Armin M. Nagel1
1Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-University (FAU) Erlangen-Nürnberg, Germany
Synopsis
Magnetic resonance imaging (MRI) techniques can usually be classified into spin-echo (SE) and gradient-echo (GRE) pulse sequences. In this presentation, the basic physical principles of GRE imaging, as well as different mechanisms to generate image contrast will be explained. Differences between SE and GRE MRI will be discussed. Additionally, the influence of different pulse sequence parameters (e.g. echo time, repetition time, flip angle; as well as spoiling techniques and preparation pulses) on the image contrast will be covered. Clinical applications of GRE imaging techniques will be shown exemplarily.
Target Audience
Clinicians,
technologists, and scientists interested in the fundamental physics and
clinical applications of gradient echo Imaging.Objectives
-
The basic principles and mechanisms to generate
different image contrasts in GRE imaging will be explained.
- Differences between spin-echo and gradient echo imaging
will be highlighted.
- Clinical applications of GRE imaging will be described.
Introduction
Magnetic resonance imaging (MRI) techniques
can usually be classified into spin-echo (SE) and gradient-echo (GRE) pulse
sequences. However, also hybrids such as the GRASE (gradient and spin echo)
technique exist (1). The physical principles and clinical applications of GRE
imaging have been reviewed in several articles (e.g. by Markl and Leupold (2) and
Hargreaves (3)).
In SE pulse sequences, refocusing RF pulses
are applied that lead to a re-phasing of the magnetization even in the presence
of static magnetic field inhomogeneities. Refocusing RF pulses generate
spin-echoes (4) and result in robust imaging techniques.
In GRE techniques, refocusing pulses are
lacking. As a consequence, shorter echo (TE) and repetition times (TR) can be
used. Thus, 3D imaging or other applications where fast data acquisition is
required are common applications of GRE imaging. In addition, the lack of
refocusing RF pulses in GRE imaging leads to a lower specific absorption rate
(SAR), which can be beneficial in high (B0 = 3T) and ultra-high
field (B0 ≥ 7T) MRI. Disadvantages of GRE imaging are
that these techniques are more prone to off-resonance artifacts. However, the
latter can also be exploited for diagnostic purposes, since signal extinctions
can also be caused by micro-bleedings. This effect is utilized in T2*- and
susceptibility weighted imaging (5).
Signal decay in GRE imaging is usually
described by the T2*-relaxation time, whereas in SE imaging, the decay of the
transverse magnetization is described by the T2-time. T2* relaxation contains
effects of dephasing caused by static inhomogeneities (T2’) and time-varying field
fluctuations (T2-relaxation). The relationship between T2 and T2* is described
by equation 1. The T2’-decay originates from temporally constant (“static”)
magnetic field inhomogeneities. These inhomogeneities can be caused by local
differences of the magnetic tissue properties. Dephasing caused by T2’-decay
can be fully reversed in SE pulse sequences but is not compensated in GRE
imaging. Thus, T2* is always shorter than T2.
$$\frac{1}{T_2^*}=\frac{1}{T_2}+\frac{1}{T_2'} (equation
1) $$Contrast Mechanisms in GRE Imaging
Image contrast in GRE imaging can be generated
by a number of mechanisms and techniques. These techniques include the adaption
of the basic pulse sequence parameters TE, TR and flip angle, the application
of spoiling techniques and contrast preparation.
TE, TR and flip angle: Whereas 90° excitation pulses are applied in
SE pulse sequences, the flip angles in GRE pulse sequences are usually much
smaller than 90°. For example, for a given TR (e.g. 50 ms) and short TE (e.g. 3
ms), image contrast can be changed from a spin-density weighted contrast to a
heavily T1-weighted image by solely increasing the flip angle. In this case,
small flip angles (e.g. 5°) lead to a spin-density-weighted contrast, whereas a
lager flip angle (e.g. > 20°) generates a T1-weighted image.
Spoiling techniques: The term “spoiling” describes different methods
(RF-spoiling or gradient spoiling) that can be applied to cancel (or “to
destroy”) transverse magnetization at the end of each TR interval. A
combination of gradient and RF spoiling leads to the FLASH-technique (6) (also
known as spoiled-GRE (SPGR), T1-Fast-field-echo (FFE)). FLASH imaging enables
pure T1 contrast and is often used for 3D contrast-enhanced MRI.
In balanced steady-state free precession (balanced
SSFP) (7), no spoiling is performed and all gradients are fully rewound. In
balanced SSFP, image contrast is based on T2/T1 and the signal-to-noise ratio
(SNR) is usually superior to the SNR of spoiled imaging techniques. This comes
at the expense of a higher sensitivity to off-resonance effects. An application
of the balanced SSFP technique is rapid morphological imaging, especially of
the beating heart.
Additional Modification of Image Contrast: Preparation pulses can be used to further
optimize the contrast for certain applications. For example, magnetization
preparation by an 180° inversion pulse followed by an delay time and a rapid 3D
GRE readout is used in magnetization-prepared rapid gradient-echo imaging (3D
MP-RAGE) to optimize image contrast (8). 3D MP-RAGE belongs to the most popular
sequences for high-resolution T1-weighted brain MRI. Applications include
voxel-based morphometry and contrast-enhanced imaging.Conclusions
GRE based
MRI techniques provide a large variety of image contrasts and are included in
most clinical imaging protocols. Compared to SE techniques, GRE is less robust
to off-resonance artifacts. The speed of GRE imaging renders it well suited for
rapid 3D imaging or real-time MRI.Acknowledgements
No acknowledgement found.References
-
Oshio K, Feinberg DA. GRASE (Gradient- and spin-echo)
imaging: a novel fast MRI technique. Magn Reson Med 1991;20(2):344-349.
- Markl M, Leupold J. Gradient echo imaging. J Magn Reson
Imaging 2012;35(6):1274-1289.
- Hargreaves BA. Rapid gradient-echo imaging. J Magn
Reson Imaging 2012;36(6):1300-1313.
- Hahn EL. Spin echoes. Phys Rev 1950;80:580-594.
- Haacke EM, Mittal S, Wu Z, Neelavalli J, Cheng YC.
Susceptibility-weighted imaging: technical aspects and clinical
applications, part 1. AJNR Am J Neuroradiol 2009;30(1):19-30.
- Haase A, Frahm J, Matthaei D, Hanicke W, Merboldt KD.
FLASH imaging. Rapid NMR imaging using low flip-angle pulses. Journal of
Magnetic Resonance (1969) 1986;67(2):258-266.
- Bieri O, Scheffler K. Fundamentals of balanced steady
state free precession MRI. J Magn Reson Imaging 2013;38(1):2-11.
- Mugler JP, 3rd, Brookeman JR. Three-dimensional
magnetization-prepared rapid gradient-echo imaging (3D MP RAGE). Magn
Reson Med 1990;15(1):152-157.