosen, and
different dipole designs were simulated using an electromagnetic (EM)
simulation tool based on finite-difference-time-domain (FDTD) approach
(Sim4Life v5.0, ZMT, Zurich, Switzerland). The prototype coils were tested on a
clinical 3T MRI system (Prisma, Siemens, Erlangen, Germany) in a homogeneous
phantom and a human volunteer.
5 dipoles with planar
structure (Fig. 1B-F) and 8 with multi-planar structure (one of them is shown
in Fig. 1G) were designed based on average human adult mandibular dimensions (5). The dipole arms are formed either by a 10
mm-wide ribbon, a single wire (Ø1.2 mm), or multiple wires separated by
5 mm. Multi-planar dipoles have additional wires (Ø1.2 mm) at a
distance of 8.6 mm to the central plane. To increase the effective
wavelength of the dipole, tuning inductors are used as shown in Fig. 1.
The transmit field
B1+(x,y,z) efficiency, i.e., transmit field amplitude
over conducted input power,
||B1+||/Pin
, were
compared between the reference loop coil and dipole antennae (Fig. 1A-G). Dielectric-coated
tips for dipole antennae were evaluated for improvement of field homogeneity
similar to (6).
Similar to loops and
dipoles (7,8) SLRs and dipoles are geometrically decoupled
when coil planes are parallel to each other. We propose to use the dipole with
high-εr cap as the Tx element, and a 4-channel SLR
array as the Rx element for optimal performance.
Based
on the SAR10g simulation results, power limits were set such that
the corresponding maximum SAR10g value was at least 20-fold less
than the input power that generates maximum SAR10g of 10 W/kg.Results
In
Table 1, the maximum
||B1+||/Pin along the axial profiles at
z=10mm
and z=25mm
planes, mean absolute deviation along the axial profile, peak
||B1+|| along the longitudinal profile at levels
y1 and
y2,
and longitudinal variation calculated as
100(1-||B1+(z2)||) are given. 7-wire dipoles have higher maximum
||B1+||/Pin values, whereas the ribbon based multi-planar
design performs slightly better in terms of longitudinal homogeneity. The
problem of the inhomogeneity along the axial profile is resolved with the
addition of a high-er cap, as shown in Table 1. According to the
simulations, high-
cap of 15 mm length results in a more
flat axial profile along the dipole arms (Fig. 2). SAR10g maps for
loop and dipole coil in the Duke anatomical model showed hotspots around feed
ports of the loop and dipoles, and SAR increase at the end of the dipole arms.
S-parameters are given
in Fig. 3C. Unloaded and loaded Q factors for loop, dipole, and SLR elements
are Qunloaded/Qloaded = 70.8/6.2, 8.2/3.9, 64.5/20.3,
respectively. SLR elements and dipole are decoupled by 33±4 dB when
positioned as illustrated in Fig. 3D, and the reflection coefficient of the
dipole is not affected from the nearby SLR elements.
In
Fig. 4B-F, in vivo images of the human mandible are shown in transverse and a
coronal plane. The transverse plane was positioned close to the intraoral coil
plane (z=4mm),
and the SNR values from dentine at the right first molar were measured and
found to be 14.8/4.6/40.5/141.7/189.6 for the loop, the dipole, SLR only, and
the SLR combined with the loop and the dipole, respectively.Discussion and Conclusion
We
demonstrated feasibility of using a dipole antenna as an intraoral coil and
improved the transmit efficiency and homogeneity of this coil. The intraoral
dipole allows free tongue movement. Transmission efficiency of the intraoral
loop and dipole are similar within 15 mm distance to the coil planes,
however, dipole outperforms the loop when the distance is larger than
20 mm. The combination of the intraoral dipole and the extraoral SLR Rx
array is required to achieve SNR comparable to or higher than the loop coil.Acknowledgements
We
gratefully acknowledge Lasby Fellowship awarded to A. C. Ozen by the School of
Dentistry of the University of Minnesota in 2019. We are thankful to CMRR
scientists and faculty for their help, especially to R. Lagore, J. Radder, Y.
Eryaman, A. Sadeghi-Tarakameh, A. Grant, and J. Thotland. We also thank J.
Fischer and S. Ilbey of University Medical Center Freiburg for their help with
the measurement setup.References
1. Bracher AK, Hofmann C, Bornstedt
A, et al. Ultrashort echo time (UTE) MRI for the assessment of caries lesions.
Dentomaxillofacial Radiol. 2013;42:20120321 doi: 10.1259/dmfr.20120321.
2. Prager M,
Heiland S, Gareis D, Hilgenfeld T, Bendszus M, Gaudino C. Dental MRI using a
dedicated RF-coil at 3 Tesla. J. Cranio-Maxillofacial Surg. 2015;43:2175–2182
doi: 10.1016/j.jcms.2015.10.011.
3. Sedlacik J,
Kutzner D, Khokale A, et al. Optimized 14+1 receive coil array and position
system for 3D high-resolution MRI of dental and maxillomandibular structures.
Dentomaxillofacial Radiol. 2015;45:2–7 doi: 10.1259/dmfr.20150177.
4. Idiyatullin D,
Corum C a, Nixdorf DR, Garwood M. Intraoral approach for imaging teeth using
the transverse B 1 field components of an occlusally oriented loop coil. Magn.
Reson. Med. 2014;72:160–165 doi: 10.1002/mrm.24893.
5. Raberin M,
Laumon B, Martin JL, Brunner F. Dimensions and form of dental arches in
subjects with normal occlusions. Am J Orthod Dentofac. Orthop 1993;104:67–72
doi: 10.1016/0889-5406(93)70029.
6.
Sadeghi-Tarakameh A, Jungst S, Wu X, et al. A New Coil Element for Highly-Dense
Transmit Arrays : An Introduction to Non-Uniform Dielectric Substrate (NODES)
Antenna. In: Proc. Intl. Soc. Mag. Res. Med. 27. Montreal, Quebec; 2019. p.
732.
7. Eryaman Y, Guerin B, Kosior R,
Adalsteinsson E, Wald LL. Combined loop +
dipole arrays for 7 T brain imaging. In: Proc. Intl. Soc. Mag. Reson. Med. 21.
Salt Lake City, Utah; 2013. p. 393.
8. Ertürk MA,
Raaijmakers AJE, Adriany G, Uğurbil K, Metzger GJ. A 16-channel combined
loop-dipole transceiver array for 7 Tesla body MRI. Magn. Reson. Med.
2017;77:884–894 doi: 10.1002/mrm.26153.
9. IEC
60601-2-33:2010+AMD1:2013+AMD2:2015 Medical electrical equipment—Part 2-33:
particular requirements for the basic safety and essential performance of
magnetic resonance equipment for medical diagnosis.