Bruno Pinho Meneses1,2 and Alexis Amadon1
1Neurospin/CEA-Saclay, Gif-sur-Yvette, France, 2Université Paris-Saclay, Saclay, France
Synopsis
Design
of optimized shimming equipment is under way for UHF MRI of the human brain.
Here shimming theoretical limits are explored. With a simple model, proof is
given that, depending on field source distribution around a zone of interest,
no shimming hardware external to such a zone can fully mitigate the
inhomogeneous field. A simulation is performed on a 100-subject database to
establish hard shim limits on the whole brain. On the other hand, 3D region-specific
shimming is shown to be a very effective way to improve homogeneity in critical
zones such as the pre-frontal cortex and around ear canals.
Introduction
Static
field shimming of the human brain is essential for allowing artifact-less
high-resolution 3D-acquisitions at Ultra-High Field. However, $$$\Delta{B_0}$$$ inhomogeneity becomes stronger as the main
magnetic field $$$B_0$$$ increases. The mitigation of $$$B_z{\equiv}B_0$$$ tends to be more challenging at interfaces
with air cavities in the head, which generate a susceptibility gradient in the brain.
Here theoretical
limits to brain shimming are explored.
Although increasing orders of Spherical
Harmonics (SH) are proposed to zero-out this inhomogeneity, we show that, despite $$$\nabla^2B_z(\textbf{x})=0$$$ in the brain, SH-based fields have limited
performance.
The solution to
Laplace’s equation is1: $$B_z(r,\theta,\phi) = \sum_{n=0}^{+\infty}\sum_{m=-n}^{n}{C_{n}^{m}\mathcal{R}_{n}^{m}(r,\theta,\phi)+D_{n}^{m}\mathcal{I}_{n}^{m}(r,\theta,\phi)}$$ with $$\mathcal{R}_{n}^{m}(r,\theta,\phi)=r^{n}Y_{n}^{m}(\theta,\phi)$$ $$\mathcal{I}_{n}^{m}(r,\theta,\phi)=\frac{1}{r^{n+1}}Y_{n}^{m}(\theta,\phi)$$ and $$$Y_{n}^{m}(\theta,\phi)$$$ a spherical harmonic of order $$$n$$$ and degree $$$m$$$. Any field shimming apparatus placed outside the region of
interest (ROI), whatever its design, can only produce fields that are described
by Regular Solid Harmonics (RSH) $$$\mathcal{R}_{n}^{m}$$$. However, depending on the location of the
sources of inhomogeneity relative to the anatomy being shimmed, both RSH and
Irregular Solid Harmonics (ISH) $$$\mathcal{I}_{n}^{m}$$$ will describe the inhomogeneous field. ISH field components cannot be
zeroed-out by RSH fields produced by the external shimming system, limiting the
shimming capabilities of any system.
By localizing
field sources that cause inhomogeneity in the brain field, we verify the
impossibility of a perfect magnetic field homogenization by external shim
hardware. Then, the best homogeneity achievable by any such system when
performing global and slice-by-slice shimming is evaluated. Finally, we explore
the benefits of performing localized shimming in 3D brain ROIs.Methods
Using
a 3D model of the human head, with magnetic susceptibilities of air and tissues
set to $$$0,36.10^{-6}$$$ and $$$-9,03.10^{-6}$$$,
respectively,
the magnetic field source distribution around it is computed from the right side
of the following equation2(adapted):
$$\nabla^2(B_z(\textbf{x}))=\left(\nabla^2\chi(\textbf{x})-3\frac{\partial^2\chi(\textbf{x})}{\partial^2{z}}\right)\frac{B_0}{3}$$ where $$$\chi$$$ is the magnetic susceptibility of the medium. Voxels
with non-zero Laplacian are considered magnetic sources.
If field sources such as dipoles are located
within the smallest sphere enclosing the shimming target, perfect
homogenization using RSH is impossible. This can be illustrated by the
configuration in Fig.1. The dipolar field source of inhomogeneity $$$\textbf{m}=m\hat{\textbf{z}}$$$, located at $$$(0,0,z')$$$, generates a magnetic field $$B_z(r,\theta)=\frac{\mu_0{m}}{2\pi}\left[r^2\frac{3\cos^2\theta-1}{2}-2rz'\cos\theta+z'^2\right]\frac{1}{\left(r^2-2rz'\cos\theta+z'^2\right)^{5/2}}$$ of Solid
Harmonics expansion: $$B_z(r,\theta,\phi)=\frac{\mu_{0}m}{2\pi|z'|^3}\sum_{n=0}^{+\infty}\frac{(n+1)(n+2)}{2z'^n}\mathcal{R}_{n}^{0}(r,\theta,\phi)\;\text{for}\;\left|\frac{r}{z'}\right|<1\;(r\in\mathcal{V}_{<})$$ $$B_z(r,\theta,\phi)=\frac{\mu_{0}m}{2\pi{}z'^3}\sum_{n=2}^{+\infty}\frac{n(n-1)z'^{n+1}}{2}\mathcal{I}_{n}^{0}(r,\theta,\phi)\;\text{for}\;\left|\frac{r}{z'}\right|>1\;(r\in\mathcal{V}_{>}).$$ Thus, although $$$\nabla^2B_z=0$$$ in $$$\mathcal{V}_{<}\cup\mathcal{V}_{>}$$$, inhomogeneity can never be zeroed-out by RSH
field patterns for $$$r>|z'|$$$, i.e. as long as the dipole is in the sphere
encompassing the ROI.
