Jin Zhang1, Karl Kiser1, and Sungheon Gene Kim1
1New York University School of Medicine, New York, NY, United States
Synopsis
It remains challenging to acquire 3D isotropic
high resolution T1 mapping
in in vivo MRI experiments. The
increase of T1 mapping
resolution and coverage are typically limited by the scan time. The purpose of
this study is to investigate the feasibility of estimate 3D isotropic high
resolution T1 mapping
using the variable flip angle (VFA) method1 and 3D-UTE-GRASP sequence2
for small animal imaging at 7T within 3 minutes.
Introduction
T1 mapping plays an important role in quantitative
dynamic contrast enhanced (DCE) MRI studies. Owing to the limited scan time for
in vivo studies, it remains
challenging to obtain high resolution T1
mapping for quantitative DCE-MRI analysis. In our previous study2, we developed the 3D-UTE-GRASP method for 3D high resolution DCE-MRI
study based on the Golden-angle Radial Sparse and Parallel (GRASP) technique3.
In this study, we investigated the feasibility of using the same 3D-UTE-GRASP
sequence, combined with the variable flip angle (VFA) method1 to
achieve 3D isotropic high resolution T1
mapping within 3 minutes. This method was tested for small animal imaging at 7T.Methods
Six to eight-week-old
C57BL6 mice (n = 5) (2 with GL261
mouse glioma models) were used for this study. MRI experiments were performed
on a Bruker 7T micro-MRI system, with 1H four-channel phased array
receive-only MRI coil. The mice were treated in strict accordance with the
National Institutes of Health Guide for the Care and Use of Laboratory Animals,
and this study was approved by the Institutional Animal Care and Use Committee.
Data Acquisition: T1 measurement was performed with three methods listed
in Table 1. In method I, the conventional UTE3D sequence provided by the vendor
was run three times (51,360 center-out spokes in each run) with flip angles 8o,
2o and 12o respectively. In method II, the UTE sequence
was modified to use the 3D golden angle sampling followed by image
reconstruction using compressed sensing and parallel processing, referred to as
3D-UTE-GRASP2 as reported in our previous study. The 3D-UTE-GRASP
was run for 3 minutes using the same three flip angles in series with 12,776
spokes for each flip angle. In method III, 2D RARE-VTR sequence was used to
measure T1 in three
orthogonal planes for comparison with the 3D methods.
T1 Map Estimation: For method I, fully
sampled images from 3D UTE with three flip angles were used to estimate T1 maps. For method II, the 3D-UTE-GRASP
image reconstruction2 was conducted to generate images with temporal
resolution T = 10 s (5 frames for each
flip angle with totally 15 frames). T1
estimation was carried out using the VFA method with the center frames from
each flip angle segment. For method III, the Image Sequence Analysis (ISA)
function of the manufacturer was used to obtain T1 maps with 64 x 64 matrix size and extrapolated to 128 x 128 to match the 3D maps from other methods. Results
Figure 1 shows an example of the 3D volume rendering
and mid-coronal/sagittal/axial slice images for method I/II/II. Figure 2 shows corresponding
estimated T1 maps in sagittal/coronal/axial
planes for each method. Figure 3 shows the Bland Altman plots for comparison
between the methods with pooled T1
values from the sagittal/coronal/axial directions as showed in Figure 2. It can
be seen that the proposed method (method II) has better consistence with Bruker
RARE-VTR measurement (method III) with the smallest mean or median difference
and the regression slope. The relatively larger difference was observed between
method II and method I; mean ± standard deviation = -7.5% ± 29.2%, with median
value -11.0% and 25 ~ 75 quartile range [-19.9% ~ -1.2%]. The mean and standard
deviation of relative error between method III and method I is -1.39% ± 33%,
with median value -6.7% and 25 ~ 75 quartile range [-14.4% ~ -4.0%]). Hence the
proposed method (method II) provided T1
values similar to that what RARE-VTR (method III) would do, but for a
substantially higher resolution and image matrix size within 3 minutes. Experiments from the other four mice showed
similar performance. Discussion and Conclusion
This study shows the possibility of using the 3D-UTE-GRASP
method for fast 3D isotropic high resolution T1 map estimation, which is essential for whole tumor
high resolution quantitative DCE-MRI studies as well as other T1-based tissue
characterization.Acknowledgements
NIH R01CA160620, NIH R01CA219964, P41EB017183, NIH/NCI
5P30CA016087References
- H. M. Cheng and G. A. Wirght, MRM 55:566-574, 2006
- J. Zhang, L. Feng, R. Otazo and S. Kim, MRM, 2019
Jan;81(1):140-152.
-
L. Feng, R. Grimm, K. T.
Block, H. Chandarana, S. Kim, J. Xu, L. Axel, D. K. Sodickson, R. Otazo, MRM,
2014;72(3):707-717.