Sebastian Theilenberg1, Chathura Kumaragamage2, Scott McIntyre2, Terry W. Nixon2, Christoph Juchem1,3, and Robin A. de Graaf2
1Biomedical Engineering, Columbia University, New York, NY, United States, 2Department of Radiology and Biomedical Imaging, Magnetic Resonance Research Center, Yale University School of Medicine, New Haven, CT, United States, 3Radiology, Columbia University Medical Center, New York, NY, United States
Synopsis
In line with recent developments in scanner design for more
cost effective and more accessible scanners, we propose a low-angle
combined-echo sequence similar to a conventional spin echo sequence capable of
producing high quality images in the presence of strong B0 inhomogeneity. We
present simulation results investigating the signal dependence on the sequence’s
timings and flip angles as well as image contrast for typical relaxation times
of normal brain white and gray matter. The simulations were validated in vivo at
4 T. Lastly, we show the feasibility to utilize multi-coil generated image encoding
fields for this sequence.
Introduction
In
recent years, a variety of new approaches to scanner design have been presented
aiming to lower the costs of MRI and to increase accessibility. One of these
approaches is relaxing the homogeneity requirements for the main B0
magnetic field from typically less than a few hundred hertz to 20 kHz (equivalent
to a few ppm to 300 ppm at 1.5 T) or
more across
the volume of interest in a recently proposed 1.5 T head-only scanner to reduce
scanner cost1. This
development necessitates the design of MRI pulse sequences robust against these
conditions. Here we present a low-angle combined-echo (LACE) sequence capable of
producing high quality images in the presence of strong B0
inhomogeneity.Methods
The proposed LACE sequence is,
similar to a conventional spin echo sequence, based on two broadband radio-frequency pulses separated by the time $$$T_E/2$$$, signal readout at
the time of the echo $$$T_E$$$ and a delay to complete the repetition time $$$T_R$$$.
With the use of short $$$T_R$$$ delays, the echo formed at steady-state contains
both spin echo and stimulated echoes (Figure 1). Simulations based on the product
operator formalism2
were performed in MATLAB (MathWorks, Natick, MA, USA) to investigate the
dependence of the signal strength on the sequence parameters. The transverse
magnetization was simulated at flip angles $$$\alpha_1/\alpha_2\leq40^\circ$$$, $$$T_E\leq100\,\mathrm{ms}$$$ and $$$T_R\leq150\,\mathrm{ms}$$$. The simulations have
been performed with two combinations of $$$T_1$$$ and $$$T_2$$$ values
($$$T_1/T_2=900/50\,\mathrm{ms}$$$ and $$$1400/60\,\mathrm{ms}$$$)
representing experimentally determined values for human brain white and gray
matter (WM/GM) at 4 T.
To validate the simulation results, in vivo
brain images were acquired from 6 healthy subjects on a 4 T Bruker scanner at
the Yale MR Research Center (MRRC) with the following sequence parameters: $$$T_E=12\,\mathrm{ms}$$$, $$$T_R=100\,\mathrm{ms}$$$, AFPST4 pulse
excitation (1 ms) with 20 kHz bandwidth, readout bandwidth BWRO = 100
kHz. A paper clip was
fixed to the bottom of the head rest below the subjects’ head to validate the B0
insensitivity of the MRI sequence at hand. The proposed 1.5 T head-only scanner will be
equipped with a multi-coil (MC) array3 instead of linear gradient coils for the
generation of image encoding fields1,4, therefore, we also tested LACE with MC generated encoding fields similar to
previous work5.Results
The simulations show the transverse magnetization rapidly
increases in the first few repetitions of the sequence and after reaching a maximum
slowly decreases to a steady-state (Figure 2A).
The steady-state magnetization is larger if α2 is larger than α1, and in general combinations of larger flip angles lead to a
stronger steady-state magnetization, with a maximum in the investigated
range around $$$\alpha_1/\alpha_2=30^\circ/40^\circ$$$ (Figure 2B).
Steady-state magnetization increases with longer $$$T_R$$$ and shorter $$$T_E$$$
(Figure
2C).
Absolute contrast between GM and WM, calculated as the absolute difference between the steady-state transverse magnetization of the two sets of $$$T_1/T_2$$$ values, was found to
be stronger with shorter $$$T_E$$$ and longer $$$T_R$$$ in general.
