Synopsis
Strong spherical diffusion encoding enables visualization of isotropically restricted regions in the human brain, believed to resemble densely packed cells. As this forces imaging in a low SNR regime, visualization has been limited to a low resolution to avoid deleterious signal bias caused by the noise floor. In this work, we propose a novel method based on super-resolution reconstruction to enable high-resolution visualization of isotropically restricted diffusion in human brain in vivo. We show that our method is superior over conventional methods with acquisition times that are compatible with clinical routine.
Introduction
Diffusion
MRI with spherical tensor encoding at ultrahigh b-values emphasizes regions that
are isotropically restricted (Fig. 1). Brain imaging shows such regions in the
cerebellar cortex, which are affected in
diseases as spinocerebellar ataxis and Alzheimer disease1,2. However, imaging with ultrahigh b-values causes
low SNR which can bias the signal due to the rectified noise floor. This
problem cannot be alleviated by averaging multiple magnitude images3. Current methods employing ultrahigh b-values
resort to low spatial resolution to avoid the noise floor (Fig. 2). In this
work, we demonstrate unprecedented spatial resolution with spherical diffusion
encoding at ultrahigh b-values by the use of super-resolution reconstruction. Results
show that it is vastly superior compared to direct sampling in terms of
contrast and accuracy.Principle
When diffusion encoding is performed at
very high b-values, the signal that remains can be attributed to restrictions
that cause the apparent diffusivity to approach zero4,5. For conventional diffusion encoding (along a single direction at a
time) the remaining signal may originate from any structure that happens to be
restricted along the diffusion encoding direction, including water trapped in
cell bodies or inside axons. However, with spherical diffusion encoding only signal
from compartments restricted in all directions remains (Fig.1). Therefore,
whatever signal remains for spherical encoding at high b-values, must be
isotopically restricted, and we refer to it as the “dot fraction” 6,7.Methods
Imaging the dot fraction
We
assume a compartment with signal fraction $$$f_{\mathrm{dot}}$$$ and
isotropic diffusivity $$$D_{\mathrm{dot}}$$$ close to zero, accompanied
by a fraction of other tissue $$$(f_{\mathrm{other}} = 1 - f_{\mathrm{dot}})$$$ with
non-zero isotropic diffusivity. The
diffusion weighted signal $$$S(b)$$$ when using spherical tensor encoding is then given by $$S(b) = S0\cdot(f_{\mathrm{dot}}\exp(-b\cdot D_\mathrm{dot}) + f_\mathrm{other}\exp(-b\cdot D_\mathrm{other})),\qquad(1) $$ where $$$S0$$$ is the non-diffusion weighted signal. When
approaching ultrahigh b-values, the
contribution from other compartments is suppressed $$$(\exp(-b_{\mathrm{high}}\cdot D_{\mathrm{other}}) \approx 0)$$$ whereas
signal in the dot compartment remains $$$(b_{\mathrm{high}}\cdot D_{\mathrm{dot}} \approx 0 \rightarrow \exp(-b_{\mathrm{high}}\cdot D_{\mathrm{dot}}) \approx 1),$$$which simplifies
Eq. 1 to6,7 $$S(b) \approx S0\cdot f_{\mathrm{dot}}.\qquad(2) $$ The resulting
map is weighted by both T2 and $$$f_{\mathrm{dot}}$$$, and will show high intensity where diffusivity is isotropically
restricted. It
should be noted that $$$f_{\mathrm{dot}}$$$ only sets an upper limit on the dot fraction7.