A database
of 100 three-dimensional $$$\Delta{}B_{0}$$$ brain maps was acquired after 2nd-order
shimming at a 3T Siemens Magnetom Prisma imager with 1.7-mm isotropic
resolution, then scaled up by 7/3 for investigation at 7T. FSL’s brain
extraction tool was used to ignore non-brain voxels. For each map in the
database, both global and slice-by-slice RSH shimming are simulated from orders
3 to 14. Inhomogeneity is defined as the standard deviation of the $$$B_0$$$ field in
the ROI while shimming performance is evaluated by the percentage reduction in
inhomogeneity from the scanner 2nd-order baseline.
Finally,
for a single subject, a mask is created, consisting of three spheres, one of
6-cm diameter in the pre-frontal lobe, two of 5-cm in the temporal lobes,
covering susceptibility-critical regions while avoiding intersection with
sources. Then full 11th-order RSH shimming is simulated in this mask.
Results and Discussion
Figure
2 illustrates that magnetization-induced sources of inhomogeneity are disposed
around the brain in such a way that it is impossible to obtain any brain-enclosing
sphere that does not enclose sources. Therefore, RSH cannot zero-out the
magnetic field excursion inside the brain even though $$$\Delta{}B_z^{\text{brain}}$$$
The
impossibility of perfect shimming arising from the exposed configuration is
reflected in the simulated shimming results in Figure 3. A plateau in
inhomogeneity reduction is observed starting at order 11. It was checked that
this result does not depend on image resolution by simulating shimming from
downgraded $$$B_0$$$-maps with 3-mm voxels.
The
plateau defines supremum inhomogeneity reduction, relative to baseline, of 43.4%(6.5)
and 72.5%(5.8) for whole-brain and slice-by-slice shimming, respectively.
At
maximum performance in whole-brain shimming, there are still wide regions with high-field
excursion (over 100 Hz), notably close to the ear canals and around nasal
conchae (cf. Figure 5).
While with
slice-by-slice shimming, global inhomogeneity can ideally be brought to
inferior levels than with whole-brain shimming, if high-resolution is
envisioned, 2D-acquisitions can be a problem due to the lower signal emitted by
each slice.
On the
other hand, by targeting the most critical shimming regions with spherical ROIs,
inhomogeneity decreases from 88.8 Hz to only 9.6 Hz, with a drop of the number
of voxels over 100 Hz from 17.3% to a remarkable 0.1% (cf. Fig. 5).
Inhomogeneity outside the target region is deeply compromised, but if only that
shimming target is of interest for imaging, tailored inner volume selection could
be performed to avoid aliasing artefacts.Conclusion and perspectives
Due
to intrinsic characteristics of the human head anatomy, static field
inhomogeneity inside the human brain can never be fully mitigated by
whole-brain shimming whatever the external hardware used. Nevertheless, by
focusing shimming in brain ROIs, there is a potential for great improvement
especially in regions close to the ear canals. Although the design of such
focused shimming systems was not addressed here, we expect recent methodologies
for optimal $$$B_0$$$ shimming Multi-Coil Array design3,4 to be pertinent.Acknowledgements
We would like to thank Guy Aubert and Nicolas Boulant for the insightful discussions leading to this work.References
1. J. D. Jackson, Classical
electrodynamics. New York: Wiley, 1962.
2. R. Salomir, B. D. de Senneville, and C. T. Moonen. A fast calculation method for magnetic field inhomogeneity due to an arbitrary distribution of bulk susceptibility. Concepts Magn. Reson. 2003; 19(1):26-34.
3. B. Pinho Meneses, M. Luong,
and A. Amadon. Optimized multi-coil array design for human brain shimming at
Ultra-High Field. Proc. ISMRM 2019; p.1477.
4. B. Pinho Meneses and A.
Amadon. Static-magnetic-field shimming coil system for magnetic resonance imaging.
European Patent 072081EP/EPR 2019.