Florian Wiesinger1, Timothy Deller2, Floris Jansen2, Jose de Arcos Rodriguez1, Ronny R Buechel3, Philipp A Kaufmann3, and Edwin EGW ter Voert3
1GE Healthcare, Munich, Germany, 2GE Healthcare, Waukesha, WI, United States, 3Department of Nuclear Medicine, University Hospital Zurich, Zurich, Switzerland
Synopsis
Respiratory
motion correction is a long-standing problem in hybrid PET/MR imaging with many
partial solutions. Here we present a
novel method based on extracting respiratory motion information from the PET
data using the ultra-fast
listmode reconstruction framework. Doing so a highly accurate respiratory waveform
derived from the inside of the body (i.e. lung liver interface) is obtained for
free without requiring an extra motion sensor or complicating the PET/MR imaging
workflow.
Introduction
PET/MR is a hybrid imaging modality allowing simultaneous acquisition of
MR and PET information, providing unique opportunities in terms of joint image
reconstruction and motion correction. In current
methods often MR is used to extract motion information which is then used for
PET and/or MR motion gating and/or correction.
This requires dedicated MR sequences and limits the availability of
motion information to the duration of the MR scan. Alternatively, also auxiliary motion sensors
(respiratory belt, optical camera, …) are used.
However, these are typically expensive, complicate workflow, require cross-calibration,
might drift, or pick-up additional motions.
Recently Spangler-Bickell et al1 introduced a PET
reconstruction framework which reconstructs the PET listmode raw data into
dynamic PET frames with high temporal resolution of ≤ 1 sec (in addition to standard
PET images). Here we describe the
adaptation of this method for MR and PET respiratory motion correction. Methods
All scans were performed on a GE SIGNA PET/MR scanner (GE Healthcare,
Chicago, IL) using either [13N]-Ammonia or [18F]-FDG PET
tracers. The PET/MR patient imaging (with
signed informed consent by the patients and ethics approval by the local ethics
commission) focused on the chest region and was performed in free breathing. PET data were collected continuously
throughout the duration of the bed position and recorded sequentially in the
listmode file. The ultra-fast
listmode reconstruction framework1 was used to
reconstruct temporal PET frames at 1.0 sec temporal resolution for the whole
duration of the PET bed position. A
respiratory waveform (RW) was then extracted from the high-temporal resolution
PET images using either #1) a cylindrical pencil beam navigator across the
lung-liver interface (RWNAV), or #2) Principal Component Analysis (PCA)
of the dominant motion component (RWPCA). The PET-derived respiratory waveform was then
used for retrospective soft-gating of both MR and PET data.
Simultaneous to the PET also MR scanning was
performed including six repetitions of a 3D radial Zero TE (ZTE) sequence for structural
lunging imaging2 [FOV=400x400x288mm3, res=1.6x1.6x1.6mm3, FA=2deg,
BW=±62.5kHz, time=71sec]. The ZTE
sequence contained multiple sequential repetitions (m) to allow soft-gating of
the acquired data into different respiratory phases (n) according to: wn,m=exp[-(bm-refn)2/α2]; with bm the respiratory amplitude of the mth spoke
relative to a reference position (refn) and α defining a soft acceptance window. The soft-gated
k-space data were then reconstructed into a dynamic 4D ZTE image using 3D gridding.Results
The animated Figures 1 and 2 show example results of a patient
injected with [13N]-Ammonia (injected dose 200-600 MBq dependent on BMI and stress/rest) where the ZTE imaging started ~8min
after the PET tracer injection. Figure 1
illustrates the dynamic PET images at 1 sec temporal resolution (top). The normalized respiratory waveforms (bottom)
obtained using the pencil beam navigator and the PCA method demonstrate excellent
agreement. The position
and duration of the six ZTE acquisitions are indicated as black thick lines
along the time axis. This patient
demonstrates a deep, regular diaphragmatic breathing pattern. Figure 2 illustrates corresponding ZTE images. Because of the deep diaphragmatic breathing
the uncorrected images (left) show strong motion blurring especially at the
lung-liver interface. Soft-gated respiratory binning (7 phases, 2nd column)
resolves the diaphragmatic breathing cycle into 7 phases. Most of the data are acquired during end-expiration
(3rd column) which also provides the sharpest image. Its Maximum Intensity Project (MIP, right)
depicts the vascular anatomy and lung lesions in fine detail.
The animated Figures 3 and 4 illustrate a patient injected with [18F]-FDG
(injected dose = 250 MBq) where the ZTE imaging started ~100min after the PET tracer injection. Again, the navigator and PCA-based respiratory
waveforms (Fig. 3) illustrate excellent agreement. Because of the
shallow chest breathing the uncorrected ZTE (Fig. 4, left) shows only benign
motion blurring which is also demonstrated by the respiratory binning (7
phases, 2nd column). The end-expiratory
phase ZTE image (3rd column) and its MIP (right) are illustrated as
well.Discussion
Respiratory motion correction is a long-standing
problem in hybrid PET/MR imaging with many partial solutions. Here we present a novel method based on
extracting respiratory motion information from the PET data using the ultra-fast listmode
reconstruction framework. Doing so a highly accurate respiratory waveform derived from the
inside of the body (i.e. lung liver interface) is obtained for free without
requiring an extra motion sensor or complicating the PET/MR imaging workflow. The obtained respiratory waveform can then be used
to soft-gate PET and MR images to the very same breathing phase and thereby assure
perfect geometric alignment in presence of respiratory motion.Acknowledgements
No acknowledgement found.References
- Spangler-Bickell et al (2020). Ultra-Fast List-Mode Reconstruction of Short PET Frames and Example Applications. Journal of Nuclear Medicine.
- Gibiino et al (2015). Free-breathing, zero-TE MR lung imaging. Magnetic Resonance Materials in Physics, Biology and Medicine, 28(3), 207-215.