Larowin Toni1, Olga Smelianskaia 1, Elizabeth Morris1, and Yolanda Bryce1
1Radiology, Memorial Sloan Kettering Cancer Center, New York, NY, United States
Synopsis
An MRI-guided breast biopsy can be lengthy and
uncomfortable for the patient, which potentially can be mitigated by
computer-aided software that helps localize the lesion, compared with manual
localization. We performed a retrospective trial to see if a computer-aided
software system reduces procedure length, increases accuracy, and increases
confidence of the clinician performing the procedure. Our study demonstrates
that computer-aided software can significantly decrease length of time of the
procedure but does not impact accuracy of the biopsy or confidence of the
clinician performing the procedure.
Background
MRI-guided vacuum-assisted breast biopsies are performed to
sample suspicious lesions not seen on conventional imaging (1)(2). For a biopsy, a woman lies on the MRI table in breast coils, with breast in
compression, while multiple scans, administration of contrast, and localization
of the lesion via a grid and skin marker are performed (3). The procedure can be
lengthy and uncomfortable for the patient, which potentially can be mitigated
by a computer-aided software to localize the lesion, decreasing
patient's time of discomfort, compared to manual localization,
which may be prone to human error and increase intraprocedural time (4). We
performed a retrospective trial to see if computer-aided software system
reduces procedure length, increases accuracy, and increases confidence of the
clinician performing the procedure.
Materials and Methods
MRI-guided biopsy with localization:
A Sentinelle Table is connected to the 1.5 Tesla 450W GE
Discovery MRI Unit which uses the DV 26 software (Figure 1). The breast
is then adequately compressed to prevent it from moving but at the same time allow flow of contrast into the
breast. The center of the grid is then marked within the interventional
grid. With manual localization, a vitamin E marker is placed on the patient's
skin within the grid (Figure 2). Alternatively with a computer-aided software,
the manufacturer's special fiducial marker is placed (Figure 3). This
fiducial marker is substantially larger than the common vitamin E
marker.
With computer-aided
software, entire coverage of the manufacturer's fiducial is important as
localization will otherwise fail. Localizing the lesion manually is done
with the knowledge of the breast lesion position relative to the vitamin E marker
within the grid. Depth is calculated by subtracting the image number
with the lesion from the image number with skin and multiplied by the slice
thickness. A computer-aided software takes all these parameters into account
and provides the location of the lesion relative to the fiducial and depth of
the lesion. Once the lesion is localized, a needle holder is placed, which
itself is a small grid to help optimally target the lesion in smaller
increments. The needle is then advanced into the breast and its positioning and
sampling confirmed with before and after T1 post contrast images.
Retrospective study was performed with evaluation of electronic medical records and imaging of 100 sequential women biopsied under MRI-guidance with manual localization and 100 sequential women biopsied under MRI-guidance with computer-aided software, from December 2018 extending retrospectively to January 2016. Information regarding procedure time, number of scans, histology, post procedure hematoma, follow-up recommendations, outcome of surgery if performed, interval cancers, and follow up period was collected (Figure 4).Results
Retrospective study demonstrates a significant decrease in
both average procedure time and number of scans with computer-aided
software compared with manual localization (25.63 minutes vs. 32.96
minutes, p = <0.001 and 6.92 vs. 9.64, p = < 0.001, respectively) (Figure 4). There was no difference in size of breast lesions (computer aided software
1.57 cm vs. manual 1.508 cm, p = 0.7374) (Figure 5). Post procedural hematoma
rate was similar between both groups (computer aided software 30% vs. manual
21%, p = 0.157). Rates of short interval follow up after benign histology a
surrogate for performing clinicians confidence, was not significantly different
(computer aided software 13% vs. manual 16%, p = 0.6885). There was also no
significant difference in upgrade rates at the time of surgery, discordant
results, and interval cancers. Follow up period was significantly longer in the
manual cohort compared with the computer-aided software cohort (12.96 months
vs. 9.77 months, respectively).Teaching points
-
The protocol uses LAVA VIBRANT-Flex for both sagittal and axial views
- Sagittal protocol: TR of 17.2ms, TE of 4.2ms, slice thickness of 3mm; matrix size 256 x 192, field of view (FOV) 34-36, bandwidth of 200KHz, flip angle 15.
- Axial protocol: TR of 5.1ms, TE of 2.4ms, slice thickness of 2mm, matrix size of 256 x 160, FOV of 24-26, bandwidth of 41.67KHz, flip angle of 10.
- Localizing the lesion
manually is done with the knowledge of the position of the vitamin E marker and
the breast lesion within the grid and the depth of the lesion from the skin. With
computer-aided software, the manufacturer’s fiducial is used. Entire coverage
of the fiducial is important as localization will fail.
- Computer-aided software
may decrease procedure time compared to manual localization.
Summary
MRI guided biopsy is a vital technique to help sample breast
lesions not seen with conventional imaging. Procedural imaging parameters are
well established. Tools such as computer-aided software can potentially help
mitigate a lengthy procedure and patient discomfort. However, it does not impact
accuracy of the biopsy or confidence of the clinician performing the
procedural. Limitations of the study include small retrospective study and
inherent bias. Acknowledgements
DISCLAIMER
This disclaimer informs readers that the views, thoughts, and opinions expressed in this
text solely to the author, and not necessarily to the author’s employer,
organization, committee or other group or individual. We do not have any affiliation with the Hologic
company other than what is described within this content.
References
1.Boo-Kyung Han, Mitchell D. Schnall,
Susan G. Orel, and Mark Rosen. Outcome of MRI-Guided Breast Biopsy. American
Journal of Roentgenology 2008 191:6, 1798-1804
2.Chevrier
M-C, David J, Khoury ME, Lalonde L, Labelle M, Trop I. Breast biopsies under
magnetic resonance imaging guidance: challenges of an essential but imperfect
technique. Curr Probl Diagn Radiol
2016;45(3):193-204.
4.Gombos EC, Jagadeesan J, Richman DM, Kacher DF. Magnetic Resonance Imaging-Guided Breast Interventions: Role in Biopsy Targeting and Lumpectomies. Magn Reson Imaging Clin N Am. 2015;23(4):547–561. doi:10.1016/j.mric.2015.05.004
3.Raza S, Chikarmane SA, Gombos EC,
Georgian-Smith D, Frost EP. Optimizing Success and Avoiding Mishaps in the Most
Difficult Image-guided Breast Biopsies. Seminars
in Ultrasound, CT and MRI 2017;0(0).