Yihe Hua1, Desmond T.B. Yeo1, and Thomas K.F. Foo1
1GE Global Research, Niskayuna, NY, United States
Synopsis
Head-only gradient coil is not only more power efficient than the whole body system for high quality MR neuroimaging, but also reduces risk of peripheral nerve stimulation. In this abstract we compared two different head-only transverse gradient coil designs -- a non-folded design and a patient-end folded design, pointed out that the current turns on the connection region of the folded design that close to human body is more efficient to induce electric field and eddy current than the non-folded design in the neck, shoulder, chest and back. A connection-region turn-constrained folded coil design confirmed the observation.
Introduction
High
quality MR neuroimaging is important to advancing our understanding of the
brain and to improve clinical diagnosis. Since the FOV needed for brain imaging
is about 22cm to 24cm, which is much smaller than that required in whole-body
applications, smaller and more efficient head-only gradient coils that are
closer to the patient can be designed to produce higher quality images [1]
with lower power requirements. The use of head-only gradients for advanced
neuroimaging also reduces the risks of peripheral nerve stimulation (PNS) when
compared to whole-body gradients, which is important especially at high gradient
strengths and slew rates [1,2]. Although human studies have been
done to compare PNS thresholds across different MRI systems, the impact of
gradient coil design on PNS has not been thoroughly studied.
Folded
gradient coil design [1,3], that have conductors on, or across, the
connection region usually leads to even lower inductance and higher efficiency
compared to separate primary layer and shield layer coil designs (non-folded
coil). However, the impact of these coil architecture choices on PNS
performance is not well documented. Here, we compared the induced electric fields and eddy currents in a realistic human body model for non-folded and folded asymmetric
head gradient designs. Method
In this abstract, we only compared the x transverse coil design of the most efficient folded
architecture (folded on the patient-end) to the non-folded architecture since x head gradient coil usually provides the lowest PNS threshold among the x,y,z coils. The dimensions and
the specifications in our designs are similar to those in the reference paper [3]
but simplified without loss of generality. They are summarized in Tab.1. A
Boundary Element Method was used to find the optimal current distribution [4].
The wire patterns after quantization of the non-folded and folded designs can be
found in Fig.1(a) and (b). After being normalized to $$$10 mT/m$$$ , the magnetic field ($$$\vec{B}$$$-field) and corresponding vector
potential field ($$$\vec{A}$$$-field) of the designs were calculated
directly by the Biot-Savart Law and used to calculate the electric field and
eddy current distribution inside of the Duke human body model (IT'IS Foundation)
under a 1000 Hz sine wave excitation with the commercially available Sim4Life
software (Zurich MedTech, Switzerland). Duke model was positioned in the coils with whose glabella being about at the iso-center of the FOV.Results and Discussions
Fig.
2 (a) and (b) show the $$$|\vec{E}|$$$-fields of the non-folded and folded
designs. For the voxels of same $$$(x,z)$$$ coordinates,
only the maximum $$$|\vec{E}|$$$-value in $$$y$$$ direction in the model is shown. It is
noted that $$$|\vec{E}|$$$
is higher in the glabellar region for non-folded
architecture, but in the folded coil case, it is higher in the neck, shoulder, chest and back. Fig. 3(a) and (b) are the flow
plots of the current density $$$\vec{J}$$$ in the Duke
model for the two gradient architectures, the color shows its amplitude. Only the right half of the body is shown since the
geometry is about left-right symmetric. The $$$\vec{E}$$$-fields in
the bone, cartilage and cerebrospinal fluid were set to zero to avoid
the non-interested high values [2,5], and the same rule was also applied to $$$\vec{J}$$$-flow plots. The $$$\vec{J}$$$-flow plots confirmed the $$$|\vec{E}|$$$-fields.
It
can be easily seen that there are current turns on the connection region (Fig.1(b))
between the primary and shield layer for the folded architecture. While these
turns are efficiently contributing to establish the gradient field, they are
also very close to the human body such that they can induce electric field and
eddy currents in the body more easily. Under such circumstances, a 3rd
design was performed to constrain the current density on the connection region
of the folded architecture so that only half number of turns are in this
region compared to the original folded design. The wire pattern of this design is shown in the
Fig.1(c) and the amplitude of $$$\vec{E}$$$ and $$$\vec{J}$$$ are both
reduced compared to the original folded case (refer to Fig.2(c) and Fig.3(c)).Conclusions
For
asymmetric x head gradient coil design, the current turns on the connection region of the
patient-end folded coil are not only close to the FOV but also very close to
patient body. They are efficient to
contribute to gradient strength meanwhile can induce higher electric field and
eddy current in the closeby region of human body. A design with reduced turn number in connection region confirms the observation. This information can be used to guide the high performance head gradient coil design for neuroimaging applications. Future work using a neuro
stimulation simulation model will include an analysis of each design in terms
of whether the regions with larger $$$|\vec{E}|$$$ or $$$|\vec{J}|$$$ are more
likely to induce PNS [5].Acknowledgements
This work was funded in part by CDMRP W81XWH-16-2-0054.References
[1] Foo TKF, Laskaris E, Verilyea M, Xu M, Thompson P, Conte G, Van Epps C, Immer C, Lee SK, Tan ET, Graziani D, Mathieu JB, Hardy CJ, Schenck JF, Fiveland E, Stautner W, Ricci J, Piel J, Park K, Hua Y, Bai Y, Kagan A, Stanley D, Weavers PT, Gray E, Shu Y, Frick MA, Campeau N, Trzasko J, Huston J, Bernstein MA. Lightweight, Compact, and High Performance 3T MR System for Imaging the Brain and Extremities, Magn Reson Med. 2018; 80: 2232–2245
[2] Tan ET, Hua Y, Fiveland E, Vermilyea M, Piel J, Park K, Ho V, Foo TKF, Peripheral Nerve Stimulation Limits of a High Gradient Amplitude and Slew Rate Head-Gradient. Magn Reson Med, 2020;83:352–366
[3]Tang F, Liu F, Freschi F, Li Y, Repetto M, Giaccone L, Wang Y and Crozier S, An improved asymmetric gradient coil design for high-resolution MRI head imaging, Phys. Med. Biol. 2016; 61(24):8875-8889
[4] Poole M, Bowtell R, Novel Gradient Coils Designed Using a Boundary Element Method, Concepts in Magnetic Resonance Part B, 2007;31B(3)162-175
[5] Davids M, Guérin B, Vom Endt A, Schad LR, Wald LL, Prediction of peripheral nerve stimulation thresholds of MRI gradient coils using coupled electromagnetic and neurodynamic simulations. Magn Reson Med. 2019; 81(1):686-701