Andreas Georg Berg1 and Martin Börner2
1Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, Vienna, Austria, 2Karlsruhe Nano Micro Facility (KNMF), Karlsruhe Institute of Technology (KIT), Eggenstein-Leopoldshafen, Germany
Synopsis
Quality
control for systematic improvements in spatial resolution up to the microscopy
range becomes increasingly relevant not only for preclinical imaging but also
for High Field MR human scanners. The design, manufacturing technology, a
prototype phantom and an exemplary evaluation for qualitative and quantitative
measurement of the spatial resolution of MR-scanners claimed to offer spatial
resolutions better than 100µm is presented. The spatial resolution for 2D-FT
anisotropic imaging can be checked qualitatively and quantitatively by
interpolation of the Modulation-Transfer-Function in two different spatial
encoding directions at the same time with slice-thickness below 120µm up to 64
line-pairs/mm (pixel-size: 8µm).
Purpose and Introduction
The improvements in methodology of spatial
encoding (e.g. sensitivity encoding by multichannel rf-coils), hardware in
connection with optimized pulse sequence protocols for targeted organs in
principle allow for high nominal spatial resolution in MRI with isotropic pixel-sizes
in the microscopic range (ps<100x100µm2), not only on preclinical
animal and investigational vertical high-field MR-microscopy devices but also
recently even on human MR-scanners. However the spatial resolution of an
MR-imaging device is only restricted ultimately by the voxel size; the actual
spatial resolution is dependent on several different technical and sample
specific features, e.g. signal-to-noise ratio (SNR), line width (T2), susceptibility
and chemical shift differences, gradient-strength available, gradient switching
behaviour and shimming performance.
Further improvements in methodology, hard-/software
and proof of resolution quality demand for standardized objective quality
control on the actually achieved spatial resolution instead of simple voxel size
criterions. Concepts for the quantitative measurement of spatial resolution are
based on the Line-Spread-Function (LSF) and Modulation-Transfer-Function (MTF)1.
Design proposals for standardized resolution phantoms for slice-selective 2D-FT
do exist2,3. However, the proposed protocols and phantoms are designed
and prescribed for minimum sized structures typically down to about 0.7 mm 3.
Proposals and realizations for phantoms do exist for proving spatial resolution
up to 5 line-pairs (lp) per mm in one spatial encoding direction4,
but the corresponding plate-thickness of 100µm still does not allow for the
proof of resolution with pixel-sizes<100µm as possible on preclinical and
recently also on human MR-scanners.
We here present an investigational
resolution-phantom and evaluation concept, which allows not only for a quick
qualitative check on actually obtained spatial resolution with periodic micro-structures
down to 8µm, but also the quantitative evaluation of the MTF up to about 64 lp/mm
based on a set of grids with varying periodicity. The demands on spatial
accuracy, aspect-ratios, duty-cycle, reproducibility and mechanical stability
on these micro-structures (phantoms) including lamellae of 140µm depth at 1µm
thickness have been ambitious. The design concept, manufacturing based on
micro-structuring technology, evaluation procedure and an example of MR-evaluation
are shown within this contribution.Subjects and Methods
The
phantom-design is based on standard proposals for the quality control on human
scanners featuring a range of grid structures with reducing periods2,3 (fig. 1).
With
difference, the proposed test-device covers periodic structures between periods
of a1=256
down to a8=2µm for the coverage of spatial resolution up to the
best-performing MR microscopy scanners. The design is optimized for 2D-FT pulse
sequences with high in-plane resolution at comparable high slice-thickness. Two
orthogonal grid sets for independent analysis of e.g. phase and frequency
encoding direction feature strongly anisotropic grid bars (fig. 1b) with
lamellae depth to lamellae width (aspect ratio) better than 10. However such
anisotropic micro-phantoms can only be obtained by special micro-manufacturing technologies
as deep X-ray lithography (fig. 2) featuring high dose rates and very low
divergence. The corresponding challenge was met using a Synchrotron source at
KNMF (KARA, KIT, Karlsruhe, Germany), available for research5.Results
As
a result of several optimizing manufacturing set-ups with different PMMA
structural depth we obtained a prototype phantom with structural depth of 120µm
featuring mechanically stable lamellae with width between a1/2 = 128
down to a5/2 = 8µm (aspect-ratio: 15). Excellent duty-cycles (ratio
of cavity to bar-width) close to 1 and straight grid walls are obtained (fig.
