Sisi Li1, Yishi Wang2, Zhangxuan Hu3, Zhe Zhang4, Bing Wu3, and Hua Guo1
1Center for Biomedical Imaging Research, Beijing, China, 2Philips Healthcare, Beijing, China, 3GE Healthcare, MR Research China, Beijing, China, 4China National Clinical Research Center for Neurological Diseases, Beijing Tiantan Hospital, Capital Medical University, Beijing, China
Synopsis
Diffusion tensor imaging (DTI) holds
great potential to aid the diagnosis of spinal cord pathologies. Single shot
EPI (SS-EPI) is mostly implemented for DTI but is significantly limited in
clinics by susceptibility inhomogeneity-induced distortions. This is especially
problematic in the thoracic region for increased inhomogeneities near lungs. To
solve this problem, multi-shot EPI and reduced-FOV methods have been proposed.
However, these methods cannot remove distortions completely. In this study, we
achieved high-fidelity DTI of the thoracic spinal cord using a distortion-free
MS-EPI technique, Point-Spread-Function Encoded EPI (PSF-EPI). Both the
efficacy of PSF-EPI in distortion correction and quantitative reproducibility
were evaluated.
Introduction
Diffusion tensor imaging (DTI) of the
spinal cord holds great potential in the detection and evaluation of diverse
spinal cord pathologies1-4.
However, this can be technically challenging due to the small size of the cord,
physiological motions and field inhomogeneity surrounding the spine. In
particular, this is further confounded in the thoracic region by the more
complicated susceptibility inhomogeneities due to air-bone interfaces near the
pulmonary parenchyma1 as well as increased motions including cardiac, respiratory and CSF
flow5. These factors necessitate a robust high-fidelity technique with
high resolution and practical acquisition efficiency for thoracic DTI.
The traditional scheme for DTI is single
shot EPI (SS-EPI) which is prone to susceptibility-induced distortions. This is
especially problematic for cases with spinal canal narrowing, such as disk herniation, spondylolisthesis and
trauma1. To remedy this problem,
advanced EPI-based methods, including multi-shot EPI (MS-EPI) methods such as readout-segmented
EPI (RS-EPI)6, interleaved EPI (iEPI)7 and reduced FOV (rFOV) methods such as zonal oblique multislice EPI
(ZOOM-EPI)8 have been proposed. However, these methods still cannot remove
distortions completely and might sacrifice the scan time or spatial coverage. To
further reduce the distortions, ZOOM-EPI and RS-EPI were combined and
implemented for thoracic DTI by using 7 blinds with 1/3 FOV 9.
This study aimed to achieve high-fidelity,
fast DTI of the thoracic spinal cord using a distortion-free multi-shot EPI
technique, Point-Spread-Function-Encoded EPI (PSF-EPI)10-12. By
using 8 shots, the acquisition time can be reduced to 6 minutes with 6
diffusion directions and 2 number of signal averages.
Methods
In PSF-EPI, an additional phase encoding
gradient is exerted to the conventional phase encoding (PE) direction of 2D EPI
in a multi-shot manner (Fig.1A). This generates a 3D k-space dataset consisting
of $$$k_x$$$, $$$k_y$$$ and $$$k_{psf}$$$. At a
given $$$k_y$$$ encoding step, all signals across the PSF
encoding dimension share the same echo time. Thus images from $$$k_x$$$-$$$k_{psf}$$$ are distortion-free. Validation of
distortion-free imaging are shown in Fig.1B. Here, tilted-CAIPI acquisition
scheme and parital Fourier undersampling were also integrated for high
acceleration (>20×).
The
scans were performed on a Philips 3T Ingenia CX scanner (Philips Healthcare,
Best, The Netherlands) and a GE 3T Signa Premier scanner (GE Healthcare,
Waukesha, Wisconsin). To evaluate the efficacy of
various EPI-based methods in distortion reduction, comparison experiments were
performed on both scanners. On the Philips scanner, PSF-EPI was compared with ZOOM-EPI using similar scan time as
well as SS-EPI (Table 1, scan 2-5). On the GE
scanner, PSF-EPI was compared with MUSE, MUSE+FOCUS, SS-EPI+FOCUS
and
SS-EPI with parallel imaging (Table 1, scan 6-10), where MUSE is a navigator-free iEPI
technique and FOCUS is a rFOV technique using ZOOM pulse.
The specific imaging parameters are summarized in Table 1. The time of the
calibration scan for PSF-EPI was excluded, which took about 30s. No triggering
or gating was used for any of the sequences. Seven healthy volunteers were recruited. This study was
approved by the Institutional Review Board and written informed consent was
obtained from each participant.
The
tensor fitting and measurement of DTI metrics were performed in FSL (FMRIB,
Oxford, UK) and DTI studio13. We also used Bland-Altman plots to test the quantitative
reproducibility through two repeated scans.Results
Fig.2
demonstrates DTI of the thoracic spinal cord using full
FOV PSF-EPI (Table 1, scan 3), rFOV ZOOM-EPI (scan 4), and full FOV SS-EPI with RPE=4 (scan 5).
Compared with full FOV SS-EPI, ZOOM-EPI method mitigated distortions by reducing the FOV along PE by 2.6 fold, though visible
distortions remained around intervertebral disks. In contrast, the PSF-EPI
images show high consistency with the anatomical reference through the entire
spinal cord. The major structural boundaries (red lines) align well with the
T2W anatomical reference. Additionally, the measured quantitative diffusivity
values from PSF-EPI generally agree with previously reported values14.
