Aiping Yao1, Manuel Murbach2, Tolga Goren1, and Niels Kuster1,3
1IT'IS Foundation, Zurich, Switzerland, 2UPV, Valencia, Spain, 3ETH Zurich, Zurich, Switzerland
Synopsis
The induced radiofrequency (RF) electrical field
resulting in unwanted heating in patients with and without implants during
magnetic resonance imaging (MRI) is a complex function of multiple factors
which must be considered in patient risk assessment, including patient anatomy,
imaging positions, RF characteristics of the implant, implant location, and MRI
scanner design and scan configuration. In this paper, the dependence of the
induced fields upon the birdcage design is investigated to determine the
minimal range of RF body coils necessary to guarantee a comprehensive risk
evaluation regarding heating and electromagnetic compatibility for any clinical
MRI scanner.
Introduction
For
patients undergoing magnetic resonance imaging (MRI), RF-induced heating is one
of the major safety concerns, particularly for patients with long, conductive
implants1. The localized RF-induced field distributions can only be
reliably estimated by simulation1-4 with computational high-resolution
anatomical human body phantoms, such as the Virtual Population5
placed at different landmark positions in a birdcage models. Risk assessment of
patient heating requires that the range of possible induced fields experienced
by patients in clinical scanners be considered.
In
this work, we performed a sensitivity analysis of the role of birdcage geometry
and structure on the induced fields and specific absorption rate (SAR) as well
as tangential E-fields (Etan)
to, and deposited power (Pdep)
around, representative generic active medical implants (pacemaker, and spinal
cord stimulator). We then compared the results to the variation observed from
patient anatomy, scan position, and scan polarization, at 64 MHz. Method
Three
studies are designed. Study 1 uses the MRIxViP1.5T database6 to extract
the head SAR, whole-body SAR, partial-body SAR, and B1+ for different anatomical phantoms as well
as the Etan to, and Pdep around, a 40-cm long generic
left-side pacemaker, for a set of ten birdcages, eight anatomical models, all
imaging positions from head to foot, at circular and near-circular
polarizations, as shown in Figure 1 (a).
The role of birdcage diameter and length are assessed based on these results.
In
study 2, a set of additional birdcages were developed to estimate the
sensitivity of these results to different design features, namely high-pass and
band-pass birdcages fed from legs and via I/Q channels (Figure 1(b)). The
patient position sensitivity was estimated by shifting model Fats 2.5 cm
laterally or 10 cm dorsally.
The
goal of study 3 was to investigate the effect of the number of birdcage rungs on
the in vivo RF-induced heating. Birdcages
with length of 70 cm and bore-diameter of 70 cm with rung numbers (N) of 8, 16, and 32 were used to
simulate the exposure of human model Duke at a typical thorax imaging position.
The magnetic field strength and the induced in
vivo electrical field distributions at the imaging position's field-of-view
(FOV) are compared. Results
Figure
1 summarizes the exposure configurations of the birdcages and anatomical models
used in this work.
First,
the comparison of head SAR (hSAR), whole-body SAR (wbSAR), and partial-body SAR
(pbSAR) under normal operating mode7 are reported together with the maximum
allowed magnetic field (B1)
exposure level. Figures 2 and 3 summarize the results of study 1 for birdcage length
and bore diameter sensitivity, respectively. As shown in Figure 2, the longer birdcage
lengths result in larger SARs under normal operating mode, leading to lower B1 field strength at the SAR
limit. In contrast, Figure 3 shows that the effect of birdcage bore-size on the
in vivo SAR distribution is very
limited, as birdcages with different bore-diameters give rise to effectively
the same values of local SAR and wbSAR, allowing, therefore, the same B1 strength.
Next,
the averaged Etan to, and Pdep around, a 40-cm long
generic left-side pacemaker, for a set of ten birdcages, eight anatomical
models, all imaging positions from head to foot, at circular and near-circular
polarizations are shown in Figure 4, as
well as the results from study 2.
The results of study 3
are summarized in Figure 4. Figure 4(a) shows the B1 field distribution over the FOV. Figure 4(b) shows
the induced in vivo electrical field distribution
of Duke, where birdcages of different rung numbers show very similar in vivo induced field distributions. These
birdcages also have the same coefficient of variation (CoV) (23%) of the
induced electrical field over the FOV. Similar results are observed for the
simulated in vivo Etan along a typical spinal
cord routing in the Duke model, as shown in Figure 4(c). The observable variations
in the Etan are due to the
non-ideality of the linear field polarization obtained with birdcages of
varying rung numbers. Conclusions
The evaluations performed
show that birdcage diameter has only a very limited effect on the in vivo RF-induced SAR distribution,
whereas birdcage length plays a critical role in determining in vivo RF-induced SAR. The sensitivity
analysis of other parameters, such as high-pass vs band-pass birdcage design,
birdcage current source, dorsal or lateral patient shifts, birdcage rung count,
led to negligible variation compared to the dominant sources of patient
anatomy, scan position, and birdcage polarization and size.
Acknowledgements
No acknowledgement found.References
1. Nyenhuis
JA. Park SM, Kamondetdacha R, et al. MRI and implanted medical devices: Basic interactions with an
emphasis on heating. IEEE Trans Device Mater Reliab 2005;5(3):467–479.
2. Muranaka
H, Horiguchi T, Usui S, et al. Dependence of rf heating on SAR and implant
position in a 1.5 T MR system. Magn Reson Med Sci 2007;6(4):199-209.
3. Kangarlu
A, Schellock F, and Chakeres D. 8.0-tesla MR system: temperature changes
associated with radiofrequency-induced heating of a head phantom.
4. Murbach M, Neufeld E, Capstick M, et al. Virtual population-based assessment of the impact of 3 Tesla
radiofrequency shimming and thermoregulation on safety and B1+ shimming. Magn Reson Med 2016;76:986-997.
5. Gosselin M-C, Neufeld E, Moser H, et al. Development of a new generation of high-resolution anatomical
models for medical device evaluation: the Virtual Population 3.0. Phys Med Biol
2014;59:5287–5303.
6. IT'IS. https://itis.swiss/virtual-population/explib/overview/
7. IEC. IEC
standard 60601-2-33. Medial electrical equipment - particular requirements for
the basic safety and essential performance of magnetic resonance equipment for medical diagnosis. 2010.