Kurt Li1, Archana Machireddy2, Alina Tudorica2, Brendan Moloney2, Karen Oh2, Neda Jafarian2, Savannah Partridge3, Xin Li2, and Wei Huang2
1International School of Beaverton, Aloha, OR, United States, 2Oregon Health & Science University, Portland, OR, United States, 3University of Washington, Seattle, WA, United States
Synopsis
The goal of this study is to
compare the diagnostic performance between intravoxel incoherent motion (IVIM)
imaging parameters derived from standard biexponential IVIM model fitting and
those from direct calculation using only 3 b-values. This was accomplished
using Monte Carlo simulations and IVIM data acquisition and analysis from 27
patients with 28 suspicious breast lesions. Results show that at low
signal-to-noise ratio, the 3 b-value approach improved parameter accuracy and
provided comparable diagnostic performance in benign-malignant classification
compared to the biexponential fitting approach.
Introduction
The apparent diffusion coefficient
(ADC) from diffusion-weighted MRI (DW-MRI) is a quantitative imaging marker shown
capable of improving breast cancer diagnostic accuracy.1 The IVIM
(intra-voxel incoherent motion) approach for DW-MRI acquisition and analysis, which
quantifies pseudo-random perfusion from capillary bed and “true” random
diffusion in the tissue, is an emerging technique for breast lesion
characterization.1-3 A recent
liver IVIM-MRI study4 proposed to simplify data fitting by using
only three b-values for quantifying perfusion fraction (f) and the “slow”
diffusion coefficient (D), consequently saving time and cost. The goal of this
study is to compare the accuracies of IVIM parameters using simulations and the
diagnostic performances for breast cancer between the two approaches of standard
biexponential IVIM fitting and 3 b-value calculations. Methods
Simulation: A noiseless biexponential decay curve was created using
literature values5 for D* (pseudodiffusivity), D and f. Ten thousand Monte Carlo simulation runs were
performed at each signal-to-noise ratio (SNR) with the only difference between
runs being the random Gaussian noises added. In each simulation, the IVIM
parameters were obtained using the standard biexponential fitting with 12 b-values (0, 10, 25, 50,
75, 100, 150, 250, 450, 800, 1000, and 1200 s/mm2) and using the 3
b-value (0, 200, 800 s/mm2) method.4 Ten SNRs evenly spaced
from 10 to 100 were simulated.
Human
Data: Twenty-seven patients were referred for biopsies due
to mammography and/or ultrasound findings of 28 suspicious breast lesions. They underwent a 3T (Siemen) pre-biopsy MRI
study, which, in addition to DCE-MRI, included axial bilateral DW-MRI using a
single-shot SE EPI sequence with the aforementioned 12 b values applied in three orthogonal directions, 192x192
in-plane matrix, 32-34 cm FOV, 5 mm slice thickness (no gap), and ~8 min
acquisition time.
Data Analysis: Clinical
reading of the anatomic and DCE images was performed according to the ACR
breast MRI BI-RADS lexicon.6 For quantitative analysis, lesion ROIs were drawn on DW-MRI in reference
to post-contrast DCE images. The voxel DW-MRI data within the ROIs were analyzed
in three ways: 1) biexponential IVIM model1 fitting with all b ≤ 1000 s/mm2 values to extract D, D*, and f; 2) 3 b-value
approach4 to extract Dest and fest (est:
estimate); 3) monoexponential fitting using two b values (0 and 800 s/mm2)
to derive ADC. The lesion mean parameter value was calculated as the average of
the voxel values within the ROI.
Pathologically determined
malignant/benign lesion status was correlated with the MRI markers. A linear support vector machine7
was used to assess diagnostic performances for individual markers, as well as
combined markers constructed by concatenating individual markers to form
feature vectors with multiple dimensions. The ROC AUC values were
compared with the Hanley and McNeil method.8 Results
Fig.1 compares the relative mean errors
of the IVIM parameters derived in simulations at different SNRs between D and Dest
(1a), f and fest (1b), respectively. Relative errors for both sets
of parameters decrease with increasing SNR.
