Liyuan Jin1, Yaan Ge1, Qilin Lu1, Qingyu Dai1, and Kun Wang1
1GE Healthcare, Beijing, China
Synopsis
This
abstract investigated the method for quantitatively predicting B0 variation
induced by human body. Magnetic field simulation was performed, with human
phantom positioned in different strength’s B0 fields, and revealed the linear relationship
between patient-induced B0 inhomogeneity and the background B0 field. The linear
relationship was proved in clinical experiment, which used the relationship to
predict B0 field map and compared the predicted B0 field map with the real B0
field map. The clinical experiment shows that 95% voxels has a prediction
accuracy within 50Hz. The finding could further benefit image quality and
improve scan workflow.
Introduction
Patient induces B0 inhomogeneous in MRI scanning is well-observed;
however, there is a lack of knowledge in quantitative prediction of the B0
variation induced by human body. The human body has an overall magnetic
susceptibility close to air that causes subtle variation in B0, while susceptibility
that varied among different tissues can lead to complicate B0 variation in
space. In this study, we revealed a linearity relationship between patient-induced
B0 inhomogeneity and the background B0 field; and provided a method to predict B0
field in clinical scan.Methods
Magnetic
field simulation was performed on SIM4LIFE software to reveal the linearity
relationship between patient-induced B0 inhomogeneity and the background B0
field. Clinical experiment was conducted to prove the relationship in clinical setting.
Simulation - The magnetic field simulations
were performed using SIM4LIFE software. A close-loop solenoid coil model, provided
by the software, was used to simulate magnetic fields of various strength by injecting
different amplitude’s current into the coil. A computable human phantom (Duke_34y_m_v3.1b02_posable)
provided by the software was placed at the center of the solenoid coil,
simulating the setting of MRI scan. The magnetic susceptibility values of
different tissues were assigned to the computable human phantom.1,2 The
magnetic field maps with and without human phantom loaded were simulated, using
the EM LF Magneto Static Vector Pot. solver, at 8 different magnetic field
strengths (1.05T, 1.25T, 1.4T, 1.5T, 1.6T, 1.7T, 1.9T and 2.1T). The field maps
without human phantom loaded were used as the background B0 field, and the
differences between the field maps with and without human phantom loaded were
used as the B0 field variation induced by human. For each voxel on human
phantom, the data was used to further analyze the relationship between the
background B0 field and B0 field variation.
Clinical
experiment - Two volunteers
were scanned on a 1.5 Tesla MRI system (GE Signa Voyager) with the use of body
coil for clinical indication of quantitatively predicting patient-induced B0
variation. A large phantom filled with uniformed CuSO4 solution was scanned
to obtain background B0 field. Each volunteer was scanned twice at different
positions (position1 and position2) along z direction to obtain B0 field maps
with human-induced inhomogeneity. We used the volunteer scan data at position1
to predict the B0 field map, assuming the volunteer was at position2. The
predicted B0 field map was compared with the actual volunteer scan B0 field map
at position2.Results
Simulation - Figure 1 shows the simulation
results by randomly selecting 9 human phantom voxels and plotting B0 field
variations of those voxels under different background B0 fields in the selected
ROI with homogeneity B0 field. A linear relationship was observed and proved by
polynomial fitting with 95% confidence bounds and all the R-square values of
the fitting lines were more than 0.99. Figure 2 shows the fitted k and b value
of each voxel in a selected ROI.
Clinical
experiment - Figure
3 shows the B0 field maps acquired in clinical experiment. The data analysis
was conducted in the ROI showed as white boxes. The position1 B0 field
variation induced by human was calculated by subtracting the background B0 field
(Figure 3-E) from the position1 B0 field map (Figure 3-C). Figure 4 shows the predicted
position2 B0 field variation of two clinical experiments (Figure 4-G),
calculated using the linear relationship of B0 field variation and background
B0 field found in simulation. The difference map (Figure 4-H) between the
predicted and the actual position2 B0 field variations was calculated to
evaluate the prediction accuracy. Around 98% voxels have a prediction error
smaller than 50Hz, which can be seen on the histogram plot of the difference
map (Figure 4-J). The other clinical experiment result (Figure 4-K, L and N) shows
that 97% voxels have a prediction error smaller than 50Hz in difference map (Figure
4-N).Discussion
Simulation
results revealed the relationship between human-induced B0 variation (ΔB) and background field (B0) of each voxel in space
that could be represented as:
ΔB(x,y,z) = k(x,y,z) · B0(x,y,z) + b(x,y,z) (1)
In
clinical practice, the B0 variation could be calculated with field map with
human loaded (F) and the background field(B0). The function could be
represented as:
F(x,y,z) - B0(x,y,z) = k(x,y,z) · B0(x,y,z) + b(x,y,z) (2)
In
clinical scan, the k and b value in function (2) could be calculated with 2
pairs of scanning data. Considering that b value matrix had a noise pattern, we
simplified the function (2) as F(x,y,z) - B0(x,y,z) = k(x,y,z) · B0(x,y,z) and used this
function to predict field variation in our clinical experiment. The result of our
clinical experiment confirmed that the simplified B0 filed prediction method showed
an acceptable field prediction accuracy.
The
findings of this study could be incorporated in the calculation of changed B0
field in clinical scan caused by patient movement. The field prediction method
will be furthermore applied to image quality improvement and scan workflow
simplification, etc.Acknowledgements
No acknowledgement found.References
1.
Jain
V, Abdulmalik O, Propert KJ, Wehrli FW. Investigating the magnetic
susceptibility properties of fresh human blood for noninvasive oxygen
saturation quantification. Magnetic resonance in medicine. 2012
Sep;68(3):863-7.
2.
Garcia
MA, Baffa O. Magnetic fields from skeletal muscles: a valuable physiological
measurement?. Frontiers in physiology. 2015 Aug 10;6:228.