Dominik Ludwig1,2, Frederik Bernd Laun3, Karel D. Klika4, Mark Edward Ladd1,2,5, Peter Bachert1,2, and Tristan Anselm Kuder1
1Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany, 2Faculty of Physics and Astronomy, Heidelberg University, Heidelberg, Germany, 3Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany, 4Molecular Structure Analysis, German Cancer Research Center (DKFZ), Heidelberg, Germany, 5Faculty of Medicine, Heidelberg University, Heidelberg, Germany
Synopsis
Diffusion
pore imaging (DPI) can be used to retrieve the pore space function of arbitrary
closed pores. In this study we show that DPI of glass capillaries is possible
even under difficult experimental conditions. By separating the long gradient
into a CPMG-like gradient echo train and matching the magnetic susceptibility, it
was possible to acquire diffraction patterns of glass capillaries that were
placed orthogonal to the main magnetic field. Furthermore, the feasibility of
doing DPI in the presence of water outside the pores was demonstrated for the
first time.
Introduction
Diffusion
pore imaging enables the direct measurement of the shape of arbitrary closed
pores that are filled with an NMR-detectable medium [1-3]. The first
implementation of diffusion pore imaging on an NMR spectrometer [1] made use of
the long–narrow approach [4] in order to retrieve the phase information
necessary for reconstruction of the pore space function. To achieve correct
measurements for $$$20$$$ µm capillaries, a CPMG train of 180° pulses was used in
[1], where the long gradient with low amplitude was split into many short pulses.
Also, the short high-amplitude gradient at the end of the pulse train was spilt
into two parts separated by a 180° pulse. More recently, it has been shown that
it is also possible to use a spin echo approach with a readout gradient in
order to counteract gradient imperfections [3]. The aim of this study was to
show that diffusion pore imaging is possible under more realistic conditions
including magnetic field gradients introduced by susceptibility effects or
water outside the pores, the latter being demonstrated for the first time.Methods
Glass capillaries of
$$$3.5$$$ mm length
and inner diameter $$$25$$$
µm were
stacked horizontally inside a
$$$5$$$-mm NMR tube, i.e. with orthogonal orientation
to the main magnetic field. The capillaries were filled with a saturated
aqueous NaCl solution – necessary to weaken the susceptibility effects
introduced by the glass capillaries – resulting in a free diffusion coefficient
$$$D=1.48$$$ μm2/ms for the water inside the capillaries at $$$T=298$$$
K. For the first experiment, a Bruker Diff30 Probe ($$$G_\text{max}=18$$$ T/m) was used and any
residual NaCl solution on the outside of the capillaries was dispersed by
filling the NMR tube with Flourinert. A schematic representation of the pulse
sequence used for both experiments is presented in Figure 1. The long gradient
of the long–narrow approach was split into a CPMG-like gradient pulse train
separated by inserting 180° pulses. The long gradient was split into $$$39$$$ segments
of $$$d_l=10.221$$$ ms each, yielding a total time of $$$\delta_\text{L}=400$$$ ms. The narrow gradient had a duration of $$$\delta_\text{S}=1.9$$$ ms with $$$G_\text{S,max}=2.7$$$ T/m, resulting in $$$q_\text{max}=1370$$$ mm-1.
In a second experiment, the influence of an NMR
detectable medium surrounding the pores was investigated. The NMR tubes were filled
with the same NaCl solution that was used to fill the capillaries. For this
experiment, we used a Bruker TBI Probe with a maximum gradient amplitude of $$$G_\text{max}=0.68$$$ T/m in z-direction. The long gradient for this experiment was split
into $$$25$$$ segments of $$$\delta_l=16$$$ ms each, yielding a total time of $$$\delta_\text{L}=400$$$ ms. The narrow gradient had a duration of $$$\delta_\text{S}=6$$$ ms with $$$G_\text{S,max}=0.68$$$ T/m, resulting in $$$q_\text{max}=1080$$$ mm-1.
q-Values and
the sampling of the gradient ramps had to be selected in order to match the
available 16-bit raster of the gradient amplifier for all experiments.Results
Figure 2A shows the signal as a function of
the q-value together with simulated data. It is seen that the experimental
result agrees well with the matrix-based simulation. Figure 2B shows the pore
space function, which is the Fourier transform of Figure 2A. The pore space
function clearly yields a diameter of 25 µm, though it is slightly smeared out
at the edges.
The results for the measurement with NaCl
solution surrounding the capillaries are shown in Figure 3A. As the signal of
the first five measurements is highly dominated by signal from the external
water, as indicated by the purple squares in Figure 3A, the first part of the
curve was extrapolated by using a parabola as a rough expansion of the Gaussian
phase approximation:
$$S(q)=S_0\cdot(1-(q/a)^2),$$
where $$$a$$$ and $$$S_0$$$ were fitted using the first two reliably measured data points. The
extrapolated curve as well as the measured data points are shown in Figure 3B. Measured
data points are again in good agreement with the matrix-based simulation. The
extrapolated data points show a moderate deviation from the matrix-based simulations.
The reconstructed pore space function is shown in Figure 4. While the diameter
of the capillaries can still be determined, minor deviations from simulation
results are observed.Discussion
The need for strong gradients is the main limiting
factor of DPI. Nevertheless, we were able to demonstrate the feasibility of DPI
with an ultra-high gradient system using $$$25$$$ µm capillaries without exploiting
the full gradient strength available. Furthermore, we were able to show that
diffusion pore imaging is possible in the presence of water outside the pores. A
denser sampling near the first zero crossing might improve extrapolation.
Further, a very high extraporal signal fraction was used here; in biological
samples, this problem will be significantly reduced. Results might be improved
by implementing a bipolar gradient pair in the first part of the CPMG train. By
using this gradient pair as a filter for the water inside the pores that is more
restricted, it might be possible to partly suppress the surrounding water
signal, enabling the direct measurement of smaller q-values instead of
extrapolating them. This work may therefore pave the way to applications with
µm-sized objects such as living cells.Acknowledgements
Financial support by the DFG (Grant No. KU 3362/1-1) is gratefully
acknowledged.References
[1]: Hertel, Stefan, Mark Hunter, and Petrik
Galvosas. "Magnetic resonance pore imaging, a tool for porous media
research." PRE 87.3 (2013): 030802.
[2]: Hertel, Stefan Andreas, et al.
"Magnetic-resonance pore imaging of nonsymmetric microscopic pore
shapes." PRE 92.1 (2015): 012808.
[3]: Bertleff, Marco, et al. "1D and 2D
diffusion pore imaging on a preclinical MR system using adaptive rephasing:
Feasibility and pulse sequence comparison." JMR 278 (2017): 39-50.
[4]: Laun, Frederik Bernd, et al.
"Determination of the defining boundary in nuclear magnetic resonance
diffusion experiments." PRL 107.4 (2011): 048102.