Dave Hitt1, Jonathan Chia1, Robert Lay1, Tom Lowe1, Michael Pawlak1, John Penatzer1, James Snicer1, Marcie Stopchinski1, Gregory Thomas1, Paul Worthington1, Kristen Williams1, and Brian Johnson2
1Philips Healthcare, Gainesville, FL, United States, 23545 SW 47th Ave, Philips Healthcare, Gainesville, FL, United States
Synopsis
Patients with MR Conditional orthopedic implants can be challenging
to scan due to susceptibility artifact from metal implants. Use of metal artifact reduction (MARS)
techniques can help reduce signal loss from metal implants, however, are not
usually sufficient to allow for detailed peri-prosthetic tissue
visualization. Metal-induced field
inhomogeneities cannot be accurately modeled, which is why O-MAR (MARS+VAT) can
improve in-plane distortions by utilizing a more robust spatial encoding
approach. Here we present the conceptual
background behind the O-MAR technique and work done to provide guidance for
optimized metal reduction imaging protocols.
Background
As
life expectancy continues to increase, so does the number of patients receiving
orthopedic implants1.
Patients receiving orthopedic implants are at risk for developing:
osteolysis, metallosis, heterotopic ossifiction, osteonecrosis, and component
loosening2. MRI provides
excellent soft-tissue contrast and is the modality of choice for the assessment
for these clinical indications, but suffer from susceptibility artifacts3,4. Therefore, when imaging orthopedic implants
with MRI steps need to be taken to reduce the metal artifact. Common approaches for reducing metal artifact
include use of spin echo based sequences over gradient echo based sequences,
reducing the echo spacing in the echo train of turbo spin echo sequences, and
increasing the receiver bandwidth. The
use of all three of these imaging strategies collectively is called MARS (Metal
Artifact Reduction) and originally described by Olsen et. al (2000) but now commonly
available from every vendor (Philips, Siemens, and GE). Improvements on MARS have taken place by
adding view angle tilt (VAT)5,6 with MARS to further reduce in-plane
susceptibility artifact and slice encoding for metal artifact correction (SEMAC)7
which also helps to reduce through-pane susceptibility artifact. On the Philips platform the MARS+VAT
technique is called O-MAR (Orthopedic Metal Artifact Reduction). Here we present the conceptual background
behind the O-MAR technique and work done to provide guidance for optimized
metal reduction imaging protocols. O-MAR (MARS+VAT) Teaching Points
Implant:
Severity
of artifact caused by the implant depends on many factors including: implant
material, size, shape, orientation, and the imaging sequence being used
(Figure1).
Susceptibility
Artifact: The
magnetic susceptibility of metal is much higher than that of tissue, such that
around metal, very large variations in Larmor frequency occur. This not only
causes signal reduction, but also signal loss. So called signal pile-ups can
also occur due to non-linear frequency-position mapping. Slice profile
selection also relies on local precession frequency variations and is equally
affected by susceptibility variations, resulting in distorted slice profiles
and consequently signal misregistration in the slice direction. Consequently,
around metal implants, anatomy can be severely obscured.
Spin
echo: Spin echo based sequences are less prone to
metal artifacts compared to gradient echo based sequences because the 180˚ RF
refocusing pulse causes signal loss at the rate of T2 compared to T2*.
Bandwidth: Geometric distortions are caused by altered
Larmor frequencies around metal implants, so increasing the receiver bandwidth
helps reduce the range of resonant frequencies over which the distortion is
spread over the pixels and containing the in-plane geometric distortion to a
smaller area. However, be aware that
increasing receiver bandwidth reduces the signal-to-noise ratio (SNR), so care
is need to find the optimal balance between SNR and artifact reduction.
Echo
spacing:
Reducing the echo spacing allows for a shorter minimum echo time (TE) and a
shorter shot length. This is also
usually achieved by increasing the receiver bandwidth. By decreasing the minimum TE, less dephasing
due to the metal susceptibility occurs.
Moreover, shortening the shot length by having a smaller echo spacing
decreases the dephasing associated with the application of multiple phase
encoding gradients in a TSE acquisition. However, image contrast needs to be
considered when choosing the minimum TE and echo spacing as all effect the
final image contrast.
Resolution: While not
necessarily included in the MARS technique, increasing resolution also helps
reduce metal artifacts by reducing geometric distortions similarly to
increasing the receiver bandwidth.
VAT: VAT is an efficient technique for in-plane
artifact correction5. VAT
uses an extra gradient in the slice select direction during the signal
read-out, and the slice is effectively viewed from an angle (Figure 2). This VAT tilt works because the sum of
artifacts from the frequency shifts in the slice select and the read-out direction
caused by the metal results in a frequency shift with an oblique orientation. By tilting or using a read-out with an
oblique angle the received signals can be projected into the correct
pixel. VAT can cause blurring but this
is minimized by the high receiver bandwidth already implemented by the MARS7.
Parallel
Imaging: Use
of parallel imaging is allowed with O-MAR in conjunction with parallel array
coils. However, parallel imaging relies
on the use of a reference scan to determine coil sensitivities that can lead to
artifacts during image reconstruction.
Additional O-MAR Parameter Optimization Considerations
(Figure 3)Summary
Patients with MR
Conditional orthopedic implants can be challenging to scan due to
susceptibility artifact from metal implants.
Use of MARS techniques (TSE sequences with short echo spacing and high
receiver bandwidth) can help reduce signal loss from metal implants, however,
are not usually sufficient to allow for peri-prosthetic tissue
visualization. Metal-induced field
inhomogeneities cannot be accurately modeled, therefore O-MAR (MARS+VAT) which
relies on a more robust spatial encoding approach should be implemented in the
presence of metal implants to improve in-plane distortions. With O-MAR all
relevant image contrasts can be obtained (T1, T2, PD, and STIR) at 1.5T and 3T
filed strengths (Figure 4).Acknowledgements
No acknowledgement found.References
1. Kurtz, S., et al., Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030. J Bone Joint Surg Am, 2007. (4): p. 780-785.
2. Ulrike Dapunt et al, HYPERLINK “/pmc/articles/ PMC3994416/”On the inflammatory response in metalon- metal implants , J Transl Med. 2014; 12: 74.
3. B.A. Hargreaves, P. W. Worters, K.B. Pauly, J.M.Koch and G.E. Gold, Metal-Induced artifacts in MRI, AJR: 197, sept 2011.
4. L.M. White, J.K. Kim, M. Mehta, N. Merchant, M.E. Schweitzer, W.B. Morrison, C.R. Hutchinson, A.E. Gross. Complications of total hip arthroplasty: MR imaging— initial experience. Radiology. 2000;215:254-62.
5. J.E.1 Vandevenne, F.M. Vanhoenacker, P.M. Parizel, K. Butts, J.M. Pauly, R.K. Lang. Reduction of metal artefacts in musculoskeletal MR imaging, JBR-BTR. 2007 Sep-Oct;90(5):345-9.
6. R.V. Olsen, P.L. Munk, M.J. Lee, et al.: Metal artifact reduction sequence: early clinical applications. Radiographics, 2000, 20: 699-712.
7. K. Butts, J.M. Pauly, G.E. Gold: Reduction of blurring in view angle tilting MRI. Magn Reson Med, 2005, 53: 418- 424..