Terry Edward Perkins1, Jonathan Goodwin2, and Peter Greer3
1Sydney university, Rankin Park, Australia, 2Radiation Oncology, Calvary Mater Newcastle, Newcastle, Australia, 3Radiation Oncology, Calvary Mater Hopsital Newcastle, Newcastle, Australia
Synopsis
Using
a golden angle radial sampling technique (Siemens WIP#1104), the motion range
detection capabilities of 4D-MRI were compared to 4D-CT simulating lung tumour
motion using a MODUS QA 4D MRI phantom. Thirteen patient respiratory patterns
were used to simulate realistic breathing motion. Repeated measures ANOVA
(F=0.052, p=0.948), a post-hoc paired t-test (all p>0.677) and the RMSE (RMSE12
= 1.9 mm) were used to evaluate differences between the detected motion ranges.
Results showed no statistically significant difference between the motion range
detection capabilities of 4D-CT and 4D-MRI in the phantom. This observation
provides evidence for further evaluation in a patient population.
Introduction
The
introduction of 4D-CT allowed the successful clinical implementation of lung
SABR through the definition and measurement of each patient’s lung tumour
motion prior to treatment (1). However, this method suffers from low tissue
contrast, making visualisation of some tumours difficult, and according to Steiner et al. (2) may underpredict the
true extent of lung tumour motion by a factor of 1.7 times in the
superior-inferior (SI) direction. The aim of this study is to compare the
performance of existing 4D-CT techniques for phantom motion range definition
and visibility with a novel 4D-MRI approach, prior to implementation in a patient
cohort.Methods
The
QUASAR MR-4D phantom is an MRI compatible phantom which allows programmable
respiratory patterns (Varian Medical Systems, Palo Alto). Two 3 cm diameter
spheres act as a tumour surrogate, one of which has SI translational motion.
The Calvary Mater Newcastle’s (CMN) 3T MRI scanner (Siemens Skyra Magnetom) and
CT scanner (Siemens Somatom Confidence) were used to scan the phantom. 4D-CT
with Varians RPM system, and Siemens WIP-#1104 radial stack of stars MRI
sequence with motion control averaging (MoCo Av) reconstruction was used to
scan the phantom. The phantom was programmed with a sinusoidal pattern and 13
patient respiratory patterns from (3) to assess the motion range detection of
4D-CT and 4D-MRI. The 95% quantiles of the input respiratory signal were used
to assess the efficacy of the measured target motion range. The respiratory
pattern baseline drift (max to min of rolling 20 s average) and amplitude
variability (median peak-peak and interquartile range of 68.2%) were quantified
to determine their impact on total range detection. The MRI navigator and RPM
system signals were compared to the input waveforms using time series cross
convolution to ensure the viability of the respiratory signal used for phase
binning and data reconstruction. Repeated measures ANOVA was used to determine
if there was a statistically significant difference between the motion range
population means detected by the 4D-CT, 4D-MRI and the input respiratory
signal. The Root Mean Square Error (RMSE) was used as an additional
confirmation of the proximity of the 4D-CT and 4D-MRI. As Patient 12 was an
outlier in the data, both RMSE for all 13 patients (RMSE13) and
excluding Patient 12 (RMSE12) was included.Results
Results
for motion range detection are shown for the 13 patient respiratory patterns in
Figure 1. The repeated measures ANOVA result gave an F-statistic of F=0.052, with
a p=0.948. The RMSE12 = 1.9 mm and the RMSE13 = 2.6 mm.
With the post-hoc paired t-test results, all p-values > 0.65. In Figure 3,
the ‘MRI factor’ is the measured MRI motion range divided by the 95% quantiles
of the input respiratory motion in the phantom. Similar for the ‘CT factor’.
Amplitude variability is defined as the interquartile range divided by the
median amplitude, and baseline drift is measured as a proportion of median
respiratory amplitude.Discussion
For
the repeated measures ANOVA test, the low F statistic and high p-value suggest
that there is insufficient evidence to reject the null hypothesis (p>0.05)
with no statistically significant difference between the means of the paired
motion ranges of 4D-CT and 4D-MRI. The post-hoc paired t-test has shown that
using the 4D-CT measured motion range is a reasonable predictor for the 4D-MRI
motion range adding further evidence to the proximity of the measured motion
ranges from the two modalities when using the 4D-MRI phantom. The high p-vale
for the post-hoc paired t-test means failure to reject the null hypothesis and
the mean difference between paired observations is zero. To visualise this, the
corresponding mean of the differences between paired modalities is shown in
Figure 2 where the average difference is close to zero, also suggesting no
difference between the detected motion ranges between 4D-CT and 4D-MRI. Patient
12 has a high amplitude variability and high baseline drift with a mostly low
respiratory amplitude representative of someone in respiratory distress. This
is shown through poor acquisition of the total motion range from patient 12
(Figure 3) where the self-navigation technique has difficulty acquiring the low
amplitude respiratory signal. The Field of View (FOV) was altered to acquire a
high Spearman’s rank correlation (0.976) between the input respiratory signal
and the self-navigation signal however the final phase was a smearing of high
amplitude respiratory data. This is made more difficult by the small amount of
phantom that is moving into and out of the FOV, which the self-navigator relies
on. This is expected to improve as this study moves to a patient cohort.Conclusions
Whilst
the motion range detection between 4D-MRI and 4D-CT are similar in this phantom
study, further validation of the 4D-MRI technique for lung cancer patients is currently
being investigated. Preliminary patient results have demonstrated superior soft
tissue contrast and usability in shallow or erratic breathing patients, when
compared with the 4D-CT approach, providing further evidence for the inclusion
of 4D-MRI in the clinical workflow for lung cancer treatment.
Acknowledgements
No acknowledgement found.References
1.Acquiring a four dimensional computed tomography
dataset using an external respiratory signal. (S. Vedam et al.) Phys Med Biol; 48: 45–62 (2003).
2. Both 4D-CT and 4D-CBCT under-predict lung target
motion during radiotherapy (E. Steiner et al.) Rad. and Onc., 135, 65-73 (2019)
3. Lee, D., Greer, P. B.,
Ludbrook, J., Arm, J., Hunter, P., Pollock, S., . . . Keall, P. (2016).
Audiovisual Biofeedback Improves Cine–Magnetic Resonance Imaging Measured Lung
Tumor Motion Consistency. International Journal of Radiation Oncology, Biology,
Physics, 94(3), 628-636. doi:10.1016/j.ijrobp.2015.11.017
4. Grimm R. (2015).
Reconstruction Techniques for Dynamic Radial MRI. der
Friedrich-Alexander-Universität