Alessandro Arduino1, Oriano Bottauscio1, Mario Chiampi1, and Luca Zilberti1
1INRiM, Torino, Italy
Synopsis
The temperature increase induced by MRI gradient fields switched according to EPI sequences in a metallic hip implant is evaluated with an accurate numerical model. The role of the body position within the scanner, of the frequency encoding direction and of the thermoregulation is studied and discussed. The results show that the eddy currents generated by the gradient coils in the implant are sufficient to induce a temperature increase of some kelvin (without accounting for the RF power deposition); moreover, it results that the biological effect can be controlled by choosing properly the frequency encoding direction.
Introduction
The
increasing number of patients with metallic orthopaedic implants imposes safety
issues associated with their MRI examination, because of the strong interaction
between the time-varying electromagnetic fields (EMF) and the high conductivity
components.
While the direct
heating of tissues correlated with the exposure to radiofrequency fields is
largely analysed in literature, less attention is devoted to the thermal
effects due to the switching Gradient Coil (GC) fields. However, experimental1,2,
theoretical3 and computational4,5 results show possible
significant temperature increase in the tissues due to the diffusion of the
heat produced inside the implant by eddy currents.
Here we
use a computational code, specifically developed, to evaluate the heating
generated in a patient with a unilateral hip prosthesis when undergoing an
Echo-Planar Imaging (EPI) sequence and analyse the influence of some parameters
(body position, frequency encoding direction, thermoregulation) on the thermal stress.
The results reported in this digest represent a selection of data recently
published by the Authors7.Method
The limited
temperature increase (few kelvin) allows us to handle separately the
electromagnetic and the thermal problems. The electromagnetic problem is solved
only inside the implant, discretized into voxels, assuming that the currents
induced into the body tissues neither modify the GC magnetic field, nor generate
significant thermal power. The signals of the GC switching sequence are divided
into sub-signals. Each sub-signal is expanded in a truncated Fourier series and
the related EMF solutions are computed in the frequency domain through a hybrid
Finite Element/Boundary Element method. The electric fields in each implant voxel are moved in
time domain and superimposed, sub-signal by sub-signal and coil by coil, to
reconstruct the instantaneous evolution of the electric field and the Joule
power density. The strategy of signal subdivision significantly reduces the
computational burden (a reduction around 90 % is estimated for the EPI
sequence).
The thermal problem is modelled through Pennes’
equation inside the human body using the Joule power densities
previously computed as forcing terms. The voxelized human
model allows the application of the finite difference method with the
Douglas–Gunn time split implemented on GPUs.6 The whole
computational procedure has been validated by comparison with experiments on an
acetabular cup inside a phantom2,7.Results
A unilateral right implant with acetabular shell
and liner is inserted in the “Duke” anatomical model of the Virtual Population,
involving 77 tissues, whose electric and thermal properties are included in the
IT’IS Foundation database. The metallic components are made of CoCrMo alloy, Ti6Al4V
alloy or austenitic stainless steel. The human model is placed inside a tubular
MRI scanner, whose GC generates the magnetic flux density spatial distributions
of Fig. 1. The EPI sequence, shown in Fig. 2, is executed continuously for
about 12 minutes.
The computations are performed moving the body along
the scanner axis, varying the coordinate zf of the femoral head with
respect to the MR isocenter in the range from -450 mm to +300 mm. The x and y
coordinates of the implant are maintained in the “natural” position. The results,
summarized in Fig. 3 in terms of temperature increase with respect to the local
temperature before the exposure, show that, for the CoCrMo alloy implant, the maximum
and minimum heating are reached for |zf|=300 mm and zf=0,
respectively.
The change of the imaging plane, and the
consequent role of the three GCs, affects the thermal stress sensibly. Figure 4
compares the maximum temperature elevation reached for all possible configurations
and implant materials with the femoral head at zf=-300 mm. The
values indicate that the six configurations could be grouped in three couples
having the same frequency encoding direction and the same maximum temperature
increase. These results highlight the essential role of the direction of the
frequency encoding whose signals produce most of the power losses. Moreover, it
can be appreciated how the temperature increase is proportional to the electric
conductivity of the implant.
The effect
of the thermoregulation is found to be very low (the variation of the maximum
temperature is always less than 0.15 K), even assuming an instantaneous
biological reaction.Discussion
The
effects of GCs are usually stronger when the implant is placed in the scanner
periphery, where the Bz component is higher and significant concomitant components
Bx (or By) are generated by X and Y GCs.
For the EPI
sequence, the frequency encoding direction determines the amount of the
heating. Hence, if the Y GC plays this role, the temperature elevation
significantly reduces, because of the relatively low field produced by such a
coil throughout the volume taken up by the hip implant.
The
thermoregulation affects the results only slightly, due to both the limited
temperature increase and the relatively low basal blood perfusion coefficient
of the tissues surrounding the prosthesis.Conclusion
The computed results put in evidence that the GC
heating effects must be taken into account for a reliable evaluation of the
safety conditions in a patient with metallic hip prosthesis when undergoing MRI.
The same investigations also suggest safety measures which could reduce the
thermal stress (e.g., change of the imaging plane, use of 'less aggressive'
sequences or introduction of waiting times).Acknowledgements
The results here presented have been developed in the framework of the
17IND01 MIMAS Project. This project has received funding from the EMPIR
Programme, co-financed by the Participating States and from the European
Union’s Horizon 2020 Research and Innovation Programme.References
-
Graf H, Steidle G, Schick F Heating of
Metallic Implants and Instruments Induced by Gradient Switching in a 1.5-Tesla
Whole-Body Unit. J Magn Reson Im. 2007;26:1328–1333.
- Bruehl R, Ihlenfeld A, Ittermann B. Gradient
heating of bulk metallic implants can be a safety concern in MRI. Magn. Reson.
Med. 2017; 77:1739-1740.
- Zilberti L, Arduino A, Bottauscio O, et al. The
Underestimated Role of Gradient Coils in MRI Safety. Magn Reson Med. 2017,
77:13-15.
- El Bannan K, Handler W, Chronik B, et al.
Heating of Metallic Rods Induced by Time-Varying Gradient Fields in MRI. J Magn
Reson Im. 2013, 38:411-416.
- Zilberti L, Bottauscio O, Chiampi M, et al. Numerical
Prediction of Temperature Elevation Induced around Metallic Hip Prostheses by
Traditional, Split, and Uniplanar Gradient Coils. Magn Reson in Med. 2015; 74:272–279.
- Arduino A, Bottauscio O, Chiampi M, et al. Douglas-Gunn
Method Applied to Dosimetric Assessment in Magnetic Resonance Imaging. IEEE
Trans Magn 2017; 53:5000204.
- Arduino A, Bottauscio O, Bruehl R, et al. In silico evaluation of
the thermal stress induced by MRI switched gradient fields in patients with
metallic hip implant. Phys Med Biol, in press. DOI: 10.1088/1361-6560/AB5428.