Huijun Vicky Liao1, Eduardo Jorge Uribe Coello1, and Alexander P. Lin1
1Radiology, Brigham and Women's Hospital, Boston, MA, United States
Synopsis
For
magnetic resonance spectroscopy, automated B0 shim mode may save scan time in a
fast-paced clinical environment but is strongly dependent on performance. Two shim modes were compared in multiple brain
regions at Siemens 3T and 7T. B0 maps
were used to evaluate the shimming performance of each mode in the left
hippocampus region. The results show that advanced shim mode is preferred on 7T
(syngo MR VE12) and brain shim mode is preferred on 3T (syngo MR VE11) for brain
MRS. However, more sophisticated shimming methods are still suggested for scanning
more inhomogeneous regions if time permits.
Introduction
B0
shimming is particularly important for brain MR spectroscopy. A good shim with narrow
spectral lines will provide more reliable separation and quantification of
metabolites which in turn improves diagnostic accuracy1,2. Siemens MR
software such as Syngo MR VE11 and VE12 provided different automatic B0 shim
mode options that can save scan time by automatically optimizing the B0 field within
the MRS voxel. This avoids
time-consuming manual shimming which is not possible in a routine clinical
environment. However, to our knowledge, the performance of these shim modes for
brain MRS scan has not been tested. The goal of this study is to evaluate the
shim modes’ performance and determine the ideal shim mode to use for brain MRS
scan in clinically-relevant scenarios using both linewidth and B0 mapping
comparisons.Methods
MR exams
were performed in 14 subjects after informed consent on a Siemens Magnetom Terra 7 Tesla MRI scanner
(Syngo MR VE12) using a Nova medical head coil (1TX / 32RX) with dielectric
pads6 and on a Siemens Magnetom Skyra 3T MRI scanner (Syngo MR VE11)
using 32 and 20 channel head coils.
Full
Width Half Maximum (FWHM),
the spectral linewidth reported by Siemens Syngo system, were compared in
different MRS voxel locations for two common B0 shim modes: “Brain” and “Advanced” (Figure 1) at 7T and 3T. Locations included:
anterior cingulate gyrus (ACG), posterior cingulate gyrus (PCG), left
hippocampus (L-hippo), dorsolateral prefrontal cortex (DLPFC), ventromedial
prefrontal cortex (VMPFC) as shown in Figure 2. These regions were selected as
they are often used across many different studies such as psychiatric diseases such as schizophrenia1,4
and bipolar 1,5 disorder. A
focus on the left hippocampus was made because it was a region close to sinus
with more field distortion and difficult to shim. At 3T, the comparison was also done in
patient with a tumor adjacent to the right posterior ventricle. Paired t-tests
was performed to compare the FWHM from the two shim modes.
3D B0
Mapping: of the left
hippocampus region (3.5x1.5x1.5cm voxel) using the following parameters:
TR/TE1/TE2=4.4/1.02/3.06ms, flip angle 10°, 44 slices and image acquisition
time of 8.3s. As the adjust volume (shim box) was copied from the MRS voxel, B0
was set to the same shim channel values on X, Y, Z, Z², ZX, ZY, X²-Z² and XY as
the MRS voxel in advanced shim mode.
This was then repeated for brain shim mode. B0 maps were processed
by MATLAB.
Additional
B0 Shim Modes: A braino
phantom with a 3x3x3 cm3 voxel was scanned with different shim modes
to evaluate how shim channels being optimized on 7T. Shim modes included: Tune
up, Standard, Advanced, Cardiac, Brain, Breast, and Prostate (Figure 1). Results
FWHM: Figure 3 shows a wide range of
FWHM from different MRS voxel locations with higher FWHM in left-hippo (yellow)
and lower FWHM in PCG (purple) and ACG (blue) in both shim modes as expected. However,
brain shim mode (mean FWHM 14.89±6.53Hz) performs significantly better than
advanced shim mode at 3T (mean FWHM 21.85±8.43Hz; p=0.0016), but at 7T, brain
shim mode (mean FWHM 44.42±12.76Hz) performs significantly worse than advanced
shim mode (mean 30.16±11.40Hz; p=0.00038).
3D B0
Mapping: In a challenging region like the left hippocampus, brain and advanced shim mode give
similar FWHM values (Brain shim:44.8Hz and Advanced shim:44.1Hz) on 7T. However,
even with similar FWHM values, B0 maps show that advanced shim mode can
generate a more homogeneous region (mean ΔB0 =111Hz) as the selected volume
included the left hippocampus than brain shim mode (mean ΔB0 =141Hz) which is
shown in Figure 4b and c (especially on the coronal and sagittal planes). However,
when a smaller volume (the size of the left hippocampus voxel) is selected,
both modes give the same mean ΔB0=80Hz.
Additional
B0 Shim Modes: Figure
5a shows FWHM values rank from low to high (left to right) from different B0 shim
modes. Interestingly, both breast and advanced shim modes give the best FWHM
but brain gives the worst. FWHM from breast, advanced, prostate and standard shim
modes are similar (5.8-6.6Hz) since they give similar values in first-order (X,
Y, Z) channel (Figure 5b), but tune up, cardiac and brain shim modes give double
even triple of FWHM since channel Y of tune up, X of cardiac and Z of brain give
much lower values than other shim modes (Figure 5b). Also, all second order
channels (Z², ZX, ZY, X²-Z², XY) behaved completely differently than others. Discussion and Conclusion
The
results show that advanced shim mode is preferred on 7T and brain shim mode is
preferred on 3T for brain MRS exams. By choosing the correct shim mode on each
scanner can help acquire optimal data in a shorter time and make acquisition
method more reproducible. However, B0 homogeneity is still highly region-dependent. If time permits, more sophisticated shimming methods such as
manual/interactive shimming or FASTMAP 2,3 are still suggested for
difficult regions like the left hippocampus. Other B0 shim modes such as breast,
prostate and standard on 7T (Syngo MR VE12) may also generate good shim quality
for brain MRS scan, however, further research is needed to confirm each shim
mode’s performance.Acknowledgements
No acknowledgement found.References
1.Lin
A, Tran T, Bluml S, et al. Guidelines for Acquiring and Reporting Clinical
Neurospectroscopy. Semin Neurol. 2012;32:432–453.
2.Juchem
C and Graaf R A, B0 Magnetic Field Homogeneity and Shimming for In Vivo MR
Spectroscopy. Anal Biochem. 2017 July 15; 529: 17–29.
3.Wilson
M, Andronesi O, Barker P, et al. Methodological consensus on clinical proton
MRS of the brain: Review and recommendations. Magn Reson Med. 2019;82:527–550.
4.Singh
S, Khushu S, Kumar P, et al. Evidence for regional hippocampal damage in
patients with schizophrenia. Neuroradiology. 2018 Feb;60(2):199-205.
5.Gigante
AD, Lafer B, Yatham LN. (1)H-MRS of hippocampus in patients after first manic
episode. World J Biol Psychiatry. 2014 Feb;15(2):145-54.
6.Liao
H et al. The Effects of Dielectric Pads on Brain MR Images and Spectroscopy at
7T. Proc Soc Magn Rad Tech. SMRT 28th
Annual Meeting, Montreal, 2019.