1b). The phantom was checked for its suitability for qualitative and quantitative
quality control on the spatial resolution using a slice-selective 2D-FT spin-echo
pulse sequence with a pixel-size of 31x31µm2 at slice-thickness of
120µm (SNR6 ≈ 8,5; fig. 3). The MR-measurements are performed on a High-Field (B=7T) human MR-scanner
using a micro-imaging insert7. The slice could be adjusted to the
layer of the 2D-grid such, that both of the orthogonal grids could be observed
(fig. 3a). Visual qualitative inspection in vertical direction indicates, that
the grid period of a3=64µm (lamellae-width a3/2 =32µm)
could be differentiated. However the same grid in horizontal direction is
hardly to be visualized at pixel-size of 31µm.
The
evaluation of the relative modulation depth for the corresponding profile with fitting of the
pixel position to the grid modulation phase yields a value Mr = 0.25. According
to the resolution criterion (Mrmin=0,5!)8 this structure
is not resolved. By interpolation of the MTF the resolution can be quantified
(Mr(Kres)=0,5!): Kres≈10.4
lp/mm; resolution: a/2 ≈ 48µm.Discussion/Conclusion
The
conceptual design of a prototype phantom for quick qualitative and quantitative
checking of the spatial resolution for the MR-microscopic range in two
independent spatial encoding directions is presented. The orthogonal sets of grids
are designed for slice-selective MRI with anisotropic voxels with pixel-size between
128 and 1µm. A prototype phantom for this design using Deep-Xray-Lithography
and Synchrotron irradiation was manufactured. The second level optimized prototypes
featured smallest lamellae width of 8µm. The application of the evaluation concept
is demonstrated on an example for slice-selective 2D-FT microscopy on a human
high-field MR-scanner with a micro-imaging insert, showing quantitatively the
difference between pixel-size and actual spatial resolution. Prototype phantoms
can be borrowed from the author for quantitative checks of MR-apparatus for
objective quality control, proof of spatial resolution, hardware and software
improvements.Acknowledgements
This project was supported by the
Karlsruhe Nano Micro Facility (KNMF); project nr. 2015-013-006488; A. Berg; 2-3-Dimensional
μ-Structures for Magnetic Resonance Microscopy.
References
1 Rossmann K; Point Spread-Finction; Line Spread-Function and Modulation transfer
function; Radiology 93, 257- 272
(1969).
2 European
Norm EN 61223-3-7 (IEC 62B/575/CDV) 2005.08.01; Evaluation and routine
testing in medical imaging departments;
Part 3-7: Acceptance and constancy tests – Determination of essential image
characteristics of magnetic resonance
equipment (2005).
3 American College of Radiology; MR accreditation program; Testing Instructions, small phantom guidance (2018) (https://www.acraccreditation.org/-/media/ACRAccreditation/Documents/MRI/SmallPhantomGuidance.pdf).
4 Fellner C, Muller W, Georgia J, Taubenreuther U, Fellner FA, Kalender WA; A high-resolution phantom for MRI; Magnetic Resonance Imaging 19,
899–904 (2001).
5 Saile V, (ed.), Wallrabe U, (ed.), Tabata
O (ed.), Korvink JG (ed.); LIGA and its applications; in: Advanced
Micro & Nanosystems 7,
1; Wiley-VCH (2009) ISBN 978-3-527-31698-4.
6 Henkelman, RM; Measurement of signal intensities in the
presence of noise in MR images; Med. Phys.12, 232-233 (1984).
7 Berg A, Potthast A, Starewicz
P; MR-MICROSCOPY
ON A HUMAN 7T-SCANNER; Proc. ISMRM 2010, progr nr. 1048,
Stockholm, Sweden, 01.-07.05.2010 (2010).
8 Lerski
RA, de Certaines JD; II Performance assessment and quality control in MRI by
Eurospin test objects and protocols;
Magnetic Resonance Imaging 11 (6) 817-833 (1993).