Maarten J. Versluis1, Kim van de Ven1, Velmurugan Gnanaprakasam1, Viswanath Kasireddy2, Suthambhara Nagaraj2, and Silke Hey1
1BIU MR, Philips Healthcare, Best, Netherlands, 2BIU MR, Philips Healthcare, Bangalore, India
Synopsis
In this study we compare SENSE accelerated multi-shot variable density
spiral diffusion to the current clinical standards: single shot EPI and MultiVane TSE diffusion. A variable density sampling strategy was employed to
correct for the phase of the different shots and iterative SENSE was used to
reduce the number of shots and scanning duration. This technique was applied on
a clinical platform with clinically acceptable reconstruction times. We showed
that spiral diffusion reduces distortions in difficult to shim brain regions
compared to ssh-EPI, and spiral diffusion has at a reduced scan duration compared to the TSE-based
approach.
Introduction
Diffusion weighted imaging (DWI) commonly uses a single shot
(ssh) echo planar imaging (EPI) readout preceded by diffusion sensitizing
gradients. The advantage is that this is a very efficient technique, resulting
in a short scan duration. However, due to the low imaging bandwidth ssh-EPI is highly
sensitive to inhomogeneous magnetic fields resulting in geometric distortions. For
anatomies or regions with a homogeneous magnetic field this is typically not an
issue, however in more challenging areas, such as e.g. spine, brain stem, or body
imaging
ssh-EPI-based
diffusion results
in severely distorted images, which can limit the application
in e.g. oncology(1). Using magnetic field inhomogeneity information can correct for the distortion(2), but areas of very
strong magnetic field variations remain non-diagnostic. A possible alternative
is to use a turbo SE (TSE) readout, which is geometrically accurate but is
suffering from a lower SNR, or much longer scan duration. In this study we have
implemented a variable density multi-shot spiral acquisition(3,4)
to overcome the drawbacks of a low imaging bandwidth, while having a moderate
scan duration increase, due to the high efficiency of a spiral readout(5–7). This technique is compared
against the most commonly used other techniques, ssh-EPI DWI and MultiVane (MV)
TSE DWI. Methods
After obtaining informed consent, 4 healthy subjects underwent three
DWI sequences on a standard Philips Ingenia Elition 3T
platform (Philips Healthcare, Best, The Netherlands). 1. DWI ssh-EPI, 2. DWI spiral and 3. MV TSE
DWI all with identical resolution and coverage: 1.5mmx1.5mmx5mm, 24 slices and max
b-factor 1000 s/mm2. For the spiral acquisition, a spiral out variable
density scheme was employed with a fully sampled center k-space radius of 10%,
which was used to apply intra-shot phase correction. The acquisition used the
standard gradient corrections implemented on the clinical system. No additional
gradient adjustments or eddy corrections were needed or applied. Reconstruction was
performed online on the standard clinical system and included deblurring based
on a B0 field map , iterative SENSE and intra-shot phase correction. An
overview of parameters is given in table 1. The DWI ssh-EPI was reconstructed
twice, with and without EPI geometry correction (using a B0-map and reversed
phase encoding direction).Results & Discussion
Figure 1 shows the comparison between the different
acquisition methods for a slice through the center of the brain of the b1000
image. The results clearly show the distortion as expected on the non-corrected
ssh-EPI acquisition (top right), which can be corrected at this location using
EPI geometry correction. The multishot spiral DWI matches the true anatomy
better, when comparing to the TSE DWI acquisition. The outline of the slice is measured
on the TSE-DWI image and copied to the other acquisitions. Figure 2 shows the
same comparison for a different subject and at a lower slice location. At this
level where strong magnetic field inhomogeneities are present, the ssh-EPI
acquisition shows pronounced distortions and areas of signal void and pile-up. The
overall shape can be recovered using EPI geometry correction, however it cannot
correct for the signal voids and signal pile-up. Multishot Spiral DWI matches
the anatomy of the MV-TSE DWI acquisition and does not suffer from areas of signal
loss or pile-up, similarly to the MV-TSE DWI acquisition, however at only half
the scan duration (2:08min vs. 5:10min, for spiral-DWI and TSE-DWI,
respectively).
The image deblurring algorithm used in the spiral
reconstruction can remove the majority of spatial blurring, however some
residual blurring is visible in figure 2, at the region of the temporal lobes
which exhibits strong spatial magnetic field inhomogeneities. This
effect is estimated to be much less detrimental in clinical imaging than the signal voids present at the EPI
acquisition.
Recon latencies are larger using Spiral DWI but still within
acceptable ranges and are expected to reduce further with the introduction of
more powerful processing hardware.Conclusion
Multishot spiral DWI is much less sensitive to areas in the
brain that exhibit high magnetic field inhomogeneities than ssh EPI based approaches.
Even if geometry correction is applied, ssh EPI still suffers from signal voids and pile-up.
MultiVane TSE DWI has a high anatomical accuracy, however at the cost of a
prolonged scan time. Spiral DWI is more robust to distortions than ssh-EPI and is more than twice as fast than MV-TSE DWI acquisition.
The combination of a variable density spiral acquisition scheme, with
intra-shot phase correction and iterative sense was implemented on a clinical
3T platform and is compatible with the clinical standard workflow.Acknowledgements
No acknowledgement found.References
1. White NS, Cancer Res. 2014;74:4638–4652 doi: 10.1158/0008-5472.CAN-13-3534. 2.Andersson, NeuroImage, 20(2):870-888, 2003. 3. Holtrop JL, J. Med. Imaging Bellingham Wash 2016;3:023501
doi: 10.1117/1.JMI.3.2.023501. 4. Liao C, Magn. Reson. Med. 2017;77:1359–1366 doi:
10.1002/mrm.26199. 5. Börnert P, Magma N. Y. N 1999;9:29–41 doi:
10.1007/bf02634590.
6. Li Z, Am. J. Neuroradiol. 2016;37:642–647 doi: 10.3174/ajnr.A4600. 7. Li Z, Magn. Reson. Med. 2020;83:170–177 doi: 10.1002/mrm.27911.