Thomas Lottner1, Simon Reiss1, Lukas Vincent Haas1, Johannes Erhardt2, Ali Caglar Özen1,3, Lars Bielak1, Martin Schüttler4, and Michael Bock1
1Dept. of Radiology, Medical Physics, Medical Center – University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Germany, 2Department of Microsystems Engineering—IMTEK, University of Freiburg, Freiburg, Germany, 3German Consortium for Translational Cancer Research Partner Site Freiburg, German Cancer Research Center (DKFZ), Heidelberg, Germany, 4CorTec GmbH, Freiburg, Germany
Synopsis
Active
implantable medical devices (AIMD) can restore body function in patients with
neurological or other disorders which increases their quality of life
considerably. In this work we measure RF-induced heating by
MRI thermometry in combination with FOTP measurements and simulations and apply
it to a 64-channel electrode grid. The
combination of MR thermometry with high spatial resolution, FOTPs with high thermal
resolution and supplementary simulations of electromagnetic and thermal
processes shows how limitations of the individual methods are overcome and
result in a more comprehensive understanding of the thermal dynamics resulting
from the exposure of implants to RF-fields.
Introduction
Active
implantable medical devices (AIMD) can restore body function in patients with
neurological or other disorders which increases their quality of life
considerably. Unfortunately, patients with AMIDs are exposed to a significant
safety hazard during an MRI exam, as the radio-frequency (RF) fields can couple
resonantly to the metallic structures of the AIMD resulting in tissue heating.
To ensure patient safety, AIMDs are tested with temperature measurements using fiber-optic
temperature probes (FOTP)1. This method is tedious for
high-channel-count implants (e.g., 64-channel intracranial (ic) EEG electrode
grids), since FOTP systems only have 4-8 sensors. To overcome this, the most
relevant electrodes may be identified by simulating the scattered fields that
are associated with tissue heating according to ISO‑TS‑109742.
It has
recently been shown, that MR thermometry is possible in the vicinity of planar
electrodes, so that heating information can be directly extracted from a
rapidly acquired temperature map. In this work we measure RF-induced heating by
MRI thermometry in combination with FOTP measurements and simulations and apply
it to a 64-channel electrode grid.Methods
A commercially
available °AirRay Cortical Electrode (CorTec GmbH,
Germany) with 8x8=64-channels made of 25µm thick Pt90/Ir10 (Figure 1),
where each electrode contact (opening diameter: 2.7mm) on the grid with
silicone rubber substrate was connected via metal tracks to a 45cm-long cable
was placed on an acrylic-glass sample holder with the leads parallel to B0 and off-center (Dx = 21cm) in a phantom (15x15x50cm) filled with
HEC gel (31g/L Hydroxyelthycellulose, 1.55g/L NaCl1). Heating was
measured in a clinical 3T MRI system (Magnetom TRIO, Siemens Healthineers, Erlangen, Germany) using a proton resonance
frequency (PRF) thermometry sequence3 with an off-resonant RF pulse
for heating4. The following parameters were used: TE/TR = 10/21 ms,
αimaging = 11°, αheating = 380°, 1.6x1.6x4 mm³
resolution, TA =5.6 s x 80 repetitions = 7.5 min, nominal whole body SAR: 4 W/kg.
The heating pulse was active for 6min followed by a 1 min cooling period. The
lower bound of the PRF imaging slice was placed 4mm above the electrodes, and temperature
change maps were calculated from the difference between 15 baseline images and
phase maps acquired during heating. Additionally, temperature was measured
using a FOTP (FOTEMP9-16 Optocon AG, Dresden, Germany) placed 2 mm above the electrode
contact with the highest temperature change that was seen in preliminary PRF thermometry measurements. Furthermore,
different immersion lengths of the implant cables (fully immersed (45cm), 41,
37, 33cm immersion) were measured.
Electro-magnetic
FDTD simulations were performed using a single electrode contact in Sim4Life
(Version 5.0, ZMT-AG, Zurich, Switzerland). The resulting SAR values were further
used to simulate thermal behavior, where the input power was scaled to minimize
the difference between FOTP/Thermometry dynamics and maximum temperature
changes for the corresponding location in the simulationResults
Figure 2
shows the maximum temperature changes over time ΔTmax, which were
located above the electrode marked in Figure 1. ΔTmax, is largest
for the immersion length of 12cm in air. Comparing ΔTmax, for
FOTP and PRF shows a R2=0.88 with a scaling factor of 6.58. Figure 3 shows an
MRI image of the electrode with an overlay indicating the temperature increase
in a slice 4mm above the electrode (immersion length 12cm). The dissipation of heat
concentrates around the electrode contacts and shows an increase towards the
left side of the implant. Figure 4 shows a slice perpendicular to the electrode
of simulated ΔTmax, with a resolution of 0.2x0.2x0.2mm3.
Indicated simulated ΔT-t curves of the FOTP and PRF show good
agreement with the measurements.Discussion
This study
shows that simultaneous heat measurement of 64 individual electrode contacts is
feasible. The visible asymmetry of the heat dissipation in Figure 3 can be
attributed to the positioning relative to the body coil. All temperature
measurements and simulations show the expected behavior. The difference in the
maximum temperature between FOTP and MR thermometry can be attributed to
differences between MR and FOTP position.
FOTP
measurements are characterized by high temporal (0.1 samples/s) and thermal
resolution (0.1K), but are limited by the number of sensors and hence spatial
resolution. Here, MR thermometry can be used to cover a whole implant with
millimeter resolution and a temperature resolution of about 1K . Although this temperature
resolution may not be sufficient for safety assessment, which requires a temperature
sensitivity well below 1K, it may help to locate hot spots at which the FOTP should
be placed.Conclusion
The
combination of MR thermometry with high spatial resolution, FOTPs with high thermal
resolution and supplementary simulations of electromagnetic and thermal
processes shows how limitations of the individual methods are overcome and
result in a more comprehensive understanding of the thermal dynamics resulting
from the exposure of implants to RF-fields.Acknowledgements
We gratefully acknowledge support for Sim4Life by ZMT, www.zurichmedtech.comReferences
1ASTM standard F 2182-2011a, 2011. ASTM 2182
Standard test method for measurement of radio frequency induced heating near
passive implants during magnetic resonance imaging. ASTM Int. 1–14
2ISO, "ISO TS 10974: Assessment of the
safety of magnetic resonance imaging for patients with an active implantable
medical device," International Organization for Standardization, Geneva,
Switzerland; 2012.
3Ehses P., et al, 2008 MRI Thermometry: Fast
Mapping of RF-Induced Heating Along Conductive Wires, MRM, 60:457–461
4Quesson B., et al, 2000. Magnetic resonance
temperature imaging for guidance of thermotherapy. JMRI 12, 525–533