Xin Chen1 and Michael Steckner1
1Canon Medical Research USA, Inc., Mayfield Village, OH, United States
Synopsis
It is widely understood that the B1 field used in MRI can
cause significant tissue heating by inducing eddy currents. Additionally, the
B1 field simultaneously induces heating in all RF coils. Patient heating is indirectly
constrained by SAR limits, as defined in IEC 60601-2-33, and the surface
contact coil temperature limit (41°C) is defined by IEC 60601-1. An
examination of the simultaneous application of both heating sources is
investigated by electromagnetic and thermal modeling. Initial results suggest
that while the presence of a “hot” coil elevates tissue temperatures, the two
simultaneous sources of heating don’t pose a patient safety risk.
INTRODUCTION
RF burns are believed to be thermal injuries due to
excessive local SAR hotspots. Thermal dose metrics like CEM431 are useful for
investigating MR exposures and assessing risk to any given tissue and their
specific damage thresholds. For cases of skin contact against hot
surfaces, standards exist2 which characterize the contact material type,
exposure time, and acceptable temperature exposure limits. 41°C is
considered a safe, no justification necessary, long-term contact exposure
temperature that is universally used by the MR vendors3,4. The limiting use
case is the non-responsive patient strapped into a coil. The juxtaposition of
both heat sources and potential consequences for patients has not been
previously considered and is the focus of this abstract.
Humans nor coils instantly heat on exposure to B1 fields. The
general heating response of RF coils is an exponential function which rises to
a steady-state temperature and the human tissue temperature rise is a more
complex function of SAR and thermoregulatory responses and other cooling
mechanisms (e.g. convection). EM/thermal modeling packages are presently designed
to only consider the standard human exposure condition and do not consider time
variable boundary conditions, such as surface contact with a coil housing. Simple
conservative approximations permit an initial investigation on whether two
simultaneous heat sources are a patient risk concern.METHODS
EM modeling (Sim4Life v4.4, ZMT) used 1.5T birdcage whole
body transmit coil (16-rung
high-pass birdcage, diameter 750mm, length 650mm. RF shield diameter 790mm, length
850mm)5,6. The unclothed human model Duke7,8 was positioned in the
coil to model pelvis or wrist imaging situation (Figure 1). The coil was tuned to
64MHz and driven in QD mode with two voltage sources. Resultant RF power
deposition into Duke was scaled to 4W/kg WB SAR and used as heating source for
the subsequent thermal simulation using Pennes Bioheat Equation. Tissue
parameters with thermo-regulated skin, fat (including subcutaneous adipose
tissue) and muscle were applied.9 Thermal simulations started with 37°C initial
tissue temperature and 25°C ambient air temperature. A two hour wait
time (no RF heating) was applied to reach basal thermal equilibrium before the
RF source was applied continuously for one hour. Thermal simulation was then
repeated with hot Rx coil surfaces introduced simultaneously with RF source. The
top of the abdomen and entire wrist, separately, were fitted with a completely
conformal Dirichlet boundary condition with a constant temperature of 40°C,
representing the coil housing (Figure 2). While an uncommon coil combination, this
reflects the locations of thermal hotspots given by the simulation without hot
coil and permitted one single modelling. In practice, coil housings have
particular surface “hot-spots” associated with specific electronic components
within the structure and are not uniformly hot for the entire surface. While
this uniform surface temperature is highly conservative (e.g. near worst case),
a 40°C surface temperature is only one degree below the permitted 41°C
surface contact temperature. While a given coil could potentially be used on
several patients in succession and thus approach its steady-state limiting
temperature on later patients, typically there is opportunity for the coil to
cool in the gaps between sequences and reconfiguration between patients. Thus,
a constant temperature of 40°C is also considered conservative. The pelvis
(with wrist at side) landmark was selected to enable a realistic maximum 4 W/kg
whole body exposure scenario, for a continuous one-hour exposure. This is also
considered to be a worst-case conservative exposure. The implications of an
unclothed model are complex with respect to insulation that keeps body heat in,
but would also insulate against the heat of the coil.RESULTS
With only RF heating peak temperature in tissue 41.2 °C
was observed in abdomen (Figure 3a) and right wrist (Figure 4a). Under the
simultaneous combination of 4 W/kg exposure and broad area, constant high
temperature coil housing contact, temperature was further increased in the skin
in contact with the coil housing, while temperature distribution in all the
other tissues and the hotspot remain unchanged (Figures 3,4).DISCUSSION
While the maximum modeled temperature, with or without the
warm coil housing, is marginally above the 41°C “no justification
required” limit, the multiple conservative assumptions noted above suggest that
the otherwise healthy patient will not be exposed to risks considered unacceptable
in the relevant standards. Comparison of the “no coil” and “hot coil” results
show that the hot coil certainly modifies the temperature distribution in the
skin in direct contact with the coil. Other modeling results with a 2 W/kg WB SAR
(not shown) showed a peak tissue temperature of 40°C. Modeling coil housing surface as a different
boundary condition (e.g. 40°C constant temperature and 100 W/m^2 heat flux
toward body) did not produce significant difference.CONCLUSION
An analysis of the simultaneous application of SAR induced
and coil housing heat on patient safety suggests that there are no risks
outside of acceptable limits.Acknowledgements
References
- Sapareto SA, Dewey WC. Thermal dose determination in cancer therapy. Int J Radiat Oncol Biol Phys 1984;10:787–800.
- ISO 13732-1: Ergonomics of the thermal environment — Methods
for the assessment of human responses to contact with surfaces — Part 1: Hot
surfaces
-
IEC 60601-1: Medical electrical equipment - Part 1: General
requirements for basic safety and essential performance
-
NEMA/MITA MS-14: Characterization of RF Coil Heating in
Magnetic Resonance Imaging Systems (in press)
- Murbach M, Neufeld E, Kainz W, et al. Whole-body and
local RF absorption in human models as a function of anatomy and position
within 1.5T MR body coil. Magn. Reson.
Med. 2014; 71(2):839-845.
- Murbach M, Cabot E, Neufeld E, et al. Local SAR
enhancements in anatomically correct children and adult models as a function of
position within 1.5 T MR body coil. Prog Biophys Mol Biol. 2011; 107(3):428-433.
- Christ A, Kainz W, Hahn EG, et al. The Virtual Family—development of surface-based anatomical models of two adults and two children for dosimetric simulations. Phys Med Biol. 2010;55:N23.
- Gosselin M-C, Neufeld E, Moser H, et al. Development of a new generation of high-resolution anatomical models for medical device evaluation: The virtual population 3.0. Phys Med Biol. 2014;59:5287.
- Murbach M, Neufeld E, Capstick M, et al. Thermal Tissue Damage Model Analyzed for Different Whole-Body SAR and Scan Durations for Standard MR Body Coils. Mag. Reson. Med. 2014;71: 421-431.