Theresa Reiter1,2, David Lohr2, Michael Hock2, Markus Johannes Ankenbrand2, Maria Roxana Stefanescu2, Maxim Terekhov2, and Laura Maria Schreiber2
1Department of Internal Medicine, University Hospital Wuerzburg, Wuerzburg, Germany, 2Chair of Cellular and Molecular Imaging, Comprehensive Heart Failure Centre (CHCF), University Hospital Wuerzburg, Wuerzburg, Germany
Synopsis
The establishment of human cardiac MRI in the setting of ultra
highfield MRI puts high demands on hardware, sequences and handling
of the examinees. Practical and safety relevant limitations are set
for extensive modifications regarding hardware and sequence. As a
consequence, human based factors play a paramount role in the success
of a routinely used cardiac imaging study at 7T. We report our
experiences from 71 volunteers with cardiac MRI scans with the aim to
increase the understanding of the practical implementation of cardiac
MRI and factors that influence and do not influence image quality in
this specific setting.
Introduction
Cardiac MRI at Ultra Highfield MRI is both highly
alluring and challenging, offering the promise of high resolution
myocardial images on one side, technical pitfalls such as pronounced
magnetic field inhomogeneities and sequence
limitations on the other side1-3.
Smaller cardiac studies have offered a glimpse on what we can expect
and hope for under ideal conditions4-6.
Larger studies without a focus on cardiac 7T MRI have demonstrated
the general acceptance in those examined albeit side effects such as
vertigo and nausea during positioning7,8.
The path towards more routinely used cardiac MRI at 7T is currently
still troublesome, as it still requires the definition of optimal
radiofrequency coils, pulse sequences and clinical practices.Motivation
The clinical benchmark with regard to the
usability of cardiac MRI at 7T is image quality. It ensures the
usability of the images, be it for scientific or for diagnostic
purposes. This, a sufficient diagnostic image quality is the
prerequisite for larger clinically orientated studies. Not all
factors that play a role for image quality are easily influencable or
are not even known at this point where a wider basis of clinical
studies is still missing. Understanding the current capabilities with
existing commercially available 7T MRI systems in
this setting thus becomes an important starting point for the future
optimization cardiac 7T MRI radiofrequency coil and pulse sequence
development, and general measurement setup.Materials and Methods
With permission of the local ethics committee,
cardiac 7T MRI was performed in 71 healthy volunteers (mean
age 32 years, weight 71 kg, 173 cm, BMI 23.6)
using a latest technology 7T system with 3rd-order
shim system (Magnetom Terra,
Siemens Healthcare, Erlangen, Germany) between 2017 and 2019. All
images were obtained using the vendor’s 1Tx/16Rx thorax coil. For
retrospective cardiac triggering the scanner’s vector ECG and, as
an alternative, the Easy ACT acoustic trigger system were used. The
imaging protocol was part of the implementation and research work at
the scanner, and among others consisted of survey, scouts, shim scans
and functional CINE imaging in short and long axis views. For B0
shimming, a dual gradient-echo sequence, provided by the vendor was
used. The shim volume covered the whole heart. For CINE imaging a
gradient echo sequence (FOV 340 x 320 mm, slice thickness 6.0 mm, TE
3.56 ms, flip angle 12 - 47°, Grappa factor 2 or 3) was used. In 40
volunteers, a full short axis stack was obtained , which covered the
complete ventricles with no interslice gap and 18 to 25 cardiac
phases. The diagnostic imaging quality of all full short axis stacks
were analyzed with a score focusing on artifact burden, noise and
overall quality. Variation of image quality as a function of
anterior and posterior coil positioning was analysed. Shim currents
and ECG trigger quality was obtained from the DICOM image headers.
Subjective acceptance of 7T MRI was analyzed, using a standardized
questionnaire and compared with 42 volunteers, who had received a
brain MRI at the same scanner.Results
The diagnostic image quality and left ventricular
function considered 680 left and 280 right ventricular segments. The
mean left ventricular ejection fraction was 53%. In the left
ventricle 14/680, in the right ventricle, 4/280 segments were non
diagnostic. Overall image quality scores in the left and in the right
ventricle were 2.25 and 2.47, respectively (Fig. 1). Acoustic and
ECG triggering were used equally often. A switch in the trigger
technique,
however, proved to be a useful tool to improve image quality when the
trigger system did not ensure reliable detection of the cardiac
activity and occurred in 20% of all scans
(Fig.2).
The
data from the shim coils covering first to third order of spherical
harmonics are given in Fig3. Coil position and deviations from the
optimal coil positioning did not significantly influence image
quality. The subjective acceptance showed that the duration of the
whole 7T exam (59%, mean value 3.02/10) was the most frequent, and
paresthetic discomfort was rarest (22.5%, mean value 3.38/10) reason
for discomfort. In addition, feelings of narrowness (38%, mean value
1.49/10), noise (40.8%, mean value 1.57/10) and vertigo during
movement (26.8%, mean value 1.31) were noted. In comparison, during
brain scans paresthetic discomfort (68,4%, mean value 3.45/10) was
the most frequent and worst remark.Discussion
With only 18/960
analyzed segments being non diagnostic, the overall image quality
allowed a diagnostic
interpretation in all included volunteers. The data from coil
positioning, shim currents and triggering showed that the
commercially available equipment allows to produce stable results for
cardiac CINE imaging. Both trigger options are currently helpful
during the initial preparation of the subject. However, it currently
appears helpful in some cases to have an triggering alternative. The
vendor based 3rd-order
B0 shimming shows satisfactory results, independently of suboptimal coil placement. The overall subjective acceptance for
cardiac studies consistently was good. However, it has to be noted
that the high end resolution images require further and in depths
focus on factors others than the human based ones.Conclusion
On the way to routine cardiac MRI at 7T, the commercially available
tools are already useful and robust, however, for highest resolution,
further improvements are needed.Acknowledgements
Financial support was obtained from the German Ministry of Education
and Research (BMBF) under grants: 01EO1004& 01EO1504.References
[1] Sievers et al.; (2011), Influence of the trigger technique on
ventricular function measurement using 3-Tesla magnetic resonance
imaging: comparison of ECG versus pulse wave triggering. Acta
radiologica 2011;52:385-392.
[2]
Stäb et al.; (2016): ECG Triggering in Ultra- High Field
Cardiovascular MRI. Tomography.org, Vol.2, No. 3, 2016
[3]
Frauenrath et al.; (2008): Acoustic cardiac triggering: a practical
solution for synchronization and gating of cardiovascular magnetic
resonance imaging at 7Tesla. Acta Acust United Ac (2008) 94:148–155
[4]
Prothmann et al.; (2016) High Spatial Resolution Cardiovascular
Magnetic Resonance at 7.0T in Patients with hypertrophic
Cardiomyopathy- First Experiences: Lesson learned from 7T. Plos
One 11(2):e0148066
[5]
Knobelsdorff-Brenkenhoff et al.; (2013), Assessment of the right
ventricle with cardiovascular magnetic resonance at 7 Tesla. JCMR
2013, 15:23
[6]
Alfakih et al.; (2003), Normal human left and right ventricular
dimensions for MRI as assessed by turbo gradient echo and
steady-state free precession imaging sequences. J Magn Reson.
Imaging. 2003; 17:323-29
[7]
Rauschenberg et al.; (2013), Multicenter Study of subjective
acceptance during magnetic resonance imaging an 7 and 9.4 T. Invest
Radiol 2014;49:249-259
[8]
Klix et al.; (2015), On the subjective acceptance during
Cardiovascular Magnetic Resonance Imaging at 7.0Tesla. Plos One
10(1):e0117095