When SNR ≤ 50, the 12-b value biexponential fit generates larger
parameter errors than the 3-b value fit.
The errors from the two fitting approaches converge and reach a plateau near
zero when SNR ≥ 60.
Fourteen
lesions were pathologically proven malignant and the other 14 benign. Fig.2
shows post-contrast DCE images and color f, fest, D, Dest,
and ADC maps of a benign (top) and malignant (bottom) lesion, where the
malignant lesion demonstrated lower D, Dest, and ADC values compared
to the benign lesion. The mean ± SD values of the individual markers for the
two lesion groups are summarized in Table
1. Significant (P < 0.05)
differences were found only in ADC and Dest. Table
2 shows the ROC AUC, sensitivity, and specificity values for breast cancer
diagnosis. The clinical MRI reading
produced 86% sensitivity and 21% specificity.
For fair comparison, the specificities of all the other markers were
calculated with cut-off values that generated 86% sensitivity. No statistically significant difference was
detected in any pair-wise comparison of AUC values. The AUCs of Dest and fest are comparable to those of D and f, respectively, with Dest showing
the highest specificity. Further, the
AUCs of the two combinations of IVIM markers, obtained with either of the two modeling
approaches, were similar to each other, but higher than those of most
individual markers. Discussion
The measured SNRs for the human breast DW-MRI data ranged 25 to 50.
Results from the simulation and human data suggest that, at low SNR, the simple
3 b-value approach4 could deliver characterization and diagnostic
performance in breast lesions similar to the conventional IVIM-MRI method employing
many more b values. The “ill-conditioned” problem9 for biexponential
fitting at low SNR could be the major source of error for parameter estimation.
The 3 b-value approach, therefore, can be a simpler, time- and cost-saving
alternative for breast IVIM quantification. One drawback of this method is that
it doesn’t quantify D*. With a very limited cohort size, the study
findings need to be validated in a much larger population. Acknowledgements
Oregon Health & Science
University Center for Women’s Health Circle of Giving Award.
Thorsten Feiweier (Siemens) for
providing the work-in-progress sequence for IVIM-MRI data acquisition.
References
1.
Partridge
SC, Nissan N, Rahbar H, Kitsch AE, Sigmund EE.
Diffusion-weighted breast MRI: clinical applications and emerging
techniques. J Magn Reson Imaging 2017;45:337-355.
2.
Iima
M, Le Bihan D. Clinical intravoxel
incoherent motion and diffusion MR imaging: past, present, and future. Radiology
2016;278:13-32.
3.
Bokacheva
L, Kaplan JB, Giri DD, Patil S, Gnanasigamani M, Nyman CG, et al. Intravoxel
incoherent motion diffusion-weighted MRI at 3.0T differentiates malignant
breast lesions from benign lesions and breast parenchyma. J Magn
Reson Imaging 2014;40:813-823.
4.
Jalnefjord
O, Montelius M, Starck G, Ljungberg M. Optimization of b‐value schemes for estimation of the
diffusion coefficient and the perfusion fraction with segmented intravoxel
incoherent motion model fitting. Magn Reson Med (2019). Epub ahead of print.
5.
Liu
C, Liang C, Liu Z, Zhang S, Huang B. Intravoxel incoherent motion (IVIM) in
evaluation of breast lesions: comparison with conventional DWI. Eur J Radiol 2013; 82:e782-e789.
6.
"American College of Radiology. Breast imaging reporting
and data system (BI-RADS): breast imaging atlas," Reston, VA: American College of Radiology (2003).
7.
Cortes
C, Vapnik V. Support-vector networks. Machine Learning 1995;20:273-297.
8.
Hanley
JA, McNeil BJ. A method of comparing the areas under receiver operating
characteristic curves derived from the same cases. Radiology
1983;148:839-843.
9.
Kroeker,
RM and Henkelman RM. Analysis of Biological NMR Relaxation Data with Continuous
Distributions of Relaxation Times. J.
Magn. Reson. 1986; 69: 218-235.