Natalia Gudino1
1National Institutes of Health, United States
Synopsis
The aim of this presentation is to review the main hardware components in the MRI system and provide understanding of specifications based on the MRI physics.
Introduction
In MRI a very large magnet generates a strong static magnetic field (B0) (>10,000 x Bearth)
to align the nuclear magnetic moments in the body, which are otherwise in
random orientation. This creates a bulk nuclear magnetisation (M0) proportional to the static field and nuclear
spin density. The precession of M0 around the
static field is determined by the gyromagnetic ratio (γ) which is a measure
of the magnetic moment-to-angular momentum
ratio of the nucleus. To detect
a particular nucleus, a high-power RF pulse excitation with angular frequency
equal γB0 (Larmor frequency) is applied by a Tx coil
with axis (in x-y plane) perpendicular to B0 (z-axis).
This generates a small magnetic field B1 that rotates
at the same speed as the spins
around B0 perturbing M0 from its equilibrium state (precessing around B0) to precesses, only temporarily, around B1 field. After this pulsed excitation, M0 returns back to its equilibrium state aligned
with B0. This free precession back to
equilibrium induces a time variant voltage (ε ) (MR signal) on a tuned receive (Rx)
coil with axis also perpendicular to B0 (could
be same Tx coil). The amplitude and transient response (oscillating at the
Larmor Frequency) of the magnetization depends on tissue relaxation constants
(T1, T2, T2*) and nuclei spin density (ρ). These
parameters can differentiate healthy from diseased tissue as well as to provide
insights about biological processes like blood perfusion. Since the
magnetization precession frequency is proportional to the static field, the
signal generated can be spatially encoded by well-defined varying magnetic field gradients (𝐺⃗).
Field gradients in three directions (x, y, z) can be generated by three sets of coil arrangements that are concentrically located inside
the magnet coil. A linear relationship between precession frequency and a
spatial coordinate makes possible the generation of an image by a simple Fourier transform. All events involved in
the generation of this image (RF excitation, gradient encoding and signal
acquisition) are controlled by a pulse sequence executed in the scanner console
which communicates with all system hardware.Superconducting Magnets
Most MRI magnets today are built with several kilometers of superconducting wire, mostly
niobium-titanium (Nb-Ti) and niobium-tin (Nb3-Sn). Because in MRI we want image
contrast to be associated to tissue properties, we need extremely high magnetic
field homogeneity ( ΔB0 < 1 ppm). This stringent homogeneity specification
differentiates these magnets from those with similar strength used in particle
accelerators and nuclear fusion. In search of higher spatial resolution, higher
signal to noise ratio (SNR) and higher signal contrast (CNR), stronger magnets
are being developed for both human and animal studies.
Today we have field strengths for human MRI systems
up to 11.7 T and for NMR systems up to 32 T. Up to 21 T superconducting magnets are built with Nb3-Sn or Nb-Ti or a combination of
both. These low temperature superconductors
(LTS) need to be kept at temperatures as low as 2 °K. This requires a cryogen
system with large amount of liquid helium (He), and liquid nitrogen (N). Today,
thanks to advanced engineering of the cryogenic system, we can find almost helium-free (few liters) commercial
low field systems.
This is
important to overcome the decreasing supply and increasing price of He. Pushing
MRI to higher field strengths requires new superconductors materials. There is active research
on building magnets with high
temperature superconductors (HTS), the most promising so far being magnesium diboride
(MgB2). These HTS are kept to temperature close
to or above liquid nitrogen (> 77 °K) eliminating the need
of liquid He in the cryogenic system.
