Arjama Halder1,2, Ali Attaran2, William B Handler2, and Blaine A Chronik2
1Medical Biophysics, Western University, London, ON, Canada, 2Physics and Astronomy, Western University, London, ON, Canada
Synopsis
A validation study to ensure the EM behavior predicted by the quasi-static solvers in SEMCAD and Sim4Life is accurate.
INTRODUCTION
SEMCAD and Sim4Life are commercially used for device
testing purposes, however, there are no validation studies performed that
suggest the behavior predicted by these solvers are correct or consistent
across different versions. The electromagnetic interaction between active
implantable medical devices (AIMDs) and gradient coils, something that can be
studied using these solvers, is a cause of concern as suggested in the ISO/TS
10974:20181,2, which makes it necessary to validate these solvers. Such
a validation will ensure that the behavior predicted by the solvers are
accurate, allowing device testing companies to correctly identify the class of devices
that require physical testing, which will effectively reduce operating expense
and test time.
Here we perform a validation study for SEMCAD and
Sim4Life, by simulating a conductive cylindrical phantom within a test coil
that is built to produce homogeneous changing fields similar in size to those
of gradients and operate in the kHz range. These simulations were validated
with measurements of electric and magnetic fields within the coil. METHOD
For the purpose of this
abstract a cylindrical phantom made up of acrylic material with a thickness of
1.33 cm, inner diameter of 22.8 cm and length of 30.9 cm was filled with saline (conductivity of saline = 0.47 S/m) up to
the height of 10 cm and placed within a test coil as shown in figure 1. The
test coil used is a split solenoid (the split was designed for increased
homogeneity) that create a switched-field (dB/dt) exposure system3.The benefit of such an external test platform is it avoids the operating
expense and difficulties of testing in an MR scanner. The current applied to
this exposure coil was sinusoidal in nature with a peak value of 80 A.
Simulation studies were performed using low frequency magneto quasi-static
solvers available in Sim4Life and SEMCAD. The vector potential generated within
the dB/dt exposure coil was calculated using the Biot-Savart law in the
quasi-static regime4. The solver used the curl of the vector
potential to calculate the magnetic field and computed the electric field using
tricubic interpolation. This electric field was perturbed due to the presence
of a conducting phantom within the coil, as will be the case in the presence of
any AIMD. All boundary conditions were neglected as zero Neumann boundary
conditions, i.e., vanishing normal flux. The tolerance for convergence had a
relative value of 10-8.
This simulation study
was validated with measurements performed using an ultra-low frequency electric
field and a dB/dt probe. The electric field probe5 was constructed
using a short dipole connected to a 2-stage instrumentation amplification phase
that provide differential amplification followed by a difference amplifier
stage that removes the common-mode voltage. The 3D dB/dt probe which led to the
calculation of the magnetic field is composed of three sets of perpendicular
loops around a common center. The calibration factor associated with the electric
and dB/dt probe are 2.64 (V/m)/V and 50 (T/s)/V respectively. A robot equipped
with 3 Nema 23 stepper motors and driven with a controller was used to
accurately place the probes within the volume of the phantom. Data from the
probes were collected using an oscilloscope as the recording device, with a
termination impedance of 1 MΩ, to determine the peak electric field and dB/dt
at each location inside the phantom. To remove extrinsic influences on voltage
changes observed in the measurements, the saline solution was grounded. RESULTS
The experimental setup along with the schematic is
shown in figure 1. Figure 2 show the simulated electric and magnetic field on
the xy plane at the center of the coil as predicted by Sim4Life and SEMCAD.
Figure 3 compares the measured with the simulated results of the electric and
magnetic fields along the line y = 0 on the xy plane at the center of the
coil. Figure 4 shows the average percent difference error between a volume of measured and simulated data for both electric and magnetic field.DISCUSSION
Fig. 2a suggests a
circulating electric field with peak values near the edges of the phantom and a
drop off at the center as expected. Fig. 2b shows a largely homogeneous magnetic
field along the test coil axis, with peak values within the phantom while a
drop off is seen near the edges. Figure 3 suggests there is a strong agreement
between the measurements performed and the simulated behavior for both
electric and magnetic fields within the coil. The average percentage error for the electric and
magnetic field experienced between simulation platforms and measurements is
less than 10%. This allows us to conclude that these solvers can correctly map
the electromagnetic environment within a gradient coil, thus providing a method
to study the interactions between gradient coils and AIMDs.
CONCLUSIONS
The electromagnetic interaction between gradient
coils and AIMDs is an important consideration factor for device manufacturing
companies as suggested by the ISO/TS 10974:2018. This abstract aims to provide
a validation study which suggest that these low-frequency quasi-static solvers
available in Sim4Life and SEMCAD allow us to accurately simulate these
electromagnetic interactions. Acknowledgements
The authors would like to thank the research and
financial supports received from Natural Sciences and Engineering Research
Council (NSERC) of Canada, the Ontario Research Fund (ORF), and CMC
Microsystems.References
- Gimbel
JR, Kanal E. Can patients with implanted pacemakers safely undergo magnetic
resonance imaging? J. Amer. College Cardiol.2004;43(7):1325–1327.
- International Organization for Standarization. Assessment of
the safety of magnetic resonance imaging for patients with an active
implantable medical device. ISO/TS
10974. 2018;4.
- Martire
DJ, Handler WB, Chronik BA. Design of an MRI Gradient Field Exposure System for
Medical Device Testing. ISMRM. 2017;
Abstract # 4334.
- Roth BJ, Cohen LG, Hallett M. The Electric Field Induced
During Magnetic Stimulation. Magn Mot
Stimul Basic Princ Clin Exp. 1991;43:268-278.
- Halder
Arjama, Attaran A, Handler WB, Chronik BA. Electric Field Probe used for
Gradient Coil Induced Field Measurements During Medical Device Testing: Design,
Calibration and Validation. IEEE Transactions on
Antennas and Propagation. 2019; ID # AP1907-1282.R1. (Under review)