Fig.3A shows the PSF-EPI results from another
healthy volunteer (Table 1, scan 2). The acquisition time is around 6 minutes. As
shown in Fig. 3B, the DTI measures are analogous to the reference values from
previous sagittal spinal cord diffusion studies5,15,16. Additionally,
the reproducibility of quantitative measurement was also tested. As shown in
the Bland-Altman plots, the DTI metrics differences demonstrate negligible bias
or trend.
In Fig.4, SS-EPI with RPE=2
demonstrates similar image quality with that of SS-EPI+FOCUS (1/2 FOV). In
comparison, 4-shot MUSE images show reduced distortions while 4-shot MUSE+FOCUS
(1/2 FOV) images show further improved fidelity though with residual artifacts possibly
due to the failure in phase correction. Particularly, the latter images are close
to 8-shot PSF-EPI (scan 8) images. Given above, the PSF-EPI method demonstrated
high fidelity with acceptable SNR and scan time (around 6.5 minutes).Discussion and Conclusion
In this study, we demonstrated the
feasibility of high-resolution DTI for thoracic spinal cord using tilted-CAIPI accelerated PSF-EPI
sequence and compared its efficacy in distortion correction with multiple
advanced EPI-based techniques. PSF-EPI showed similar performance
of distortion correction on both Philips and GE scanners. With
clinically practical acquisition time, the 8-shot PSF-EPI results show high
anatomical fidelity with reasonable SNR levels. Additionally, the results of quantitative assessments indicate acceptable reproducibility of PSF-EPI.
For future work, patient group that includes
various pathologies will be recruited for better clinical evaluation.Acknowledgements
No acknowledgement found.References
1. Andre JB,
Bammer R. Advanced Diffusion-Weighted Magnetic Resonance Imaging Techniques of
the Human Spinal Cord. Topics in Magnetic Resonance Imaging Tmri
2010;21(6):367-378.
2. Ellingson
BM, Cohen-Adad J. Diffusion-Weighted Imaging of the Spinal Cord. Quantitative
MRI of the Spinal Cord: Elsevier; 2014. p 123-145.
3. Budzik J-F,
Balbi V, Le Thuc V, Duhamel A, Assaker R, Cotten A. Diffusion tensor imaging
and fibre tracking in cervical spondylotic myelopathy. European radiology
2011;21(2):426-433.
4. Maier SE.
Examination of Spinal Cord Tissue Architecture with Magnetic Resonance
Diffusion Tensor Imaging. Neurotherapeutics 2007;4(3):453-459.
5. Zaharchuk
G, Saritas E, Andre J, et al. Reduced field-of-view diffusion imaging of the
human spinal cord: comparison with conventional single-shot echo-planar
imaging. American journal of neuroradiology 2011;32(5):813-820.
6. Holdsworth
SJ, Skare S, Newbould RD, Bammer R. Robust GRAPPA-Accelerated
Diffusion-Weighted Readout-Segmented (RS)-EPI. Magnetic Resonance in Medicine
2009;62(6):1629-1640.
7. Bammer R,
Augustin M, Prokesch RW, Stollberger R, Fazekas F. Diffusion‐weighted imaging
of the spinal cord: interleaved echo‐planar imaging is superior to fast
spin‐echo. Journal of Magnetic Resonance Imaging: An Official Journal of the
International Society for Magnetic Resonance in Medicine 2002;15(4):364-373.
8. Wheeler-Kingshott
CA, Hickman SJ, Parker GJ, et al. Investigating cervical spinal cord structure
using axial diffusion tensor imaging. Neuroimage 2002;16(1):93-102.
9. Holdsworth
S, Skare S, O'Hallaran R, Bammer R. Reduced-FOV diffusion imaging with ZOnal
Oblique Multislice (ZOOM) combined with readout-segmented (RS)-EPI. Proc Intl
Soc Mag Reson Med, 2010. p. 1611.
10. Robson MD,
Gore JC, Constable RT. Measurement of the point spread function in MRI using
constant time imaging. Magnetic resonance in medicine 1997;38(5):733-740.
11. Dong Z, Wang
F, Reese TG, et al. Tilted‐CAIPI for highly accelerated distortion‐free EPI
with point spread function (PSF) encoding. Magnetic resonance in medicine
2019;81(1):377-392.
12. Zeng H, Constable
RT. Image distortion correction in EPI: comparison of field mapping with point
spread function mapping. Magnetic Resonance in Medicine: An Official Journal of
the International Society for Magnetic Resonance in Medicine
2002;48(1):137-146.
13. Jiang H, van
Zijl PCM, Kim J, Pearlson GD, Mori S. DtiStudio: Resource program for diffusion
tensor computation and fiber bundle tracking. Computer Methods and Programs in
Biomedicine 2006;81(2):106-116.
14. Andre JB,
Zaharchuk G, Saritas E, et al. Clinical evaluation of reduced field-of-view
diffusion-weighted imaging of the cervical and thoracic spine and spinal cord.
AJNR Am J Neuroradiol 2012;33(10):1860-1866.
15. Jeong EK, Kim
SE, Guo J, Kholmovski EG, Parker DL. High‐resolution DTI with 2D interleaved
multislice reduced FOV single‐shot diffusion‐weighted EPI (2D ss‐rFOV‐DWEPI).
Magnetic resonance in medicine 2005;54(6):1575-1579.
16. Kim
T, Zollinger L, Shi X, et al. Quantification of diffusivities of the human
cervical spinal cord using a 2D single-shot interleaved multisection inner
volume diffusion-weighted echo-planar imaging technique. American journal of
neuroradiology 2010;31(4):682-687.