Lisa J Wilmes1, David C Newitt1, Wen Li1, Evelyn Proctor1, Natsuko Onishi1, Teffany Joy Bareng1, Tom L Chenevert 2, Dariya I Malyarenko2, Patrick J Bolan3, Todor Karaulanov4, Nola M Hylton1, and Kathryn E Keenan5
1University of California San Francisco, San Francisco, CA, United States, 2University of Michigan Medical School, Ann Arbor, MI, United States, 3University of Minnesota Medical School, Minneapolis, MN, United States, 4QalibreMD, Boulder, CO, United States, 5National Institute of Standards and Technology, Boulder, CO, United States
Synopsis
A multiparametric breast phantom with an
integrated MRI-“visible” liquid crystal (LC) thermometer for real-time
monitoring of phantom temperature was evaluated at three imaging sites, on MRI
scanners from different vendors. Short-duration MRI acquisitions were optimized
for visualization of LC elements and implemented at the sites for temperature
measurements before and after DWI acquisitions. MRI-measured temperature
differences between different imaging experiments were reflected in different
water ADC values, calculated from DWI, that were consistent with the
relationship between ADC and temperature described in the literature.
Introduction
Multi-site data acquisition of MRI biomarkers is necessary
for clinical trials, and it is especially important to characterize the quality
of quantitative MRI measurements across different imaging sites. Previously, a
breast MRI phantom was developed that can be used to assess T1 relaxation time
measurements, quality of fat suppression, and the accuracy and reproducibility
of apparent diffusion coefficients (ADCs) calculated from diffusion-weighted
imaging (DWI)1,2. Especially in the case of ADC, the temperature at
the time of the measurement is necessary to assess the accuracy and
reproducibility, because diffusion is a thermally-dependent process. Previous
multi-site diffusion studies have successfully used ice-water to control the
temperature of the diffusion measurement3,4. An integrated liquid
crystal thermometer was developed that enables MRI-based measurement of
temperature5. We compared ADC measurements across multiple sites,
taking into account the temperature at the time of measurement.Methods
Breast phantom and integrated thermometer
Three
breast phantoms were constructed containing water and polyvinylpyrrolidone
(PVP) elements to mimic the range of human breast ADC values expected for
healthy tissue, benign tumor and malignant tumor tissues as previously
described1. In this updated version, a liquid crystal (LC) thermometer5
was included in the center of each of three breast phantoms, which were sent to
participating imaging sites. The liquid crystal thermometer contained ten vials
with LCs of different transition temperatures. A “visual calibration ” of the LC thermometer was
conducted by the manufacturer, QalibreMD, and the transition temperatures of
the ten consecutive LC vials ranged from 15.5C to 25.5C in approximately 1C
increments with +/-0.5C uncertainty (Figure 1) To determine the temperature from the MR image,
the viewer determines the first “on” or visible vial. Then, the recorded
temperature is the temperature of that vial +/- 0.5C uncertainty.
MRI methodsMRI data were acquired at three different sites on 3 T MRI scanners from the three main MRI vendors (GE, Philips, Siemens). Dedicated 8- or 16-channel breast coils were used at all sites. MRI data consisted of optimized 2D GRE and 2D FSE MRI thermometer measurements (acquisition time ~1 minute) acquired before and after multiple DWI acquisitions. DWI data were acquired using a standard two b-value (0,800 mm2/s) single shot echo planar imaging sequence at all sites. MRI data from all sites were centrally analyzed. Temperature was visually assessed on the central slice of MRI thermometer images, and ADC maps were calculated from DWI data using in-house software (Exelis Visual Information Solutions, Boulder, CO).
Temperature correction of ADCAt all sites, temperature was also measured using an external method in addition to the integrated thermometer (e.g. alcohol thermometer, liquid crystal thermometer strip on the exterior of the phantom, or a thermistor). Then, based on the work of Holz et al.6 to fit empirical data with the Speedy-Angel power law, the estimated water ADC was determined from:ADC = ADC0[(T/TS)-1]γ,where ADC0 = (1.635 x 10-8 +/- 2.242 x 10-11) m2 s-1, TS = (215.05 +/- 1.20) K, and γ = 2.063 +/- 0.051. The parameters determined by Holz et al are appropriate for water between 0 and 100 C.Results
MR images of the integrated thermometer showed that the
temperature at all three imaging sites did not change from pre- to post- ADC
acquisitions (Figure 1). A representative ADC map showing the ROI in the water
vial is shown
(Figure 2).
All three sites had water ADC measurements that agreed with
the Holz-model-predicted values (Table 2, Figure 3). The temperature determined
by the integrated thermometer did not always agree within the uncertainty with
the alternate external temperature measurement. However, the range of predicted
ADC values using the integrated thermometer or the alternative temperature
measurement did overlap.Discussion
This work
demonstrated the feasibility of imaging the integrated LC thermometer at
multiple imaging sites with different scanner configurations, using two
different commercial imaging sequences, with acquisitions times on the order of
one minute. When the relationship between MRI measured temperature and water ADC was evaluated, all sites measured ADC values consistent with the range of ADCs predicted for the measured temperature. It was also noted that the integrated phantom MRI thermometer measurement did not always agree with the alternative temperature measurement, which could be due to the different location of the temperature measurements (internal vs. external surface), or the temporal differences in measurements (alternative temperature measurements happened outside the scan room prior to or after the entire imaging session). Conclusion
It is important to account for temperature when
comparing DWI measurements across imaging sites. The real-time temperature
monitoring of phantom studies using the LC thermometer may help improve the
reproducibility of DWI data acquisition, by improving corrections for ADC
temperature dependenceAcknowledgements
Support: National Institutes of Health Grants: U01CA225427, 1R44CA235820, U01CA166104, P41 EB027061, R01 CA 190299References
-
Keenan
KE, Wilmes LJ, Aliu SO, Newitt DC, Jones EF, Boss MA, Stupic KF, Russek SE,
Hylton NM. 2016. Design of a breast phantom for quantitative MRI. Journal of
Magnetic Resonance Imaging, 44(3):610-619.
- Keenan
KE, Peskin AP, Wilmes LF, Aliu SO, Jones EF, Li W, Kornak J, Newitt DC, Hylton
NM. 2016. Variability and bias assessment in breast ADC measurement across multiple
systems. Journal of Magnetic Resonance Imaging, 44(4):846-855.
- Chenevert
TL, Galbán CJ, Ivancevic MK, Rohrer SE, Londy FJ, Kwee TC, Meyer CR,
Johnson TD, Rehemtulla and Ross Brian D. 2011.
Diffusion coefficient measurement using a temperature controlled fluid for
quality control in multi-center studies. Journal of Magnetic Resonance Imaging,
34(4): 983–987.
- Malyarenko DI, Swanson SD, Konar AS, LoCastro E, et al. 2019.
Multicenter Repeatability Study of a Novel Quantitative Diffusion Kurtosis
Imaging Phantom. Tomography, 5:36-43.
- Keenan KE, Mirowski E, Snow M, Stupic KF,
Russek SE, 2018. Liquid Crystal Magnetic Resonance Visible Thermometer. Proceedings of the International Society of
Magnetic Resonance in Medicine 25th Scientific Meeting. Paris,
France.
- Holz M, Heil SF,
Sacco A. 2000. Temperature-dependent self-diffusion coefficients of water and
six selected molecular liquids for calibration in accurate 1H NMR PFG
measurements. Physical Chemistry Chemical Physics, 2: 4740-4742.