Chuanli Cheng1, Qian Wan1, Yangzi Qiao1, Min Pan2, Changjun Tie1, Hairong Zheng1, Xin Liu1, and Chao Zou1
1Shenzhen Institutes of Advanced Technology,Chinese Academy of Sciences, Shenzhen, China, 2Shenzhen Hospital of Guangzhou University of Chinese Medicine, Shenzhen, China
Synopsis
Brown adipose tissue (BAT) is a special
adipose tissue which burns fat and dissipates energy in heat. As a result, heat
production capacity of BAT is considered to represent its activity directly. In
this study, we aimed to evaluate the accuracy of the proposed magnetic
resonance fat-referenced PRFS method on BAT in vivo in rat after activated by Norepinephrine.
The preliminary results show the PRFS based thermometry can describe the
temperature change in activated BAT accurately, implying that magnetic
resonance thermometry (MRT) is a useful tool to characterize the BAT activity.
Introduction
Since brown adipose tissue (BAT) burns fat
and dissipates energy in heat after activated, temperature measurement would be
a direct description of its function1. Proton resonance frequency shift
(PRFS) thermometry has been verified as a noninvasive temperature measurement
method, and it is very easy to implement in current clinical/preclinical MRI
systems2. Previously, dual-step iterative temperature estimation (DITE) fat-referenced
PRFS method has been demonstrated to measure the temperature change in BAT ex vivo
using a 3.0T MRI3. In this study, a calibration study was carried out to evaluate
the accuracy of the proposed DITE method on activated BAT in vivo in rat using
a fluorescence fiber optical thermometer as reference on a clinical 3.0T MR. In
comparison with the results from the thermometer, the measured average mean
error and RMS error using the DITE method in the interscapular BAT (iBAT) of
rat were −0.24ºC and 0.86ºC, respectively.Methods
Two SD rats with age of 11 weeks were used
in this study with IRB approval. Figure 1 shows the experimental setup of this
study. Before the experiment, an MR-compatible fluorescence fiber optical
thermometer was surgically inserted into iBAT location (Figure 1) with confirmation
of MRI scans. During the experiment, the rats were anesthetized by isoflurane (in
100% oxygen with a flow of 1.5 L/min). The MRI scans were completed using a
3.0T scanner with maximal gradient = 100mT/m (uMR 790, Shanghai United Imaging healthcare,
Shanghai, China) using an 8-echo gradient-echo sequence. A single-channel
rat/mouse surface coil was used to acquire images. MRI parameters are
summarized in Table1. The imaging slices were located in the interscapular
area. The scan time of each measurement was 5 minutes and a total of 60
successive measurements were acquired. The total acquisition time was 300
minutes. BAT was activated by injecting NE in a dose of 2 mg/kg through intraperitoneal
injection immediately after 12th measurement (60 minutes).
The proposed fat-referenced DITE proton
resonance frequency shift (PRFS) method was adopted to measure the temperature
change in iBAT area. The accuracy was evaluated by comparing the results to the
thermometer readings.Results
The temperature maps of one rat are shown
in Figure 2. The first echo magnitude image of the first measurement is shown
in Figure 2(a) with the location of the fiber optical thermometer. An iBAT ROI
is drawn around the fiber optical thermometer. The distributions of the
relative temperature change at the 5, 65, 125, 185 minutes after NE injection
in the ROI are overlapped in the magnitude image and are shown in Figure
2(b-e). Figure 3 shows the mean temperature change curve in the ROI over all of
the measurements for two rats. The temperatures estimated with the proposed
DITE method were consistent with the values measured with the thermometer. The
averaged mean error (ME) /RMSE for two rats were −0.36°C/0.88°C and -0.13°C/0.84°C,
respectively, over all of the measurements within the ROI.Discussion and conclusions
In this study, the temperature change in
iBAT after NE activation was successfully observed by the proposed DITE method
and its accuracy was demonstrated by comparing to the results from fiber optic
thermometer.
There exists a little deviation between the
measured temperatures by DITE method and fiber optical thermometer from FIG.3.
The reason may be that the accuracy of DITE method was compromised by the motion
of rat due to surgical operation. The physiological fluctuation may further
induce the susceptibility change in the small iBAT area after NE injection, and
thus yielding inferior results to the ex vivo experiment. Besides, the
experiments were conducted on a clinical 3.0T MR scanner and the accuracy was
limited due to the available SNR.
The present study demonstrated the accuracy
of the proposed method of measuring the temperature change in activated BAT in
vivo and it would promote the research of the heat product capacity of BAT noninvasively.Acknowledgements
This research was supported by the Natural Science Foundation of China, No.61901462, and the China Postdoctoral Science
Foundation, No.2019M650220.References
1. Sampath S C, Sampath S C, Bredella M A,
et al. Imaging of brown adipose tissue: state of the art[J]. Radiology, 2016,
280(1): 4-19.
2. Zou C, Tie C, Pan M, et al.
Referenceless MR thermometry-a comparison of five methods[J]. Physics in
medicine and biology, 2017, 62(1): 1-16.3.
3. Cheng C, Zou C, Wan Q, et al. Dual‐step
iterative temperature estimation method for accurate and precise fat‐referenced
PRFS temperature imaging[J]. Magnetic resonance in medicine, 2019, 81(2):
1322-1334.