Manouchehr Takrimi1 and Ergin Atalar1,2
1National Magnetic Resonance Research Center, Bilkent University, Ankara, Turkey, 2Department of Electrical and Electronics Engineering, Bilkent University, Ankara, Turkey
Synopsis
A
novel programmable set of z-gradient array equipped with an active-shield array
is introduced and simulated for diverse applications in magnetic resonance
imaging (MRI). The proposed active-shield
array can dynamically provide a proper shield for the main array in which its
magnetic profile can be programmed based on the required scenario. This is
achieved by using a set of independently tunable power amplifiers that feed
both array elements individually or in a cluster. As a proof of concept, two
sets of arrays are simulated: one for a double-gradient z-coil array and another
one possessing a large field of view (FOV).
Purpose
The
conventional application-specific gradient and its dedicated active shield sub-system
can be replaced by a set of novel programmable array versions. They are highly customizable
and offer wide range of new features in addition to conventional
functionalities.Method
Gradient
array have recently been used for shimming as well as spatial coding and for some
cases for both of them1-10 but regarding active-shield array, no
general solution has been proposed yet. Gradient active-shield array consists
of a series of pre-calculated wire patterns called coil elements. A coil
element may be fed individually by a dedicated power amplifier (PA), or a group
of elements or clusters can be fed by a common PA in a proper manner. Finding
an acceptable configuration for cluster positions under practical constrains
leads to nonlinear optimization problems. After positioning the array elements,
by calculating the magnetic field created by each element based on analytic or
numerical methods and performing appropriate optimizations, a set of suitable
feeding waveforms may be computed. The well-known target field method or a
combinational scenario of analytical and numerical solutions may be used for
better results.Results
A z-gradient
coil possessing a large FOV is ideally designed using Maple11
without taking into account the existence of cryostat. The coil consists of 10
pairs of 2x2 mm2 copper rectangular wires of diameter 680 mm and
height 1.2 m as shown in Fig. 1. The BZ field plot is shown in Fig. 2
where the contour lines indicate 0.5 $$$\mu$$$T separations. The RMS value for ten feeding
currents is 0.501 A leading to 25.4
$$$\mu$$$T/m/Arms
efficiency. The FOV diameter (at 0.3% deviation from linearity) is 600 mm. To
demonstrate the flexibility of proposed method, a simple array consisting of 10
pairs of uniformly-spaced and independently (but symmetrically) fed wire loops
of diameter 786 mm and total height of 1.6 m (look at Fig. 1) is used to design
an active-shield array for above mentioned z-gradient coil taking into account 10
mm thick cylindrical aluminum cryostat of diameter 910mm. The stray fields are
sampled 1.6 m along the surface of the cryostat to compute necessary feeding
currents. To demonstrate the effectiveness of active shielding in the presence
of cryostat, trapezoidal currents of 200 $$$\mu$$$s rise, pulse, and fall time with computed
amplitudes are fed synchronously to all array elements. Simulation results are
shown in Fig. 3 at peak input current for t=400$$$\mu$$$s and before the fall
time. The position of cryostat is also shown in the picture. Simulation shows
an efficiency of 8.7 $$$\mu$$$T/m/Arms
and 7.9% deviation from linearity within the same FOV. As a diverse application, a multi-gradient z-coil
array based on similar geometry but new wire positions and feed currents is
designed possessing two identical FOVs of diameter 300 mm, symmetrically
positioned at ±300 mm on the z-axis. The same uniformly-spaced active-shield
array with another set of optimized feeding currents is utilized to perform as
a shield. Feeding by a similar trapezoidal current as before, Fig. 4 illustrates
the BZ field for overall double FOV z-gradient array and its dedicated shield
