Manuela B. Rösler1, Roger Luechinger1, David O. Brunner1, Markus Weiger1, and Klaas P. Pruessmann1
1Institute for Biomedical Engineering, ETH Zurich and University of Zurich, Zurich, Switzerland
Synopsis
Gradient induced heating is relevant for RF coils used in
recent high-performance gradient coils. In this abstract, we investigate the
predictability of gradient induced heating on the basis of theoretical gradient
field data. Furthermore, the heating of components usually used for coil
construction e.g. different plugs, cables and shielding materials under intense
gradient usage (EPI readout with 100 mT/m and 1200 mT/m/s at almost
100% duty cycle for 1 min) was measured. Depending on the position and
orientation we observe temperature increases up to 67°C for an N-plug and 12°C
for a coaxial cable.
Introduction
Since eddy currents scale up with increasing gradient
strength and slew rate, gradient induced heating is relevant for conducting
structures placed inside e.g. RF coils (1).
In this study, we investigate the heating of components commonly used in RF
coil construction depending on their position and orientation.Methods
Theoretically, the deposited power causing heating will be
proportional to the square of the gradient fields (2,3),
if the slew rate has a constant absolute value. For objects with a negligible
extension in one dimension, only field components perpendicular to its main
surface need to be taken into account. In this work, this relation is described
by f(B_grad) and is used to estimate temperature increase.
All measurements were performed on a whole body 3 T
scanner in combination with a high-performance gradient insert (4).
The device under test (DUT) was heated by an EPI readout with almost 100% duty
cycle as well as a gradient strength of 100 mT/m, a slew rate of
1200 mT/m/ms and an in-plane resolution chosen to reduce the plateau with
constant gradient strength to zero. Meanwhile, the temperature profile was
observed with optical temperature sensors (T1C-2M-PP10 + 4 channel reflex fiber
optic conditioner, Neoptix, Canada), which were fixed to the DUT with tape. Afterwards, the linear temperature increase (dT/dt) within 4 s at the
beginning of the heating period as well as the absolute temperature difference
(ΔT) after 1 min heating was evaluated. Measurements above 120°C were
interrupted manually.
First, the heating of a FR4 board with a 35 μm thick copper
area of 7 x 7 cm² was investigated at different positions and orientations
in the gradient tube as well as under different measurement gradient orientations. One
temperature sensor was placed in the middle and three others at the edges of
the copper area.
Afterwards, this experiment was repeated for other DUTs centered
at position x = -15 cm, y = 0 cm, z = -17 cm relative to the
gradient isocenter:
-
FR4 boards with a 35 μm thick copper area of 5 x
5 cm² and 4 x 4 cm²
- A conductive textile (4715-Conductive textile
tape with standard adhesive, Holland Shielding, The Netherlands) with an area
of 7 x 7 cm²
- Different straight cable RF plugs (HUBER+SUHNER,
Switzerland) orientated parallel to the z-axis of the magnet. The temperature
sensor was fixed at the shell.
- Different RF cables (HUBER+SUHNER, Switzerland) with
a length of 30 cm routed parallel to the z-axis of the magnet. If
applicable, the plastic jacket was removed locally to place an additional
temperature sensor directly on the metallic shield. Temperature sensors were
fixed in the middle of the cable.
For comparison, the FR4 board (7 x 7cm²) was
also measured in the default gradient system (200 mT/m/ms and 40 mT/m)
at one location. Since exact field data is not available for this
gradient, the board was positioned at x = 32 cm, y = 0 cm and z =
-40 cm, where based on the technical description high field values are
expected.
Results & Discussion
The typical temperature profiles as shown in Figure 1
indicates, that the size of the copper area is too big to neglect heat
transfer. However, the chosen size was a compromise between sensitivity at
locations with limited heating and accuracy. Gradient field deviations within
the board area were checked to be small, but could be an additional error
source.
As expected from theory, the proportionality factor (figure
2b) between the measurements and f(B_grad)
is independent of gradient axis and sample orientation (figure
2a). Nonetheless, the confidence value is only 0.92, which might have been
caused by limitations in positioning, sample size, variations in cooling
effects (airflow) and the temperature measurement method itself (fixation and
positioning of sensors). Obviously, the linearity factor must be calibrated for
every object.
The wanted gradient fields in z-direction used in MRI would
only heat the board, if oriented perpendicular to the z-axis. However, for the
gradient insert field data indicate highest eddy currents when placed
orthogonal to x-axis at x = -15 cm, y = 0 cm, z = -17 cm.
Measured data (figure 3) confirms that x and y components of the gradient
field, which also lead to known concomitant field, are not negligible.
Larger components and ticker cables heat up faster as
summarized in table 1. Reducing the area of conducting structures (e.g.
slotting) and conductivity at low frequencies (conductive textile (5))
minimizes the temperature increase.
The slew rate of the insert gradient is 6-times higher than
the one of the default gradient system, which would result in 36-times higher
power deposition. However, only an approximately 9-times higher initial
temperature increase (dT/dt 4.3°C/s vs 0.47°C/s) was observed. This might be
explained by the smaller diameter and shorter length in z-direction as well as
unknown non-z-components of the standard system.
As demonstrated, temperature measurements could localize
structures with strong eddy currents in RF coils and help to reduce image
artefacts.Conclusion
In high-performance gradients eddy current induced heating
is significantly increased and may lead to serious problems. It might as well
become more relevant for implants.Acknowledgements
No acknowledgement found.References
1. Rösler
MB, Weiger M, Brunner DO et al. An eight-channel array coil for zero echo time
imaging. In Proceedings of the 27th Annual Meeting of ISMRM, Montreal, QC,
Canada 2019. p. 438.
2. Graf H, Steidle G, Schick F. Heating
of metallic implants and instruments induced by gradient switching in a
1.5‐Tesla whole‐body unit. Journal of Magnetic Resonance Imaging: An Official
Journal of the International Society for Magnetic Resonance in Medicine. 2007;26(5):1328-1333.
3. El Bannan K, Handler W, Chronik B,
Salisbury SP. Heating of metallic rods induced by time‐varying gradient fields
in MRI. J Magn Reson Imaging.
2013;38(2):411-416.
4. Weiger
M, Overweg J, Rösler MB et al. A high-performance gradient insert for
rapid and short-T2 imaging at full duty cycle. Magn Reson Med. 2018;79(6):3256-3266.
5. Weiger
M, Brunner DO, Schmid T et al. A virtually 1H‐free birdcage coil for
zero echo time MRI without background signal. Magn Reson Med.
2017;78(1):399-407.