There is an area of low contrast for intermediate values of $$$T_R$$$ whose
location is dependent on the relative proton density $$$PD_{GM}/PD_{WM}$$$ between the two tissues (Figure 3A-C). A region of high contrast is evident when mapping it against the flip angles $$$\alpha_1/\alpha_2$$$. While the shape and strength of that area differ with relative PD, small flip angles around $$$\alpha_1=10^\circ$$$ and $$$\alpha_2=15^\circ$$$ should always yield good contrast (Figure 3D-F).
Brain images acquired in vivo match the simulated contrast behavior when assuming a relative PD of
about 85% with images at $$$\alpha_1/\alpha_2=10^\circ/20^\circ$$$ showing good contrast with $$$T_E/T_R=12/100\,\mathrm{ms}$$$, but images acquired
at $$$\alpha_1/\alpha_2=10^\circ/40^\circ$$$ or with $$$T_R=50\,\mathrm{ms}$$$ showing only little contrast (Figure 4). The latter case also shows reduced signal intensity, in line with Figure 2C. While the B0
inhomogeneity induced by the paper clip effects large portions of a 3D
gradient echo scan acquired for reference, LACE imaging is able to recover the signal in this area. LACE
images acquired with MC generated encoding fields show largely the same image
quality than the ones acquired with regular gradients (Figure 5). Mismatches between regular LACE and MC-LACE images are likely due to
non-linearities in the MC encoding fields at the edge of the FoV.Discussion
LACE MRI enables imaging in the presence of strong B0
inhomogeneity by utilizing high bandwidth RF pulses and selecting combined
spin- and stimulated echoes. The overall signal intensity is ~20% of a
low-angle gradient-echo sequence (e.g. FLASH, calculated with the Ernst equation6 at flip angle 15°) and is comparable to that of a
recently described MP-SSFP method7. High GM/WM image contrast can be achieved
with modest nutation angles, providing a low-power alternative to high-powered
spin-echo-based methods (e.g. 3D FSE). Even though employing larger flip angles generates more signal, sufficient signal
can be acquired at low flip angles with good GM/WM contrast. If the background B0
field behavior is known, as it is the case for a given main magnetic field, the
resulting image distortions can be corrected. Utilizing lower flip angles
enables larger bandwidth RF pulses that are necessary to excite all spins in
the presence of large off-resonances without exceeding SAR constraints. With
only two excitation pulses the described sequence is rather simple and should
be straightforward to implement in any MRI system.Acknowledgements
This research was supported by the National Institute of
Biomedical Imaging & Bioengineering of the National Institutes of Health under
award number U01EB025153.References
1. Garwood M, Mullen M, Kobayashi N, et al. A compact vertical 1.5T human head scanner with shoulders outside the bore
and window for studying motor coordination. Proc Intl Soc Magn Reson Med. 2020;28:1278.
2. Sørensen OW, Eich GW, Levitt MH,
Bodenhausen G, Ernst RR. Product operator formalism for the description of NMR
pulse experiments. Prog Nucl Magn Reson Spectrosc. 1984;16:163-192.
3. Juchem C, Nixon TW, McIntyre S, Rothman DL, de Graaf RA. Magnetic field modeling with a set of individual localized
coils. J Magn Reson. 2010;204(2):281-289.
4. Theilenberg S, Shang Y, Kobayashi N,
Parkinson B, Juchem C. Multi-Coil Array for Combined Imaging and B0 Shimming
in a Portable Head-Only Scanner. Proc Intl
Soc Magn Reson Med. 2019;27:1480.
5. Juchem C, Theilenberg S, Kumaragamage
C, et al. Dynamic multicoil technique (DYNAMITE) MRI on human brain. Magn
Reson Med. 2020;6(84):2953-2963.
6. Ernst RR, Bodenhausen G, Wokaun A. Principles of Nuclear Magnetic Resonance in One and Two Dimensions. Oxford: Clarendon Press; 1987:610.
7. Kobayashi N, Parkinson B, Idiyatullin
D, et al. Development and validation of 3D MP‐SSFP to enable MRI in
inhomogeneous magnetic fields. Magn Reson Med. 2020;85(2):831-844.