at t=400$$$\mu$$$s. The contour lines indicate 0.2 $$$\mu$$$T separations. Simulation
shows an efficiency of 8.27 $$$\mu$$$T/m/Arms
and 10.1% deviation from linearity within both FOVs of 300 mm diameter. It
should be note that increasing the number of wires (ten wires in above
examples) or clusters greatly improves the overall efficiency of the shield and
makes it possible to optimize wide range of other important performance parameters. Discussion
In conventional designs where the main and shield coils are
connected in series, wire positions are calculated and optimized to satisfy
linearity, efficiency, gradient strength, shielding, and other performance constrains. Also the whole coil is fed by an
expensive yet powerful gradient power amplifier capable of delivering high voltages
in order to supply enough current for both coils. After construction, only the
feeding waveform can be changed to fine tune the performance of whole gradient
assembly. In the proposed active-shield array design, the wire positions were
set either uniformly or based on system-dependent optimizations and then an
optimized set of feeding waveforms (ten in above mentioned design examples) are
supplied by less expensive but independent power amplifiers to satisfy the
constrains. It is seen that a higher degree of freedom for fine tuning of overall
gradient assembly exists in this method. Note that the impedance of individual
array elements is much less than the whole coil and hence, higher switching
frequencies are also attainable.Conclusion
The
benefit of a matched set of z-gradient array and its dedicated active-shield
array is four folded: (a) it is highly customizable due to its dynamic nature; (b) it improves the overall imaging speed
since the impedance of the individual array elements are considerably less than
the main coil and its series connected shield; (c) magnetic field profiling is
possible for wide range of MRI applications; (d) if shielding is not concerned,
both of the main coil array and its shield array may be used to achieve very high gradient strengths or slew rates. We expect that this realistic simulation study will
lead to construction of the gradient system in a near future.Acknowledgements
No acknowledgement found.References
- Wintzheimer S, Driessle T,
Ledwig M, et al. A 50-channel matrix gradient system: A feasibility study.
Proc. of the ISMRM, Stockholm, Sweden. 2010:3937.
- Juchem
C, Rudrapatna SU, Nixon TW, de Graaf RA. Dynamic multi‐coil
technique (DYNAMITE) shimming for echo‐planar imaging of the human brain at 7
Tesla. NeuroImage. 2015;105:462–472.
- Juchem C, Green D, de Graaf
RA. Multi-coil magnetic field modeling. Journal of Magnetic Resonance. 2013 Nov
1;236:95-104.
- Juchem C, Nixon TW, McIntyre
S, et al. Magnetic field modeling with a set of individual localized coils.
Journal of Magnetic Resonance. 2010 Jun 1;204(2):281-9.
- Jia F, Littin S, Layton KJ, et
al. Design of a shielded coil element of a matrix gradient coil. Journal of
Magnetic Resonance. 2017 Aug 1;281:217-28.
- Littin S, Jia F, Layton KJ, et
al. Development and implementation of an 84‐channel matrix gradient coil.
Magnetic resonance in medicine. 2018 Feb;79(2):1181-91.
- Kroboth S, Layton KJ, Jia F, Littin S, et al.
Optimization of coil element configurations for a matrix gradient coil. IEEE
transactions on medical imaging. 2017 Aug 23;37(1):284-92.
- Ertan, K. , Taraghinia, S. , Sadeghi, A. and Atalar,
E. (2018), A z‐gradient array for simultaneous multi‐slice excitation with a
single‐band RF pulse. Magn. Reson. Med, 80: 400-412.
- Ertan, K, Taraghinia, S, Atalar,
E. Driving mutually coupled gradient array coils in magnetic resonance
imaging. Magn Reson Med. 2019; 82: 1187– 1198.
- Ertan, K., Taraghinia S, Sarıtas EU, Atalar E. Local optimization
of diffusion encoding gradients using a z-gradient array for echo time reduction
in DWI. In Proceedings of the 26th Annual Meeting of ISMRM, Paris, France,
2018. Abstract# 3194.
- Maple (2018.2). Maplesoft, a division of Waterloo
Maple Inc., Waterloo, Ontario.