Mark J Burton1, Robert Thomen1, Talissa A. Altes1, Carlos Leiva Salinas1, Nicholas Tustison2, and Joseph P Cousins1
1Radiology, University of Missouri Hospital, Columbia, MO, United States, 2Radiology, University of Virginia, Charlottesville, VA, United States
Synopsis
This work was to assess the trade-off between scan time and
image quality afforded by acceleration techniques GRAPPA and simultaneous
multi-slice (SMS). We imaged an ACR phantom using standard clinical T1,
T2, and FLAIR sequences. Then, without altering any other scan
parameters, we increased GRAPPA and SMS acceleration factors (2, 4, and 6). All GRAPPA and SMS accelerated scans passed
the ACR resolution criteria. Both GRAPPA and SMS passed low-contrast criteria
at acceleration factor 2, but factors of 4 and 6 did not. GRAPPA and SMS gave nearly
identical similarity indices and SNR at all acceleration factors.
Purpose
In many institutions the standard
brain MRI examination is a relatively lengthy imaging protocol. There is
potential to substantially decrease acquisition time using recent technical
developments including parallel imaging and simultaneous multi-slice [1-3]. However,
these acceleration techniques can degrade image quality. The purpose of this
work was to assess the trade-off between scan time and image quality using the
industry standard ACR imaging quality metrics and the acceleration techniques GRAPPA
and simultaneous multi-slice (SMS). Methods
MRI scans of the ACR detail
phantom were acquired on a Siemens VIDA 3.0T MRI scanner (Siemens Healthcare,
Erlangen, Germany) using a 20-channel head coil. T1-weighted, T2-weighted, and FLAIR
images were acquired using the ACR standard for evaluating SNR and CNR high and
low contrast object detectability measures.
For each weighted sequence, all sequence parameters remained the same
except the acceleration technique and acceleration factor (Table 1). Baseline scans were acquired without any
acceleration techniques. GRAPPA and SMS
were then added separately, and increased by 2, 4, and 6-times acceleration. The SNR, CNR, and minimum detectable element
size were evaluated for each acquisition. The MR images were evaluated with spatial
normalization, using quantitative comparison, and image similarity metrics [4],
including localized neighborhood cross correlation [5] and the structural
similarity index [6]. A visual evaluation of the high contrast resolution, low
contrast and artifacts as set by the ACR, as minimum passing requirements, was
also performed [7]. Results
Figure 1 demonstrates the changes in SNR and similarity index for the T1,
T2 and FLAIR acquisitions using GRAPPA and SMS acceleration factors of 2, 4 and
6. GRAPPA and SMS had nearly identical results in
SNR and similarity indices for each acceleration factor. Both GRAPPA and SMS showed a slight increase in
SNR at an acceleration factor of 2 but reduction in SNR and similarity indices
at acceleration factors of 4 and 6. All GRAPPA and SMS accelerated scans passed
the ACR resolution and high-contrast criteria, but only the acceleration factor
2 scans passed the ACR low-contrast criteria (Figure 2 and 3). Conclusions
By assessing resolution, high- and low-contrasts, SNR, and similarity indices
between images of increasing acceleration, we find that GRAPPA and SMS acceleration
factor of 2 gave the best imaging results with increased SNR while reducing the
scan time by 40%, but higher acceleration factors of 4 and 6 dramatically
reduced SNR and produced images of non-diagnostic low-contrast quality
according to ACR quality assurance guidelines.Acknowledgements
No acknowledgement found.References
1. Wiggins GC, Polimeni
JR, Potthast A, et al. 96-channel receive-only head coil for 3 Tesla: design
optimization and evaluation. Magn Reson Med 2009;62:754 –62
2. Wiggins GC,
Triantafyllou C, Potthast A, et al. 32-channel 3 Tesla receive-only
phased-array head coilwith soccer-ball element geometry. Magn Reson Med
2006;56:216–23
3. Keil B, Blau JN,
Biber S, et al. A 64-channel 3T array coil for accelerated brain MRI. Magn
Reson Med 2013;70:248 –58
4. Avants BB, Tustison
NJ, Song G, Cook PA, Klein A, Gee JC. A reproducible evaluation of ANTs
similarity metric performance in brain image registration. NeuroImage.
2011;54(3):2033-44.
5. The Insight ToolKit
image registration framework. Avants BB, Tustison NJ, Stauffer M, et al. Front
Neuroinform. 2014;8:44.
6. Image quality
assessment: from error visibility to structural similarity. Wang Z, Bovik AC,
Sheikh HR, Simoncelli EP. IEEE Trans Image Process. 2004;13:600-12.
7. American College of
Radiology MR Accreditation Program Testing Instructions. (accessed Nov 6, 2019) https://www.acraccreditation.org/-/media/ACRAccreditation/Documents/MRI/MRAccreditationTestingInstructions.pdf?la=en