Data acquisition
A
healthy volunteer was scanned on a Siemens 3T-Prisma (Siemens Healthineers,
Germany) using a 20-channel head and neck coil. We used a prototype spin-echo
EPI pulse sequence enabling spherical tensor encoding8-10. To enable super-resolution reconstruction, we
performed imaging with eight rotations of the FOV at resolution 1.6x1.6x7.2 mm3
(all shared the same phase encoding direction, Fig. 3). For each FOV rotation,
we acquired 13 repetitions of spherical b-tensors at b = 4 ms/μm2. We used TR/TE=4.2s/120ms, giving a total
acquisition time of 9:31 min. For reference, 15 repetitions of the
same b-tensors were acquired directly at 1.6x1.6x1.6 mm3 at TR/TE=14.2s/120ms
with an acquisition time of 9:18 min. All
data were denoised using Marchenko-Pastur PCA11,12, averaged over repetitions and assessed
qualitatively. The gray-to-white matter signal ratio was calculated between gray matter voxels in the
cerebellar cortex and white matter voxels in the cerebellum.
Super-resolution reconstruction
We reconstructed the acquired images
to a resolution of 1.6x1.6x1.6 mm3 using super-resolution
reconstruction. This approach produces a high-resolution image from multiple
low-resolution images, which boosts SNR and thereby suppresses noise-floor
effects (Fig. 3). We used a regularized least-squares solution using the
pseudo-inverse13,14 $$ \mathbf{x = ( A^{T}A + \lambda I)^{-1}A^{T}} \qquad(3)$$ where $$$\bf x$$$ is the high-resolution image to recover, $$$\bf y$$$ are the measured low-resolution signal samples, $$$\bf I$$$ is the identity matrix, $$$\lambda$$$ is a scalar weight determining the regularization strength set to 0.05 and $$$\bf A$$$ is the sampling matrix, which was
constructed from the geometric imaging parameters.Results
Super-resolution reconstruction shows a
marked improvement over direct sampling in visualization of the cerebellar
cortex at high b-values and high resolution, for similar acquisition times
(Fig. 4). Direct sampling leads to poor image contrast throughout the brain,
whereas the super-resolution reconstruction exhibits vastly improved contrast.
Most prominent are the tightly packed granule cells of the cerebellar cortex,
where the gray-to-white matter signal ratio is 1.06 for direct sampling and
1.82 for super-resolution reconstruction. Discussion and conclusions
Combining spherical tensor encoding,
ultrahigh b-values, and super-resolution reconstruction enables high-resolution imaging of the dot fraction with improved contrast compared to conventional imaging. We
demonstrated this in whole-brain in vivo at a resolution of 1.6x1.6x1.6 mm3
on a clinical scanner. Previous acquisitions with spherical encoding and
ultra-high b-values were done at a resolution of 4.0x4.0x4.0 mm3 on
a connectome scanner7.
Most
strikingly, a high signal can be observed in the cerebellar cortex which is
populated by densely packed granule cells (Fig. 5)7,15. This observation supports the hypothesis that granule
cells and/or their extracellular space is highly restrictive to diffusion in all spatial directions. Therefore,
we speculate that the proposed image contrast may be used to detect diseases
affecting these cells. However, the contrast may be confounded by T2
shine-through (compare S(b=4) vs. $$$f_{\mathrm{dot}}$$$ in Fig. 5), which warrants further
investigation.
In conclusion, we have presented and
demonstrated a novel methodology that facilitates isotropically restricted
fraction imaging at unprecedented spatial resolution, clinically relevant
acquisition times and clinical MRI hardware.Acknowledgements
We thank Siemens Healthcare (Erlangen,
Germany) for access to the pulse sequence programming environment.References
[1] G. Xia, K. N. McFarland, K. Wang, P. S.
Sarkar, A. T. Yachnis, and T. Ashizawa, “Purkinje cell loss is the major brain
pathology of spinocerebellar ataxia type 10.,” Journal of neurology,
neurosurgery, and psychiatry, vol. 84, no. 12. pp. 1409–1411, Dec. 2013,
doi: 10.1136/jnnp-2013-305080.
[2] Y. Fukutani, N. J. Cairns, M. N. Rossor,
and P. L. Lantos, “Purkinje cell loss and astrocytosis in the cerebellum in
familial and sporadic Alzheimer’s
disease.,” Neurosci. Lett., vol. 214, no. 1, pp. 33–36, Aug. 1996, doi:
10.1016/0304-3940(96)12875-5.
