Ria Forner1, Janot P. Tokaya2, Alexander Raaijmakers2,3, and Dennis Klomp2
1Radiology, UMC Utrecht, Utrecht, Netherlands, 2UMC Utrecht, Utrecht, Netherlands, 3Biomedical Engineering, Eindhoven University of Technology, Utrecht, Netherlands
Synopsis
This work is presenting a simulation
study on X-nuclei imaging at 7T with a birdcage body coil for sodium, carbon
and phosphorous. Three models and four imaging targets have been investigated.
Although B1+ efficiency generally
decreases with frequency, we show that the SAR efficiency, ( B1+/√SAR ) stays more or less constant over the investigated
frequencies.
Furthermore, it is shown that
although every model and every imaging target has one optimal relative phase
setting between the birdcage ports, for each frequency there exists one generic
phase setting that provides only a minimal drop in performance over all imaging
targets and models.
INTRODUCTION
At clinically relevant field strengths
such as 1.5T or 3T as opposed to 7T and beyond, proton imaging makes use of the
birdcage body coil as a general, large-field-of-view transmit coil and
reception is performed using a local receive array. The wavelength of X-nuclear
B1 fields is much larger and so, for these nuclei, the birdcage body coil can
still be used to provide a relatively uniform excitation at ultra-high field.
In
this abstract, birdcage body coils tuned to 120MHz for 31P, 79MHz
for 23Na, and 75 MHz for 13C at 7T were simulated with a
human male, female or child body model are presented. The resulting B1+ and SAR efficiencies in four organs were extracted for comparison: the heart,
the liver, the spleen and the kidneys. We have determined these metrics for the
optimal phase settings of the birdcage feed ports and for a generic phase
setting that aims to cover all investigated models and/or imaging targets.METHODS
All simulations are performed
using the numerical FDTD simulation package Sim4Life (ZMT, Zurich,
Switzerland).
The coil has an inner diameter of
60 cm, 24 rods at lengths of 40 cm, and an RF shield with 68 cm diameter. Lumped
element capacitors with parallel resistive losses were used with realistic Q
values to mimic coil losses. Three human body models (Duke, Ella, and
Thelonius, an adolescent male) were each inserted coaxially inside the three
high-pass birdcages (figure 1).
The power was normalised to 1W
total input power to assume a perfectly matched coil with no reflections at the
ports.
In post-processing a final check
of circular polarisation was performed by plotting the B-field vectors so as to
visualise polarisation. This was performed to obtain optimal settings for all
four organs, in all three body models and at all three frequencies. During
post-processing, the phase of the second port was stepped from 0° through 360°to ascertain the optimum (for
B1+) phase shim for each organ. Further, an average phase
setting was determined for each frequency and the resulting loss in efficiency
was determined when compared to the optimal phase setting for each specific
case of the organs. Gridding was manually adjusted to provide a minimum voxel
size of 2 mm3. The coil was driven in two ports separately using a
CW power source matched to the coil port impedance and the fields summed using
equal weighting. 3D B1+and SAR maps in the body for a
length of 10cm beyond the end rings of the coil were exported to MATLAB (information
about company and location of company) for analysis.
Results were analysed by
calculating the mean and standard deviation of B1+ in
each target organ.RESULTS
B1+ and SAR efficiency have been determined in each
model, organ and frequency. Results are presented in figures 2 and 3. Figure 4 shows the sinusoidal behaviour of
the mean B1+ for all organs, body models and the lowest and highest frequencies as the phase
difference between the two ports is varied from 0° through 360°. Although the
optimal phase settings vary, one generic phase setting can be identified for
each frequency at which the penalty shouldn't be too large.
Finally, in figure 5 we see
that this penalty for deviating from the best shim setting is insignificant. At
maximum, B1+ efficiency is reduced by 10%.DISCUSSION
The behaviour of the B1+ field follows similar
trends at 75MHz and at 79MHz, although markedly different at 120MHz. As
expected, at 120MHz, the mean B1+ is about half of that at the lower
frequencies. This can be attributed to the fact that the tissue loading is
increased at the higher frequency and thus, less field penetrates the body
models. However, this trend disappears when the B1+ field is normalised by the
square root of the peak local SAR.
As the optimal phase shim is approached, the B1+ mean
approaches the maximum while the homogeneity is also the highest. This is in
accordance with the fact that a circularly polarised field is achieved in most
of the organ of interest. The optimal shim settings for each organ in every
body model may differ by as much as 70° (at 31P , for Duke, 30° for heart and
100° for spleen) This may be explained by the inhomogeneous field distribution in the body due to variable loading as well as the locations of the organs.
It is possible to obtain a
trade-off value for this phase shim setting at every frequency such that the
performance is degraded by only between 5% to 10% irrespective of organ or
subject.
CONCLUSION
Although the B1+ decreases with frequency, the birdcage SAR efficiency
stays more or less stable throughout the investigated frequency range (75MHz
through 120MHz). It is possible to ‘shim’ the birdcage to obtain the maximum
possible B1+ in every organ in a body. However, these shim settings vary widely according to organ and
frequency and are very different from those used in clinical systems currently
at similar frequencies. Still it is possible to
determine a single optimal phase setting at every frequency such that the mean
B1+ in every organ is hardly reduced compared to the maximum possible B1+.Acknowledgements
This work was supported by: European
H2020-FETOPEN: NICI
References
No reference found.