Fenglei Zhou1,2, Francesco Grussu1,3, Zhanxiong Li4, and Geoff Parker1,5
1Centre for Medical Image Computing, Department of Computer Science, University College London, London, United Kingdom, 2School of Pharmacy, University College London, London, United Kingdom, 3Queen Square MS Centre, Queen Square Institute of Neurology, Faculty of Brain Sciences, University College London, London, United Kingdom, 4College of Textile and Clothing Engineering, Soochow University, Suzhou, China, 5Bioxydyn Limited, Manchester, United Kingdom
Synopsis
A spinal
cord-mimicking fibre phantom was developed to evaluate its potential for
validating dMRI. Microfibres were co-electrospun with two polymer solutions and
characterized by SEM. The phantom comprised two material samples designed with 0o
and 90o crossings. SEM revealed that fibres were uniaxially aligned
and hollow, having similar sizes to spinal cord axons. Diffusion tensor
analysis of a ZOOM dMRI acquisition demonstrated the difference in alignment of
the two samples. Diffusion kurtosis analysis demonstrated differences in axial
and radial diffusion restriction, with parameter values consistent with
published spinal cord data. Relaxation time constants were similar in two
samples.
INTRODUCTION
New approaches and
models have been developed to obtain the microstructure of spinal cord by dMRI.1
However, there has been no dedicated phantoms that can be used to validate diffusion
MR measurements in spinal cord. We have previously developed tissue-mimicking
MR phantoms from co-electrospun hollow microfibres.2-4 Here we
demonstrate a spinal cord phantom constructed from hollow microfibres and
assesses its potential as a tool for validating MR measurement on a 3T clinical
scanner.METHODS
Microstructural characterisation
and phantom construction
A ~0.5 mm thickness microfibre strip was fabricated
by co-electrospinning, as previously described.5 Pore
diameter was calculated from 5 SEM images for each sample using ImageJ.6
Two phantom samples were constructed by packing two blocks of 24 fibre layers
(10 mm×10 mm) into a 15 ml centrifuge tube filled with water having
approximate diameter of spinal cord (1-12 µm).7 Fibres in one block
were uniaxially aligned (0o) and in the other were interleaved, crossed at
90o.
The two samples assembled
into a cylindrical plastic container (inner diameter: ~140 mm; height: ~180 mm)
that can house up to 7 samples.
MR acquisition and
analysis
The phantom filled with water was scanned
on a 3T Philips Ingenia CX with the vendor’s 64-channel head coil to characterise
its diffusion and relaxation properties. The MRI protocol included: 3D
T2-weighted turbo spin echo for anatomical localisation (resolution: 1 mm
isotropic; TR: 2200 ms; TSE factor: 90; TSE echo spacing: 4.7 ms; TE
equivalent: 110 ms; compressed-SENSE factor of 7); 3D spoiled gradient echo
(SPGR) imaging for T1, T2* and proton density (PD) mapping (resolution: 1 mm
isotropic; TR: 28.5 ms; TE/echo spacing = 2.3/3.3 ms for 6 echoes; flip angles:
4° and 24°; compressed-SENSE factor of 6) and for B1 mapping (actual flip angle
imaging; resolution: 4 mm isotropic; TR1/TR2: 30/180 ms/ms; TE: 2.1 ms); ZOOM-EPI
DW imaging (reduced-field-of-view of 64x48 mm2;
resolution: 1x1 mm2; 12 3-mm thick slices acquired in 3 packages; TR
= 6000 ms; 1 b = 0 at each of TE = {20, 30, 60, 64, 90, 120, 200} ms and 24 directions
at each of b = {1200, 2500} s/mm2 with TE = 64 ms). Scans were
performed at room temperature (T = 23.3°C with intrinsic water self-diffusivity estimated to be 2.2 mm2
ms–1).
From two SPGR
scans with minimum TE and different flip angles we derived quantitative T1,8 correcting nominal flip angles with B1 map. Multiple SPGR echoes acquired at a flip
angle of 24° were instead
used to calculate T2*. Relaxometry metrics enabled the calculation of
quantitative PD with the method of pseudo proton densities,9 taking
care of normalising phantom’s PD to that of surrounding water. Then, diffusion
data were analysed by fitting DTI model to the lowest b-shell and DKI model to
the complete set of multi-shell measurements, deriving FA, axial/radial/mean
diffusivity (AD/RD/MD) and axial/radial/mean kurtosis. A quantitative T2 index
was also obtained by fitting mono-exponential decay to the set of b = 0 images
acquired at various TEs.
Model fitting was performed with
custom-written python code (T1, T2, T2*, PD) and with FSL dtifit (FA, AD, RD, MD) and DiPy (AK, RK, MK).
Finally,
a mask of two samples was drawn manually with FSLView and median and IQR of all
metrics was evaluated within the masks.