[3] H. Gudbjartsson and S. Patz, “The Rician
distribution of noisy MRI data.,” Magn. Reson. Med., vol. 34, no. 6, pp.
910–914, Dec. 1995, doi: 10.1002/mrm.1910340618.
[4] D. C. Alexander, “Axon radius
measurements in vivo from diffusion MRI: A feasibility study,” Proc. IEEE
Int. Conf. Comput. Vis., 2007, doi: 10.1109/ICCV.2007.4409141.
[5] G. J. Stanisz, G. A. Wright, R. M.
Henkelman, and A. Szafer, “An analytical model of restricted diffusion in
bovine optic nerve,” Magn. Reson. Med., vol. 37, no. 1, pp. 103–111,
1997, doi: 10.1002/mrm.1910370115.
[6] B. Dhital, E. Kellner, V. G. Kiselev,
and M. Reisert, “The absence of restricted water pool in brain white matter.,” Neuroimage,
vol. 182, pp. 398–406, Nov. 2018, doi: 10.1016/j.neuroimage.2017.10.051.
[7] C. M. W. Tax, F. Szczepankiewicz, M.
Nilsson, and D. K. Jones, “The dot-compartment revealed? Diffusion MRI with
ultra-strong gradients and spherical tensor encoding in the living human
brain,” Neuroimage, vol. 210, p. 116534, 2020, doi:
https://doi.org/10.1016/j.neuroimage.2020.116534.
[8] F. Szczepankiewicz, J. Sjölund, F.
Ståhlberg, J. Lätt, and M. Nilsson, “Tensor-valued diffusion encoding for
diffusional variance decomposition (DIVIDE): Technical feasibility in clinical
MRI systems,” PLoS One, vol. 14, no. 3, p. e0214238, Mar. 2019,
[Online]. Available: https://doi.org/10.1371/journal.pone.0214238.
[9] F. Szczepankiewicz, C.-F. Westin, and M.
Nilsson, “Maxwell-compensated design of asymmetric gradient waveforms for tensor-valued
diffusion encoding,” Magn. Reson. Med., vol. 82, no. 4, pp. 1424–1437,
Oct. 2019, doi: https://doi.org/10.1002/mrm.27828.
[10] J. Sjölund, F. Szczepankiewicz, M.
Nilsson, D. Topgaard, C. F. Westin, and H. Knutsson, “Constrained optimization
of gradient waveforms for generalized diffusion encoding,” J. Magn. Reson.,
vol. 261, pp. 157–168, 2015, doi: 10.1016/j.jmr.2015.10.012.
[11] L. Cordero-Grande, D. Christiaens, J.
Hutter, A. N. Price, and J. V Hajnal, “Complex diffusion-weighted image
estimation via matrix recovery under general noise models,” Neuroimage,
vol. 200, pp. 391–404, 2019, doi: https://doi.org/10.1016/j.neuroimage.2019.06.039.
[12] J.-D. Tournier et al., “MRtrix3: A
fast, flexible and open software framework for medical image processing and visualisation.,” Neuroimage, vol.
202, p. 116137, Nov. 2019, doi: 10.1016/j.neuroimage.2019.116137.
[13] M. Elad and A. Feuer, “Restoration of a single superresolution image from several blurred, noisy, and undersampled measured images,” IEEE Trans. Image Process., vol. 6, no. 12, pp. 1646–1658, 1997, doi: 10.1109/83.650118.
[14] A. N. Tikhonov and V. Y. Arsenin, Solutions of ill-posed problems. 1977.
[15] K. Amunts et al., “BigBrain: An
ultrahigh-resolution 3D human brain model,” Science (80-. )., vol. 340,
no. 6139, pp. 1472–1475, 2013, doi: 10.1126/science.1235381.