Lieke van den Wildenberg1, Quincy van Houtum1, Wybe van der Kemp1, Catalina Arteaga de Castro1, Alex Bhogal1, Paul Chang2, Sahar Nassirpour2, and Dennis Klomp1
1Radiology Department, UMC Utrecht, Utrecht, Netherlands, 2MR Shim GmbH, Reutlingen, Germany
Synopsis
Ultra-high-field MRI is more sensitive to variations in the B0
field specially for larger organs such as the liver. Dynamic shimming using a
local array of shim coils can potentially improve the field homogeneity by 18-46% when compared
to the standard hardware correction methods at 7T. The use of local shim coil
arrays embedded in RF receiver coils may become a good strategy when maximizing
shim performance considering the limited available number of shim channels,
particularly for larger organs such as the liver at ultra- higher field
strengths.
Introduction
Ultra-high field
MRI (≥7T)
has the potential to become a clinically viable tool due to its higher
intrinsic SNR and CNR compared to lower magnetic field. This
increase in SNR and CNR allows for higher spatial resolutions and higher
sensitivity in image acquisition than conventional clinical MRI. However, one
of the main technical challenges with increasing field strength is the static
magnetic field inhomogeneity. Variations in the B0 field lead to
local frequency offsets that can cause artefacts during MRI and MRS measurements.1 The magnetic susceptibility differences between
tissue interfaces is one of the leading factors for local magnetic field
inhomogeneities due to the linear increase in field offsets caused by tissue
susceptibility at higher fields. To reduce the spatial variations, higher order
spherical harmonics are needed in addition to the standard hardware B0
shimming methods of the scanner, which would require a high number of shim
coils and channels.2 An alternative is the use
of local shim coil arrays that can provide the high order of spatial field
variation albeit confined to a small region. Moreover, local shim coils would
couple less to the conductive bore thereby may have negligible eddy currents
thus could generate spatially varying magnetic fields that counteract the B0
field variations in real time. Here, we demonstrate that B0 field variations
in the liver can be substantially reduced using a local array of only 16 shim
coils, when compared to the standard hardware correction method at 7T.Methods
A whole-body 7T MR scanner equipped with a multi-transmit RF system
(Achieva, Philips Health Care, Cleveland, OH, USA) was used to acquire B0
maps of the liver in five healthy volunteers. Eight transceiver fractionated
dipole antennas with 16 additional receive loops interfaced to 8 parallel 2kW
peak power amplifiers, were positioned symmetrically around the body at the
position of the liver.3 RF phase shimming was
performed to maximize transmission efficiency and optimize B1+
field homogeneity in the liver. Five dual-echo B0
maps (GE, 386×410×180mm3 FOV, 6×6×6mm3 voxel size, FA=4°,
TR=6ms, TE=1.413ms, ΔTE=1ms) were acquired during free breathing as well as in inspiratory
and expiratory breath-holds. The local shim coil array consisted of 16 circular
loops, each with a diameter of 5cm. The coil diameter was
optimized using simulation libraries provided by MR Shim GmbH (Reutlingen,
Germany) between the range of 1-10cm. Each loop consisted of 20 coil windings
of enameled copper wire with a diameter of 0.56mm.4 After the diameter was optimized,
electromagnetic simulations were performed using 30 slices of the
gradient dual-echo field maps to calculate the magnetic field that is generated
by a constant electric current through the shim coil array using the
Biot-Savart Law.5 The coils were arranged in
two rows of eight around the body (Figure 1).
These simulations were validated afterwards with a measurement in the 7T MRI
scanner. B0 maps with equal sequence parameters were acquired in a
phantom (octagon, 39×40×20cm, filled with polyvinylpyrrolidone) using a single
shim coil driven with a current of 8A. The difference between the simulation
and the measured B0 map was calculated as the RMSE and the
correlation with the Pearson correlation coefficient.6Results
The simulations showed that the use of the local array of shim coils led
to 65±8.3% improvement of the B0 homogeneity during free breathing,
80±4.5% improvement during inspiratory state, and 75±12% improvement during
expiratory state when compared to unshimmed B0 map acquired during
free breathing (Figure 2, 3). Compared to
the field homogeneity when using the standard hardware-based method up to
second order terms, the B0 variation in the liver can be reduced by 9±5.1% in free breathing, 46%±16 in inspiratory state, and
18±12% in expiratory state. Comparison of the simulations with the phantom
7T measurements, shows
clear B0 differences (Figure 4, 5). The difference is close to zero further away from the shim coil. The
Pearson correlation coefficient between simulations and measurements is 0.70
and the RMSE is 196Hz with the maximum B0 field variation value
measured at 5.34×103Hz and the maximum of the simulation at 1.66×103Hz.Discussion
Substantial gain in field uniformity can be obtained with organ specific
local B0 shim coils when compared to generalized static spherical
harmonic shimming. In our study, we simulated dynamic shimming by using breath-hold,
while real-time shim updating was not implemented. Direct use of simulated
fields in real time will cause a mismatch as shown by the phantom measurements that
differed from the simulations. While this can be prevented by obtaining fast B0
calibration maps from all shim coils, the magnitude of field variation is
similar, thereby the simulation results do provide validation for the power
requirements of the coils and amplifiers.Conclusion
This study shows that local B0 field variations in the liver
at 7T can be reduced by using an array of local shim coils. B0
shimming improves by 9±5.1% in free breathing up to 46% in inspiration when
using a local shim coil array in comparison to standard hardware shimming B0
shimming. As local shimming becomes
more important at ultra-high magnetic field strengths, the use of embedding local
shim coils in RF receiver arrays may become attractive, particularly for larger
organs such as the liver.Acknowledgements
No acknowledgement found.References
1. Truong Trong-Kha, et al. Effects of static and
radiofrequency magnetic field inhomogeneity in ultra-high field magnetic
resonance imaging. Magnetic Resonance Imaging. 2006; 24:103-112.
2. Pan JW, Lo KM, Hetherington HP. Role of high order and
degree B0 shimming for spectroscopic imaging of the human brain at 7 Tesla. Magn Reson Med 2012;
68:1007-1017.
3. Raaijmakers AJE, Italiaander M, Voogt IJ, Luijten PR, Hoogduin JM, Klomp
DWJ, van den Berg CAT (2015) The fractionated dipole antenna: a new antenna for
body imaging at 7 Tesla. Magn Reson Med 75:1366–1374.
4. MR Shim GmbH, Reutelingen, Germany
5. Juchem C, Nixon TW, McIntyre S, Boer VO, Rothman DL, de Graaf RA. Dynamic
multi-coil shimming of the human brain at 7 T. Journal of Magnetic Resonance
2011; 212(2):280-288.
6. Fisher, R.A. Statistical Methods for Research Workers, 13th Ed.,
Hafner, 1958.