Xinzeng Wang1, Daniel Litwiller2, Ali Ersoz3, Marc Lebel4, Sagar Mandava5, Lloyd Estkowski3, Arnaud Guidon6, Ann Shimakawa7, and Ersin Bayram1
1Global MR Applications & Workflow, GE Healthcare, Houston, TX, United States, 2Global MR Applications & Workflow, GE Healthcare, New York, NY, United States, 3Global MR Applications & Workflow, GE Healthcare, Waukesha, WI, United States, 4Global MR Applications & Workflow, GE Healthcare, Calgary, AB, Canada, 5Global MR Applications & Workflow, GE Healthcare, Tucson, AZ, United States, 6Global MR Applications & Workflow, GE Healthcare, Boston, MA, United States, 7Global MR Applications & Workflow, GE Healthcare, Menlo Park, CA, United States
Synopsis
PROPELLER
DWI, a FSE based DWI method, is increasingly used to reduce
susceptibility artifacts and motion artifacts. Multi-shot Echo-Planar diffusion
method also can reduce susceptibility artifacts, but PROPELLER DWI shows better
image quality where susceptibility artifacts are most problematic, such as in
skull base, head-neck and pelvis. However, the acquisition time is often longer
compared to ms-DW-EPI, therefore SNR is usually compromised to reduce
acquisition time. In this work, we evaluated a deep-learning based
reconstruction method (DL Recon PROP) intended to improve image quality and ADC
measurements by reducing the noise and artifacts without increasing acquisition
time.
Introduction
PROPELLER DWI (1,2), a Fast-Spin-Echo (FSE-) based diffusion weighted
imaging (DWI) is increasingly used to reduce susceptibility artifacts and
motion artifacts. The in-plane resolution is also improved compared to
single-shot DW methods. Multi-shot Echo-Planar diffusion method (ms-DW-EPI) (3)
also can reduce susceptibility artifacts, but PROPELLER DWI shows better image
quality where susceptibility artifacts are most problematic, such as in skull
base, head-neck and pelvis with metal implant (4). However, the acquisition
time is often longer compared to ms-DW-EPI, therefore SNR is usually
compromised to reduce acquisition time. The purpose of this work was to
evaluate a convolutional neural network (CNN) based reconstruction method to
improve the PROLLER DW image quality and ADC measurements by reducing the noise
and artifacts without increasing acquisition time.Methods
The CNN
based deep-learning network was trained for PROPELLER reconstruction ((DL Recon
PROP)) with over 10,000 high SNR and high spatial resolution images. The DL
Recon PROP was designed to improve the SNR and spatial resolution, and was
embedded into the conventional reconstruction pathway to generate two sets of
image series from a single set of raw MR data. The proposed DL Recon PROP
method was first evaluated with a standard NIST diffusion phantom to demonstrate
the artifacts and noise removal as well as quantifying the ADC measurements. The
diffusion phantom images were acquired using PROPELLER DWI with NSA = 8, 4 and
2, and a parallel acceleration factor
of 2. Since diffusion phantoms have long T2 relaxation times compared to most
soft tissues, the receiver bandwidth was intentionally increased to reduce SNR for
evaluation. Finally, the proposed DL Recon PROP method was evaluated and
compared against conventional PROPELLER DWI and ms-DW-EPI in the brain,
prostate and female pelvis of 5 healthy volunteers with IRB approval and
written informed consent. All images were acquired at a 3T GE MR scanner (GE Healthcare, Waukesha, WI).Results
The
diffusion phantom images reconstructed with conventional reconstruction method
and DL Recon PROP method were shown in Figure 1. Compared to the conventional reconstruction
method, DL Recon PROP increased the SNR and image sharpness, as shown in Fig.1
(T2 image). As expected, the SNR of b1500 images was reduced by reducing NEX
factor. However, the apparent SNR was increased using DL Recon PROP, and the
b1500 image with a NEX of 2 and DL Recon PROP was not visually inferior to
b1500 image with a NEX of 8 and conventional reconstruction method.
DL Recon PROP
also improved ADC measurements, as shown in Figure 2. ADC maps were generated
from Figure 1 with both conventional reconstruction method and DL Recon PROP. The
ADC maps reconstructed with DL Recon PROP also showed improved SNR and image
sharpness. Compared to the conventional reconstruction method, DL Recon PROP
did not change the mean ADC values across different NEX or different
concentrations of PVP, as shown in the representative plot of the measured ADC
values from vial with 30% PVP. DL Recon PROP produced the same mean ADC values
as the conventional reconstruction method, but with comparatively smaller
standard deviation.
The
representative PROPELLER T2 and b1000 brain images of a healthy volunteer were
shown in Figure 3. DL Recon PROP reduced noise and improved sharpness of both
T2 and b1000 images, resulting in a better ADC estimates with smaller
variations and fewer artifacts compared to the conventional reconstruction
method.
Multishot
DW-EPI and PROPELLER DWI prostate (Figure 4) and pelvis (Figure 5) images of
healthy volunteers were acquired without endorectal coil and gel. Multishot
DW-EPI images were acquired with 3 shots and reconstructed with original
reconstruction methods. PROPELLER T2 and high b-value images were reconstructed
using both conventional reconstruction method and DL Recon PROP method. The B0
inhomogeneities in the anatomies were large due to the presence of multiple
air-tissue interfaces, resulting in large susceptibility artifacts. Susceptibility
artifacts (Figure 4 and 5, red arrows) were shown in DW-EPI images, but not in
PROPELLER images. The PROPELLER images reconstructed using conventional method
showed low SNR. Howerver, DL Recon PROP significantly improved SNR and image
sharpness of both T2 and high b-value images, resulting in a better ADC
estimates with smaller variations. The peripheral zone (Figure 4, red arrow) is
more clearly visualized in the DL Recon PROP images than in the conventional
PROPELLER DWI images.Conclusion and Discussion
In this
work, we demonstrated that DL Recon PROP method to improve the image quality of
DW PROPELLER images without increasing acquisition time. In addition to
increasing apparent SNR, DL Recon PROP also improved image sharpness. DL Recon
PROP will not change the mean ADC values, while make it more precise.Acknowledgements
No acknowledgement found.References
1. James
G. Pipe, Nicholas Zwart. Turboprop: Improved PROPELLER Imaging. MRM.
2006;55(2):380-385
2. Xiaoli
Zhao, Zhiqiang Li, Ajeetkumar Gaddipati. PROPELLER DUO: Applied to Diffusion-Weighted
Imaging. ISMRM, 2009
3. Chen,
N.-K., Guidon, A., Chang, H.-C., Song, A.W. A robust multi-shot scan strategy
for high-resolution diffusion weighted MRI enabled by multiplexed
sensitivity-encoding (MUSE). NeuroImage. 2013;72:41-47
4. Czarniecki M, Caglic
I, Grist JT, Gill AB, Lorenc K, Slough RA, Priest AN, Barrett T. Role of
PROPELLER-DWI of the prostate in reducing distortion and artefact from total
hip replacement metalwork. Eur J Radiol. 2018;102:213-219