Arjama Halder1,2, Ali Attaran2, William B Handler2, and Blaine A Chronik2
1Medical Biophysics, Western University, London, ON, Canada, 2Physics and Astronomy, Western University, London, ON, Canada
Synopsis
Validating measured gradient induced electric fields in a conductive ASTM phantom within an X-Axis gradient coil.
INTRODUCTION
The
objective of this study was to investigate gradient coil induced electric
fields using an ultra-low frequency electric field probe within a conductive ASTM
phantom in an operating gradient coil. In recent years there has been a growing population of
people equipped with active
implantable medical devices (AIMDs) who might require an MRI exam at
some point in the future. However, the electromagnetic interactions
between devices and the various sub systems of MR scanners
lead to potential hazards that can cause patient to have limited,
to no access to an MRI1. The ISO/TS 10974:20182 mentions
one such potential cause of worry,
to be resulting from the electromagnetic interaction between the
MR scanner pulsed gradient fields and AIMDs. Electric field quantification is one of the most difficult aspect of this interaction and can cause patient harm of
varying severity, depending on the AIMD type.
Most of the studies performed to address device
compatibility with the MR scanner that focuses on this issue
concentrates on designing different test
platforms that mimics the dB/dt fields produced by
gradients2,3 or developing test methods that allow electric field
measurements without the need of generating dB/dt fields4. Here we
propose the use of an ultra-low frequency electric field probe5,6,
as a supplementary test method to measure the actual variation of the electric
field caused by an operating gradient coil in a scanner. The use of this
probe is validated using a saline filled ASTM phantom as a
representation of a tissue-mimicking environment and a
high performance asymmetric
head gradient coil outside of the scanner. Performing this test
with sufficient accuracy in an operating gradient coil is difficult given the
strength of the field generated, but
can be very useful for direct field measurements in the scanner for
device testing purposes, and also validates electric
fields predicted
using the
low frequency solver in Sim4Life (Zurich Med
Tech).METHOD
To perform these measurements the ASTM phantom was
filled with saline up to the height of 9 cm (conductivity of saline = 0.47 S/m)
and placed within a gradient coil as shown in figure 1. For this set of
measurement, only the X-Axis of the gradient
coil was used, with the a sinusoidal current
of 126 A at
a frequency of 1 kHz. This operation frequency allowed the use of
quasi-static approximations7 which led us
to assume that tissues are uniform volume conductors. The ultra-low frequency electric field probe6 was held and moved by a
robotic arm which allowed precise measurements at locations near the head
region of the phantom. The probe was constructed using a 2 cm short dipole
antenna followed by a 2-stage instrumentation amplifier and a differential
amplifier that provided differential amplification and removed common-mode
voltage respectively. This probe had a differential input impedance of 0.8||2
(GΩ||pF), while the designed amplifier had a
gain of 100 and a cut-off frequency of 31 kHz. Data
was collected via a connection
to an oscilloscope with a termination impedance of 1 MΩ
using an RG58 cable. To remove the extrinsic influences
in the measurements the saline solution was grounded.
Post-processing of the collected data involved
removing high frequency noise from the measurements using a low-pass filter for
every position. A sinusoidal fit was performed on
the filtered signals to obtain the peak values at those positions.
To quantify the error of the fit a root mean squared error (RMSE) was
calculated for each fit performed. The fitted signal was then scaled with the
calibration factor associated with the probe, which
is 2.64 (V/m)/V. Simulations were performed in Sim4Life using the
magneto-quasi-static low frequency solver to compare
against the experimental results.RESULTS
The experimental setup along with a schematic is shown
in figure 1. Figure 2 shows the simulated electric field magnitude on the xz,
yz and xy plane at the center of the phantom. Figure 3 show the measured and
simulated results along a line y = 0 on 2 different yz plane that are about 3
cm off in either direction from the center of the phantom with the probe oriented
in the z direction. DISCUSSION
The experimental results show a reasonable agreement with the
simulated behavior of the electric field given the sensitivity of the field
measured, as shown in figure 3. This agreement
shows that this prototype electric
field probe makes it possible to obtain useful measurements of
electric fields induced by gradients inside tissue. Future measurements will
build upon this to explore
the
full extent of the phantom and the other gradient axes. Measurements will also be performed with an AIMD
within the phantom.
The results achieved from the this study
show possible
commercial use of such a probe for conducting gradient induced electric field
measurements in AIMDs within a scanner or a test platform. CONCLUSIONS
ISO/TS 10974:20182 considers gradient
induced extrinsic electric fields to a potential cause of worry for patients
with AIMDs. The standard proposes radiated gradient immunity test to account
for these effects. This research provides a supplementary test method that
allows direct measurement of the variation in electric field using an ultra-low
frequency electric field probe within a gradient coil.Acknowledgements
The authors would like to thank the research and
financial supports received from Natural Sciences and Engineering Research
Council (NSERC) of Canada, the Ontario Research Fund (ORF), and CMC
Microsystems.References
- Gimbel
JR, Kanal E. Can patients with implanted pacemakers safely undergo magnetic
resonance imaging? J. Amer. College Cardiol.2004;43(7):1325–1327.
- International Organization for Standarization. Assessment of
the safety of magnetic resonance imaging for patients with an active
implantable medical device. ISO/TS
10974. 2018;4.
- Martire
DJ, Handler WB, Chronik BA. Design of an MRI Gradient Field Exposure System for
Medical Device Testing. ISMRM. 2017;
Abstract # 4334.
- Xin
X, Chen XL, Sison S. A MRI Gradient Induced Electric Field Exposure System for
Active Implanted Medical Devices. ISMRM. 2019; Abstract # 1440.
-
Glover PM, Bowtell R. Measurement of Electric Fields Due to
Time-Varying Magnetic Field Gradients Using Dipole Probes. Physics in Medicine
and Biology. 2007;52: 5119-5130.
-
Halder
Arjama Attaran A, Handler WB, Chronik BA. Electric Field Probe used for
Gradient Coil Induced Field Measurements During Medical Device Testing: Design,
Calibration and Validation. IEEE Transactions on
Antennas and Propagation. 2019; ID # AP1907-1282.R1. (Under review)
- Roth BJ, Cohen LG, Hallett M. The Electric Field Induced
During Magnetic Stimulation. Magn Mot
Stimul Basic Princ Clin Exp. 1991;43:268-278.