Daeun Kim1, Stephen F. Cauley2, Krishna S. Nayak1, Richard M. Leahy1, and Justin P. Haldar1
1Electrical and Computer Engineering, University of Southern California, Los Angeles, CA, United States, 2Radiology, A. A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Charlestown, MA, United States
Synopsis
MRI acquisitions often incidentally excite spatial regions
that are not interesting for the application. This unnecessary magnetization
can lead to artifacts and/or prolonged acquisitions. We propose a novel virtual-coil approach,
called region-optimized virtual (ROVir) coils, that can localize signal from an
ROI and/or suppress signal from unwanted spatial regions while also providing
coil compression. This is achieved using
optimal beamforming principles (without requiring modification of pulse
sequences or imaging hardware), and can be applied directly to k-space data,
which enables simplified image reconstruction.
We illustrate ROVir with reduced-FOV imaging, demonstrating capabilities
to suppress aliasing artifacts from outside the nominal FOV.
Introduction
Typical MRI experiments
often incidentally excite spatial regions that are not interesting for the
application. Excitation of uninteresting
regions can complicate the design of efficient imaging methods, leading to
artifacts and/or acquisitions that are longer than necessary.
We propose a novel
approach to localizing the signal from a region-of-interest (ROI) and/or
suppressing the signal from uninteresting spatial regions. While there are many hardware- and sequence-based
approaches that address this problem, the proposed approach instead uses the
beamforming capabilities1,2 of a phased array of receiver coils to
enable the suppression of unwanted signals without modifying the acquisition
hardware or pulse sequence. The proposed approach linearly mixes the signal
from the original receiver coils to form a set of coil-compressed region-optimized
virtual (ROVir) coils that are optimal in the sense that they maximize a
signal-to-interference ratio (SIR) metric. Importantly, unlike existing
beamforming approaches that use different coil combinations at different
spatial locations and can only be used for post-processing of
already-reconstructed images3-9, ROVir uses a form of
spatially-invariant beamforming (“sensor-domain
beamforming”) that can be applied directly to raw k-space data prior to image reconstruction.
This enables simplified image reconstruction.
In addition, unlike existing coil compression approaches that only
attempt to maximize signal10-11, ROVir is also capable of actively suppressing
unwanted signal. Finally, unlike recent similar sensor-domain approaches12,
ROVir possesses mathematical optimality.Theory
We consider the standard virtual coil paradigm, in which original
multi-channel k-space data, $$$d_l(\mathbf{k})$$$,
for $$$l=1,\dots,N_c$$$ from $$$N_c$$$ original coils are linearly transformed into data $$$v_j(\mathbf{k})$$$ from a smaller number of $$$N_v$$$ virtual coils according to $$$v_j(\mathbf{k})=\sum_{l=1}^{N_c} w_{lj} d_l(\mathbf{k})$$$ for $$$j=1,\dots,N_v$$$. To achieve both coil-compression and signal localization/suppression, ROVir
chooses the $$$\left\{ w_{lj} \right\}$$$ linear coil-combination coefficients to
maximize an SIR metric, which is defined by the ratio between the signal energy
from the spatial region of interest and the interference energy from unwanted
spatial regions. After some
manipulation, it can be shown that this is equivalent to: $$ \text{SIR}(\mathbf{w}_j) \triangleq \frac{\mathbf{w}_j^H \mathbf{A} \mathbf{w}_j}{\mathbf{w}_j^H \mathbf{B} \mathbf{w}_j},$$ where $$$\mathbf{w}_j \in \mathbb{C}^{N_c}$$$ is the vector of weighting coefficients $$$ \left \{ w_{lj} \right \}$$$. The matrices $$$\mathbf{A} \in \mathbb{C}^{N_c \times N_c}$$$ and $$$\mathbf{B} \in \mathbb{C}^{N_c \times N_c}$$$ are constructed as inter-coil correlation
matrices corresponding to the spatial region of interest $$$\Omega$$$ and the unwanted spatial regions $$$\Gamma$$$, respectively, defined by: $$\mathbf{A} \triangleq \int_\Omega \mathbf{g}(\mathbf{x}) \mathbf{g} (\mathbf{x})^{H} d\mathbf{x}, ~~\text{and}~~\mathbf{B} \triangleq \int_\Gamma \mathbf{g}(\mathbf{x}) \mathbf{g} (\mathbf{x})^{H} d\mathbf{x},$$ with $$$ \mathbf{g} (\mathbf{x}) \in \mathbb{C}^{1\times N_c} \triangleq \left[ g_1(\mathbf{x}),\dots, g_{N_c}(\mathbf{x}) \right]$$$ where $$$ g_l(\mathbf{x})$$$ is the sensitivity weighted image for the $$$ l $$$th coil.
The optimal weight vectors can be found by a simple generalized eigenvalue
decomposition for $$$\mathbf{A}$$$ and $$$\mathbf{B}$$$1-2.