In addition to the magnet coils, an MRI system can have active shielding
coils, also built from superconducting wire, located at both ends of the magnet
coil to reduce stray fields outside the bore. To correct for unavoidable bulk B0 inhomogeneities passive shimming
can be performed in site by the arrangement of ferromagnetic small parts. To
compensate for B0 inhomogeneities
introduced in the MRI experiment, dynamic electrical shimming can be performed
through additional wired loops driven independently to generate different field
harmonics.Gradient Coils
MRI scanners have 3 sets of gradient coil arrangements integrated into an
annular layer inside the bore of the main magnet winding. Each of these coils
generate an independent spatial field change along x, y and z (Gx, Gy, Gz). The field change is linear in the imaging
volume. The z-gradient is generated by sets of circular loop pairs coaxially
arranged along z-axis and symmetrically spaced from the z=0 plane. The currents
on each side of the pairs run in opposite directions such as to increase
(parallel field) or decrease (anti-parallel field) the total magnetic field. In most MRI scanner today, the x-and y-gradients are generated by finger-print coils laid on a cylindrical surface concentrically
to the magnet coil. Both gradient coil layouts are similar but with axis
rotated 90 degrees. As in the z-gradient coil, loops are spatially symmetric
driven by anti-symmetric currents such as to generate fields
in parallel and antiparallel with B0 across
x and y direction. These
coils have an additional coil layer that minimizes eddy currents in surrounding
conductive surfaces specially the cryostat. Each of these gradient coils is
independently driven by a power
amplifier. In order to generate enough field change and spatially encode the
nuclear spin resonance, gradient coils can be driven by currents as high as 900
A. Gradient encoding only happens in a portion of the imaging cycle, which can
be as short as few milliseconds, and these coils are switched on-off at frequencies in the kHz well inside the audio range. Because
these large coils have
inductances on the order of thousands microhenries, the fast switching of the
currents induces voltages on the order of thousands
of volts. Gradient
hardware in the MRI scanner
handles the highest power of all other hardware (in the order of MW). At
kHz frequency range amplifier losses can be minimized by ultra-high-power efficient
switch-mode amplifiers using Pulse-Width-
Modulation (PWM). Different
from magnet coils,
gradient coils are not built with superconductors so at these high currents power dissipation in the wire can be in the order of several kW. Therefore,
water cooling is required. This is usually
implemented by chilled
water pipes in close contact
with gradient coil wires.RF Hardware
Of all the MRI hardware, the RF coil is the most accessible and
customizable hardware to improve image quality for a specific application. In
MRI scanners with field strengths up to 3 T there is a RF transmitter
integrated in the scanner bore known as body coil. This coil has a
cylindrical
shape and it is built from multiple rungs distributed along the cylindrical
surface and aligned in the direction of the static field (z-axis). Both
end-ring segments connecting each rung are capacitively or inductively coupled
to the next rung. This “birdcage” structure has 2 ports that are located 90 degrees azimuthally from each other
and receive high voltages with same amplitude but with a 90 degrees phase
delay. These voltages generate a sinusoidal current distribution along the rungs that creates a time
variant magnetic field (B1), on
resonance, rotating with the nuclear magnetization Therefore this Tx hardware
can generate a very homogenous excitation of the magnetization in a
well-defined volume. The body coil can be used as a Tx only or as a Tx-Rx coil.
It has been shown that receiving the MR signal with an array of surface loops,
instead of a volume coil, provides
higher SNR and image acceleration by skipping acquisition lines and taking advantage of coil spatial information.
This approach is known as parallel imaging. Most imaging protocols today use
the body coil as a Tx only coil and external Rx arrays with as many as 128
channels. At higher field strengths the shorter excitation wavelength
compromises the homogeneity achieved with this coil architecture, and other Tx
coil topologies need to be considered. One approach to achieve higher
homogeneity of the excitation field is to use an array of Tx coil independently
controlled. This method is known as parallel transmission (pTx). Some
ultra-high field scanners have already a pTx interface, up to 16 channels,
implemented in their control. The technology for these ultra-high field systems
continues to evolve to compensate for field homogeneities caused
by tissue properties of the body and to automate methods
for obtaining high quality
images while ensuring patient safety.Acknowledgements
References