Synopsis
Ultra-high
field (UHF) magnetic resonance imaging (MRI) enables functional brain images
with sub-millimeter spatial resolution. However, susceptibility induced magnetic
field (B0) variations within tissue are the source of various artifacts. Shim
coils of different shape and size are applied to reduce these B0 inhomogeneity.
However, most shim coils only have a regular shape and distribute current pattern
on a cylinder surface or close-fit helmet. The difference in performance
between current multi coil array and irregular shape multi coil array has not
been explored. The optimization methods for multi-coil shim arrays are
discussed, together with design and construction procedure.
Target Audience:
Ph.D. candidates and aboveIntroduction
Ultra-high
magnetic field strength provides several advantages, including much improved
T2* contrast and image signal-to-noise ratio (SNR). These advantages can be exploited
by acquiring data with reduced voxel sizes to produce sub-millimeter spatial
resolution. UHF MRI suffered from an inherent limitation: off-resonance scales
with the magnetic field strength. Accordingly, susceptibility artifacts are
worse in UHF MRI than at lower field strength.
Shim coils
are used to reduce B0 inhomogeneities by generating a compensating field to
counteracts the off-resonance field. In practice, this can be done by
whole-body spherical harmonic
(SH) shimming, which
generates a magnetic field with the spatial distribution described by spherical
harmonic functions up to the second or the third order (Eq.1)(1). While being useful, SH shimming cannot effectively
improve locally confined field inhomogeneity, where the required number of shim
coils and the SH order are high (2, 3). Alternatively, multi-coil
shim array using multiple small shim coils around the imaging object might be
applied (3). The generated local fields can contribute
effectively to counteract the existing B0 inhomogeneity. Such coils feature low
inductance and small size which is advantageous to dynamic slice-wise shimming (4, 5), integration into the RF
receive coil (6, 7), real-time correction of
temporal B0 alteration (8), generating spatial encoding magnetic fields (9), and novel parallel imaging methods (10).
$$∆B_0 (x,y,z)=B_{0,offset}+\sum_{I=0}^∞\sum_{m=-I}^{+I}C_{I,m}F_{I,m}(x,y,z)\space\space{(I)}$$
Multi-coil
shim arrays and integrated RF/B0 shim coil arrays (6, 7) typically have a simple shape, for
instance, a circular loop distributed in a regular pattern on a cylinder
surface or close-fitting helmet. Previous simulations demonstrate that increasing
the number of coils goes along with improved shimming capability (7), since more shim coils provide higher
degree of freedom in shim current design. In parallel, the increasing number of
coils would require extra amplifiers and consume more space within the bore.
Furthermore, difficulties in maintenance and troubleshooting are expected.
In the
brain region, the major field inhomogeneities are found near air-tissue
boundary, for instance, in prefrontal cortex and temporal lobe. Given the
similarity of human anatomy structure and the pattern of B0 inhomogeneity in
the human brain. Therefore, the shim array can be modified for better
performance on an identified target pattern. Previous studies placed the active
shim coils in the mouth (11) or over the nose (12, 13) to improve B0 uniformity in the
frontal lobe. Another study use genetic algorithm to optimize a shim coil with
irregular shape for 4 representative brain slices (14) and extended later for optimizing
position and geometry of 16 channel multi-coil array for 2 brain slices (15).
Here
we present two methods for the design of multi-coil shim arrays at ultra-high
field. Shim arrays are optimized to targeting the whole-brain B0 inhomogeneity,
while keeping the shim coil number in a reasonable range and feasible for in-house
construction1. Nonlinear constrained optimization for coil size and position (16)
Started
from a symmetric 32 channel multi-coil array on the cylinder surface. The coil
element had an identical square shape with a side length of 60 mm and 25 turns.
We optimize 3 DOFs of each coil (Figure 1A):
1) The
size, i.e. side length. Constrained between [20 100] mm
2) Axial
coordinate (Z) of the coil’s center on the cylinder surface. Constrained
between [-150 80] mm
3) Angular
coordinate (θ), the angle between the reference axis on a chosen plane and a
line from the origin to the projection of the coil center to that plane.
Unconstrained
We use a
nonlinear constrained optimization function, fmincon, in MATLAB (MathWork, Inc. Natick, MA, USA). The flowchart
of the optimization process is depicted in Figure 1B. First, we use a group of
8 B0 maps for training. Within iteration, the position and size were estimated
numerically by the gradient of a cost function (Eq. 2), which was defined as
the sum of the residual magnetic field of 8 training data, using sequential
quadratic programming (SQP):$$cost(τ)= \sum_{i=1}^{8}\sum_{j=1}^{32}(C_{ij}F_j(x,y,z)+B_i(x,y,z))\space\space(II)$$
,where
τ includes the size and position of
the coils in the current iteration, Bi
is the B0 map of the ith
brain, CijFj is
the magnetic field created by the shim coils for ith brain. The current of each coil was determined by a
constrained least-squares optimization, with a variable boundary from 1.5A to
3A according to the coil size. The optimization will be terminated when the
changes in the position and size of the coils are smaller than a defined
threshold.2. Stream function method(17)
The
current density on a surface is obtained from a stream function by taking the curl:
$$J(r_n )= ∇×[ψ(r_n)n(r_n)]\space\space(III)$$
The shim coil design begins by
defining two components(18): (a) a conductor shim coil surface S
where current flows, and (b) a region of interest (ROI) V, in
this case a group of 84 (12 subjects * 7 positions) field maps. The local shim
coil surface orientation is given by its normal vector at a given location . In general, the coil surface
consists of various sub-surfaces and sub regions. Here we use cylinder
geometry.