RESULTS
Fig.1 reveals axially-aligned hollow fibres and the range
of their pore sizes. There is a broad range of pore sizes within the phantom,
reminiscent of axonal distributions in spinal cord.7 The area-weighted
pore size of hollow fibres was 5.1 ± 0.4 μm, which is much smaller than those
used in brain phantoms.10
Figure 2 shows a
sagittal view of the two samples and the conventional DTI colour-encoding,
which highlights a much stronger directional coherence for the fibres within
the sample that does not contain crossings.
Figure 3 shows
quantitative maps in both samples (diffusion metrics were resampled to the SPGR
space for visualisation), while Table 1 reports median and IQR of all metrics. Diffusion
properties differ considerably in the two samples due to their different fibre
microstructure. The sample containing crossing fibres shows considerably
smaller FA and AD and higher AK than the other sample. Overall, DTI and DKI
metrics fall in a range that is compatible with values that could be observed
in vivo.11 The two samples show instead comparable relaxation
properties, with relaxation constants longer than that observed in spinal cord
white matter in vivo.12DISCUSSION
The
development of clinical spinal cord phantoms requires not only appropriate
microstructures and orientation, but also sufficient bulk material to be used
with clinical imaging protocols. The macro- and microstructures of microfibres can
be effectively controlled using co-electrospinning,5 and have similar
microstructure to spinal cord axonal structure, as demonstrated using SEM. The
similarity in geometry is reflected in the similarity between our phantom
diffusion parametrisations using a ZOOM acquisition and those seen in vivo.CONCLUSION
A novel spinal
cord-mimicking phantom has been developed, reflecting in vivo diffusion. The
phantom exhibits diffusivity and anisotropy that are in the range expected for spinal
cord, indicating that they can provide a helpful standard for diffusion
measurements on clinical MR scanners. Acknowledgements
This
research was supported by NIHR UCLH Biomedical Research Centre (BRC) grant. GJM
Parker has a shareholding and part time appointment and directorship at
Bioxydyn Ltd. which provides MRI services. FG is funded by the European Union’s
Horizon 2020 research and innovation programme under grant agreement No. 634541
and by the Engineering and Physical Sciences Research Council (EPSRC
EP/R006032/1). The QSMSC is supported by the UK Multiple Sclerosis Society
(grants 892/08 and 77/2017) and by the Department of Health's National
Institute for Health Research (NIHR) Biomedical Research Centres (BRC R&D
03/10/RAG0449). The contribution from Marco Battiston and Becky Samson to the
MR sequences is acknowledged. We thank Claudia Wheeler-Kingshott
for allowing us total freedom to acquire MR data. We also thank
Philips Healthcare for assistance in protocol development and for access to
research protocols. References
1.
Cohen-Adad J. Microstructural imaging in the spinal cord and validation
strategies. NeuroImage. 2018; 182: 169-183.
2.
Hubbard PL, Zhou FL, Eichhorn SJ, et al. Biomimetic phantom for the validation
of diffusion magnetic resonance imaging. Magn. Reson. Med. 2015; 73: 299-305.
3.
Ye AQ, Cristinacce Hubbard PL, Zhou FL, et al. In Diffusion tensor MRI phantom
exhibits anomalous diffusion. 36th Annual International Conference of the IEEE
EMBS, pp. 746-749. 26-30 Aug. 2014.
4.
Teh I, Zhou FL, Hubbard Cristinacce PL, et al. Biomimetic Phantom for Cardiac
Diffusion MRI. J. Magn. Reson. Imaging. 2015; 43: 594-600.
5.
Zhou FL, Li ZX, Gough JE, et al. Axon mimicking hydrophilic hollow
polycaprolactone microfibres for diffusion magnetic resonance imaging. Mater.
Design. 2018; 137: 394-403.
6.
Zhou FL, Eichhorn SJ, Parker GJM, et al. Production and cross-sectional characterisation
of aligned co-electrospun hollow microfibrous bulk assemblies. Mater. Charact. 2015;
109: 25-35.
7.
Duval T, Ariane Saliani A, Harris Nami H, et al. Axons morphometry in the human
spinal cord. NeuroImage. 2019; 185:119-128.
8.
Stikov N, Boudereau M, Levesque IR, et al. On the accuracy of T1 mapping:
searching for common ground, Magn.Reson. Med. 2015; 73:514-522.
9.
Volz S, Nöth U, Jurcoane A, et al. Quantitative proton density mapping:
correcting the receiver sensitivity bias via pseudo proton densities.
NeuroImage. 2012; 63:540-552.
10.
Zhou FL, McDowell A, Seunarine K, et al. Axon-mimicking hydrophilic fibre
phantoms for diffusion MRI, Proceedings of ISMRM 2019, Abstract No. 3636.
11.
Grussu F, Ianus A, Tur C, et al. Relevance of time-dependence for clinically
viable diffusion imaging of the spinal cord. Magn Reson Med. 2019;
81:1247-1264.
12.
Battiston M, Schneider T, Prados F. et al. Fast and reproducible in vivo T1
mapping of the human cervical spinal cord. Mag. Reson. Med. 2018; 79:2142-2148.