The generalized eigenvectors with the largest eigenvalues will have
maximal SIR, and SIR-optimal ROVir coils are obtained by keeping the top-
eigenvectors. In practice, the $$$\mathbf{A}$$$ and $$$\mathbf{B}$$$ matrices can be computed
using the same kind of prescan/calibration data that is already used for
parallel imaging. Once the optimal ROVir weights are calculated, they can then
be applied directly to raw k-space data of subsequent scans. Figure 1 provides a representative
illustration of ROVir.
Methods
Brain imaging example: We applied ROVir to in vivo T1-weighted brain imaging data acquired with
a 32-channel array, and performed comparisons against conventional SVD and ROI-focused
SVD (ROI-SVD) coil-compression approaches10-11. We considered a scenario
where we are interested in the signal from the brain, but not in the signal
from other parts of the head, neck, or torso.
Application to reduced-FOV: One potential application of ROVir could be to reduce the size of the
image support, enabling small FOV imaging.
As examples, we applied ROVir to reduced-FOV head imaging. Using 32-channel brain data, we
retrospectively simulated a spiral k-space trajectory designed for a small FOV,
as well as a Cartesian k-space trajectory designed for a small FOV. Reconstructions were performed with standard
gridding/Fourier transform reconstruction, with and without ROVir.Results
Figure 2 shows a comparison of ROVir against SVD and ROI-SVD coil
compression methods. The results
demonstrate that ROVir can substantially suppress the interference, compacting energy
from the ROI into the first virtual coils, while also compacting energy from
the unwanted spatial region into the last virtual coils. Keeping only the first few virtual coils
successfully suppresses unwanted spatial regions.
Results for non-cartesian reduced-FOV imaging are shown in Fig. 3 for
different ROIs, demonstrating that ROVir enables aliasing-free depictions of each
ROI by suppressing the interference from unwanted regions. Fig. 4 demonstrates
that beam-steering can be used to allow ROVir to focus on arbitrary ROIs within
the FOV. Figure 5 shows similar
reduced-FOV results in the context of Cartesian acquisition. Conclusion
This work proposed
ROVir, a simple beamforming framework that can be used to localize an ROI and suppress
signal from unwanted spatial regions.
ROVir is compatible with and complementary to existing signal localization/suppression
methods. While we only showed examples
involving simple image reconstruction methods, ROVir can be applied directly to
raw and potentially undersampled k-space data, and can potentially be used in
combination with arbitrary image reconstruction methods. We believe that this approach is potentially
useful for a wide range of different MRI applications that would benefit from
improved signal localization/suppression.Acknowledgements
This
work was supported by research grants NIH R01-MH116173, NIH R01-NS074980, and
NIH R01-NS089212 and R33-CA225400.
References
1. Van Veen BD, Buckley KM. Beamforming: A versatile approach to spatial filtering. IEEE ASSP Mag 1988;5:4-24.
2. Monzingo RA, Haupt RL, Miller TW. Introduction to Adaptive Arrays. Raleigh: SciTech Publishing. 2011.
3. Roemer PB, Edelstein WA, Hayes CE, Souza SP, Mueller OM. The NMR phased array. Magn Reson Med 1990;16:192-225.
4. Walsh DO, Gmitro AF, Marcellin MW. Adaptive reconstruction of phased array MR imagery. Magn Reson Med 2000;43:682-690.
5. Kellman P, McVeigh ER. Ghost artifact cancellation using phased array processing. Magn Reson Med 2001;46:335-343.
6. H Eggers PB P Boernert. Real-time partial parallel spiral imaging with localized sensitivities. In: Proc. Int Soc Magn Reson Med 2001; p. 1772.
7. Lin FH, Witzel T, Ze_ro TA, Belliveau JW. Linear constraint minimum variance beamformer functional magnetic resonance inverse imaging. NeuroImage 2008;43:297-311.
8. Ebbini E, DelaBarre L, Vaughan JT, Gopinath A. Digital beamforming in MRI. In:Proc. IEEE BenMAS. 2014.
9. Mandava S, Keerthivasan MB, Martin DR, Altbach MI, Bilgin A. Radial streak artifact reduction using phased array beamforming. Magn Reson Med 2019;81:3915-3923.
10. Buehrer M, Pruessmann KP, Boesiger P, Kozerke S. Array compression for MRI with large coil arrays. Magn Reson Med 2007;57:1131-1139.
11. Huang F, Vijayakumar S, Li Y, Hertel S, Duensing GR. A software channel compression technique for faster reconstruction with many channels. Magn Reson Imag 2008;26:133-141.
12. Cauley S, Polak D, Liu W, Bilgic B, Gagoski B, Grant PE, Conklin J, Kirsch J, Huang SY, Setsompop K, Wald LL. Geometric coil mixing (GCM) to dampen confounding signals in MRI reconstruction. In: Proc. Int. Soc. Magn. Reson. Med.. 2019; p. 449.
13. Jackson JI, Meyer CH, Nishimura DG, Macovski A. Selection of a convolution function for Fourier inversion using gridding. IEEE Trans Med Imag 1991;10:473-478.
14. Griswold MA, Jakob PM, Nittka M, Goldfarb JW, Haase A. Partially parallel imaging with localized sensitivities (PILS). Magn Reson Med 2000;44:602-609.