In order to
obtain a physically and technically meaningful solution, the power dissipated
by the shim coil was constrained using cgsvd
and lsqi in the MATLAB.
The final
wire pattern of shim coil array was obtained using singular values decomposition
(SVD) method. After solving the stream function for all training field maps, a
n*m matrix was formed. Where n
indicates the total number of nodes on the mesh and m corresponds to the total number of field maps entering the basis
generation algorithm. The resulting singular
vectors were used to obtain the wire pattern of coil elements of a shim coil
array.Acknowledgements
The author would like thank Ali Aghaeifar, Feng Jia, Jason Stockmann and
Bruno Pinho-Meneses for helpful documents. This work was supported by DFG SCHE658/13 and the Max Planck Society.
References
1.RA de Graaf,C Juchem.
CHAPTER 4 B0 Shimming Technology. Magnetic Resonance Technology: Hardware and
System Component Design: The Royal Society of Chemistry; 2016. p 166-207.
2.C Juchem, S Umesh Rudrapatna, TW
Nixon,RA de Graaf. Dynamic multi-coil technique (DYNAMITE) shimming for
echo-planar imaging of the human brain at 7 Tesla. Neuroimage 2015;105:462-72.
3.C Juchem, TW Nixon, S McIntyre, DL
Rothman,RA de Graaf. Magnetic field modeling with a set of individual localized
coils. J Magn Reson 2010;204(2):281-9.
4.C Juchem, TW Nixon, S McIntyre, VO
Boer, DL Rothman,RA de Graaf. Dynamic multi-coil shimming of the human brain at
7 T. J Magn Reson 2011;212(2):280-8.
5.A Aghaeifar, C Mirkes, J Bause, T
Steffen, N Avdievitch, A Henning,K Scheffler. Dynamic B0 shimming of the human
brain at 9.4 T with a 16-channel multi-coil shim setup. Magn Reson Med 2018.
6.TK Truong, D Darnell,AW Song.
Integrated RF/shim coil array for parallel reception and localized B0 shimming
in the human brain. Neuroimage 2014;103:235-40.
7.JP Stockmann,
T Witzel, B Keil, JR Polimeni, A Mareyam, C LaPierre, K Setsompop,LL Wald. A 32-channel combined RF and B0 shim array
for 3T brain imaging. Magn Reson Med 2016;75(1):441-51.
8.Topfer R, Foias A, Stikov N, Cohen‐Adad J.
Real‐time correction of respiration‐induced distortions in the human spinal
cord using a 24‐channel shim array. Magn Reson Med. 2018;80:935–946.
9.Umesh Rudrapatna S, Fluerenbrock F,
Nixon TW, de Graaf RA, Juchem C. Combined imaging and shimming with the dynamic
multi‐coil technique. Magn Reson Med. 2019;81:1424–1433
10.Scheffler K, Bause J, Loktyushin A,
Aghaeifar A, Steffen T, Schölkopf B. Spread‐spectrum magnetic resonance imaging. Magn
Reson Med. 2019;82:877–885.
11.JJ Hsu,GH Glover. Mitigation of
susceptibility-induced signal loss in neuroimaging using localized shim coils.
Magn Reson Med 2005;53(2):243-8.
12.C Juchem, TW Nixon, S McIntyre, DL
Rothman,RA de Graaf. Magnetic field homogenization of the human prefrontal
cortex with a set of localized electrical coils. Magn Reson Med 2010;63(1):171-80.
13.J Zhou, JP
Stockmann, N Arango, T Witzel, K Scheffler, LL Wald,FH Lin. An orthogonal shim coil for 3T brain
imaging. Magn Reson Med 2020;83(4):1499-1511.
14.Zivkovic I, Mirkes C, Scheffler K. B0
shimming at 9.4T using a multi‐coil approach—coil design with genetic algorithm. In 24th
Annual Meeting and Exhibition of the ISMRM, Singapore, 2016. p. 1152.
15.Zivkovic I, Tolstikhin I, Schölkopf B,
Scheffler K. B0 matrix shim array design‐optimization of the position, geometry and
the number of segments of individual coil elements. In 33rd Annual Scientific
Meeting of the European Society for Magnetic Resonance. Medicine and Biology.
Vienna, Austria, 2016:36.
16.A
Aghaeifar, J Zhou, R Heule, B
Tabibian, B Schölkopf, F Jia, M Zaitsev,K Scheffler. A 32‐channel
multi‐coil setup optimized for human brain shimming at 9.4T. Magnetic Resonance
in Medicine 2019;83(2):749-764.
17.F Jia, H Elshatlawy, A Aghaeifar, YH
Chu, YC Hsu, S Littin, S Kroboth, H Yu, P Amrein, X Gao, W Yang, P LeVan, K
Scheffler,M Zaitsev. Design of a shim coil array matched to the human brain
anatomy. Magn Reson Med 2020;83(4):1442-1457.
18.Peeren GN. Stream function approach for
determining optimal surface currents. J Comput Phys. 2003;191:305–321