Neville D Gai1 and John A Butman1
1National Institutes of Health, Bethesda, MD, United States
Synopsis
While most brain imaging sequences now favor their 3D
counterparts, diffusion imaging is an exception. This is due to large
diffusion gradients resulting in increased sensitivity to motion exhibited by
3D acquisition. Prior schemes have used limited brain coverage and/or
triggering or acquired multiple 3D slabs along with modified reconstruction
schemes. The modified sequence used here employs first-order motion compensated
diffusion gradients in addition to real-time alignment to acquire whole brain 3D-DWI images as a single slab. Relatively shorter TE (using enhanced gradients) and TR along with other modifications result in
faster, reduced artifact diffusion images while providing higher SNR.
Introduction
MRI sequences have been trending in
favor of 3D acquisition over 2D multi-slice (MS) imaging. A majority of brain
imaging sequences (including Flair, T1 and T2-w, SWI, MPRAGE) now preferably
employ their 3D counterparts. Advantages to 3D acquisition (relative to 2D)
include finer slice definition, higher SNR/time and possibility of
obtaining isotropic image data sets for processing.
Diffusion imaging has been an
exception to this trend, in large part due to the difficulty of obtaining
artifact-free images. These artifacts arise because of sensitivity of
3D acquisitions to physiologic brain motion and CSF pulsations. In the 3D case, motion related phase
inconsistencies induced by diffusion gradients aggregate along the slice
direction, a problem not present in the 2D case.
Previous work in 3D has aimed at high resolution
limited coverage of specific regions of the brain and using longer scan times1,2. Other works have employed multiple 3D slabs (relying
on through plane component of motion-induced phase to change insignificantly
between TRs) and modified reconstruction to stitch together whole brain
diffusion images3-5. A majority of these implementations required
extensive reconstruction based post-processing, difficult in a clinical
environment. In this work, we implemented a clinically feasible whole brain 3D
diffusion weighted imaging scheme and compared it to 2D multi-slice DWI.Methods
Several sequence design
modifications were utilized for 3D-DWI acquisition. Modifications related to
reducing motion were as follows:
- Intra-TR
motion was reduced through use of first-order motion compensated diffusion
encoding gradients using enhanced gradient strength (max: 80 mT/m, SR100) to
keep TE relatively short. In addition,
shorter TR was employed.
- Inter-TR motion related phase effect was reduced by
monitoring motion on the fly by registering to a previous volume and adjusting
excitation and acquisition through phase offsets and rotation matrix.
Stronger fat suppression was affected using a 90°
spectral-spatial pulse of duration 6.4 ms.
Scanning:
Seven volunteers (3M, 4F) were scanned under an IRB approved
protocol on a Philips 3T scanner (equipped with dual mode gradients) using a
32-channel head coil.
Scan parameters were as follows:
MS-DWI: FOV=23x23cm
2,
res:2x2x2mm
3, slice gap=0, TR/TE=8190/75ms, ss-EPI, HFF=0.68, b=1000s/mm
2, SENSE(y)=2, ~74 slices, scan time:1:42.
3D-DWI: FOV=23x23cm
2,
res:2x2x2mm
3, TR/TE=325/78 ms, NSA=2, ss-EPI, HFF=0.64, b=1000s/mm
2, SENSE(y)=2.5, SENSE(z)=2, ~74 slices, scan time:3:15.
Processing:
To compare MS and 3D-DWI, post-processing was performed
in SPM and Matlab®. The
following steps were carried out.
- All images were registered to a common frame.
- The b=0s/mm2 images were segmented to
provide GM and WM maps.
- These maps were employed on directional images to
obtain segmented maps.
- Directional images were scaled by b=0 image
values on a pixel by pixel basis.
- Values were compared between MS-DWI and 3D.
- Directional images were combined to produce trace
and ADC maps.
- Relative SNR measurements were done for b=0 images by drawing corresponding ROIs in sub-cortical regions and outside the brain for several matching slices.
Results
Table 1 shows values measured in GM+WM for scaled directional images across all volunteers. Global difference between MS-DWI directional SM (measurement axis), SP (phase) and SS (slice-encoding) images and corresponding 3D-DWI images was 8.7%, 3.2% and -10.2%, respectively. Table 2 provides (mean+std) global ADC values for all volunteers. Difference between MS-DWI and 3D-DWI gray matter ADC was 12.6% and for white matter ADC was -4.6%. SNR corrected for scan time was 1.7 times higher for 3D-DWI compared to MS-DWI.
Sagittal images were reformatted in the transverse plane.
Figure 1 shows two such reformatted slices of diffusion weighted images along
M, P and S directions for seven volunteers for the two acquisition schemes (MS
and 3D). Figure 2 shows trace $$$\sqrt[3]{S_M\cdot S_P\cdot S_S}$$$ images for
two volunteers. Figure 3 shows sample ADC maps for two volunteers.Discussion
In
this work, use of gating or multi-slab acquisitions was eschewed in favor of
speed (since increased scan times defeat the very purpose of curtailing motion
artifacts) and motion-compensated gradients along with real-time volume
alignment. Nevertheless, large phase inconsistencies along z direction as a
result of residual eddy currents, motion and table vibration can result in
ghosting or signal loss. Reduction in echo time through use of enhanced
gradients can be prone to increased eddy currents and mechanical vibration
related artifacts. Through plane pulsatile CSF motion can corrupt neighboring
slices due to phase encoding along the z direction. This can affect
quantitative diffusion imaging.
Quantitatively,
diffusion images along the slice encode direction showed the highest bias when
compared with their MS counterpart. This bias on our system was a result of employing
enhanced gradients which was confirmed by scanning a phantom at two different
gradient strengths. For the sequences used here, SAR for 3D-DWI was lower at ~25%
of maximum when compared to MS-DWI (~71%). PNS was 65% for 3D-DWI and 79% for
MS-DWI.
In
summary, full brain single-slab 3D-DWI with
relatively high resolution, and good image quality in a clinically feasible scan
time, was achieved. Improved quantitative accuracy will be possible with future improvements
in hardware and mechanical stability. Extension of 3D-DWI to DTI currently
poses challenges related to increased scan time and mechanical stability.Acknowledgements
No acknowledgement found.References
1. Golay X, Jiang H, van Zijl PC, Mori
S. High-resolution isotropic 3D diffusion tensor imaging of the human brain.
Magn Reson Med 2002;47(5):837-843.
2. Frank LR, Jung Y, Inati S, Tyszka JM,
Wong EC. High efficiency, low distortion 3D diffusion tensor imaging with
variable density spiral fast spin echoes (3D DW VDS RARE). Neuroimage
2010;49(2):1510-1523.
3. Engstrom M, Skare S.
Diffusion-weighted 3D multislab echo planar imaging for high signal-to-noise
ratio efficiency and isotropic image resolution. Magn Reson Med
2013;70(6):1507-1514.
4. Chang HC, Sundman M, Petit L, et al.
Human brain diffusion tensor imaging at submillimeter isotropic resolution on a
3Tesla clinical MRI scanner. Neuroimage 2015;118:667-675.
5. Wu W, Poser BA, Douaud G, et al.
High-resolution diffusion MRI at 7T using a three-dimensional multi-slab
acquisition. Neuroimage 2016;143:1-14.