Jacinta Browne1, Christin A Tiegs-Heiden1, Vance T Lehman1, Zaiyang Long1, Robert E Watson1, Gina K Hesley1, and Krzysztof R Gorny1
1Radiology, Mayo Clinic, Rochester, MN, United States
Synopsis
An MRgFUS ablation treatment of lumbar
facet joints in a patient with a traditional MRI non-conditional
pacemaker was completed. A
risk-benefit analysis by a coordinated multi-disciplinary team prior to this treatment
was performed to account for the risks associated with traditional MRI
non-conditional pacemaker. The treatment was
successfully performed as per our institution established cardiac
implanted electronic device (CIED) MRI practice and the patient had no adverse cardiac event
during or following this procedure. By
careful use of our institutional CIED MR-practice guidelines, we demonstrated that such treatments
can be safely achieved for patients with CIEDs on a case-by-case basis.
INTRODUCTION
Magnetic Resonance Imaging-guided Focused Ultrasound (MRgFUS) is a
minimally-invasive treatment modality that utilizes an ultrasound transducer
integrated in an MRI scanner. MRgFUS systems have been successfully used to treat symptomatic
uterine fibroids 1-4 and facet joint pain 5. During MRgFUS treatment,
ultrasound energy is focused within target tissues causing localized thermal
ablation; MRI is essential for treatment
planning, US beam guidance, real-time MR-thermometry and treatment assessment 3. Facet joint treatments in our practice
utilize an ExAblate 2100 MRgFUS system (Insightec, Haifa, Israel) integrated with a 1.5T
MR scanner (Signa Excite, General Electric, Waukesha , WI), which poses significant
safety challenges for patients with implanted cardiac devices potentially excluding
them from treatment. This study describes a MRgFUS ablation of lumbar
facet joints in a patient with a traditional MRI non-conditional
pacemaker, there has been a dearth of such treatments reported 6.METHODS
In this study an 80 years male with a
history of chronic axial
lower back pain underwent MRgFUS ablation of the bilateral L3-4, L4-5, and
L5-S1 facet joints. The patient had a history
of symptomatic sinus bradycardia which was managed with a dual chamber transvenous
pacemaker (Assurity DR2240, MRI non-conditional, St Jude Medical, Memphis, TN); the
patient was not pacemaker-dependent. A risk-benefit analysis was carried out prior
to this ablation treatment to account for the risks associated with MRI
non-conditional pacemaker, and the exam was performed in accordance with our
established cardiac implanted electronic device (CIED) / MRI-practice, with
over 3000 patients scanned, that includes a coordinated team of
physicians, cardiology pacing nurses, MRI physicists and technologists. Prior to entry into MR scanner room the cardiology
personnel programmed the patient’s pacemaker to DOO (dual-absence of sensing -
no response to sensed input) mode of 80bpm.
The initial transducer
localization within MRI coordinates for the ablation treatment required acquisition
calibration scans based on single-shot fast spin echo pulse sequence. For this
patient the 1.5W/kg limit was exceeded in the scanner-based SAR estimation for
the corresponding patient weight of 77kg where the SAR was predicted to be 1.73W/kg.
Therefore, this calibration scan was completed using a QA phantom. The patient was subsequently positioned into
the feet first-supine position on top of the MRgFUS table with his lower back
located directly above the transducer. After this process the table was slowly
advanced into the iso-center of the MR bore. For the entirety of the treatment the
cardiology nurse continually monitored the patient’s cardiac function through electrocardiography. The MRI was performed in normal operating
mode for both SAR and gradient switch rates. All MR sequences executed during the 206mins of the procedure were adjusted to
maintain the whole-body SAR below the threshold of 1.5W/kg, which is a
consensus value for safe scanning of patients with non-MR conditional pacemakers
at our institution 7.
Furthermore, the real-time SAR display was monitored and sequences were
stopped when the 10s SAR average exceeded the 1.5W/kg threshold.RESULTS
Following a 3D scout (0.26W/kg), a set of three anatomical T2-weighted
fast spin echo sequences, in axial, sagittal, and coronal planes, were acquired
(SAR =1.13, 0.95, and 0.94W/kg) for
treatment planning (Table 1). These
images were then transferred to the ExAblate workstation where the target locations
of the MRgFUS ablation were planned using the ExAblate treatment planning
system. MRgFUS of the bilateral selected lumbar facet joints was performed using
22 individual sonications 5.
Each sonication was monitored in the axial plane using phase-sensitive
gradient-recalled echo sequences acquired for the purpose of MR thermometry
feedback with temporal resolution of 6s. The images were transferred in real-time to
the ExAblate workstation where the thermometry and the corresponding treatment
dose maps were overlaid onto the patient anatomy. Upon completion of the
treatment T2-weighted fast spin echo sequence with fat suppression (1.12W/kg)
was acquired to assess for edema around the target joints. The patient had no
adverse cardiac event during or immediately following the MRgFUS procedure. His
pacemaker was subsequently interrogated, which confirmed no damage or
alteration, and it was reprogrammed to its original settings. Informed consent for publication purposes was
obtained from the patient.DISCUSION
MRgFUS ablation of the facet joint is
a new treatment being used for facet joint pain 5, 8, 9 which is
dependent upon the ability to accurately image the area undergoing treatment in
real-time using MRI. A large proportion of the patient group with facet joint
pain tend to be an older population with confounding cardiac morbidities often
with a CIED 10. Despite the fact that these devices
are non-MR conditional are regarded as presenting contraindications to MRI 11, 12, our institution has under certain
precautions safely scanned over 3000 such devices over the last ten years 7.
By adopting and incorporating these guidelines into the process of an
MRgFUS procedure it was possible to safely complete a facet joint ablation with
no changes in pacemaker function or pacing threshold, or any other complications. CONCLUSION
This study
reports on successful MRgFUS lumbar facet joint ablation in a patient with a MRI
non-conditional pacemaker. By careful use of our MRI CIED protocol 7, we demonstrated that the
MRgFUS ablation treatment of facet joints can be safely achieved for patients
with CIEDs on a case-by-case basis.Acknowledgements
NoneReferences
1.Gorny
KR, Woodrum DA, Brown DL, Henrichsen TL, Weaver AL, Amrami KK, et al. Magnetic
resonance–guided focused ultrasound of uterine leiomyomas: review of a 12-month
outcome of 130 clinical patients. Journal of Vascular and Interventional
Radiology. 2011;22(6):857-64.
2.Hesley GK, Felmlee JP, Gebhart JB, Dunagan
KT, Gorny KR, Kesler JB, et al., editors. Noninvasive treatment of uterine
fibroids: early Mayo Clinic experience with magnetic resonance imaging-guided
focused ultrasound. Mayo Clinic Proceedings; 2006: Elsevier.
3.Hesley GK, Gorny KR, Henrichsen TL,
Woodrum DA, Brown DL. A clinical review of focused ultrasound ablation with
magnetic resonance guidance: an option for treating uterine fibroids.
Ultrasound quarterly. 2008;24(2):131-9.
4.Hesley GK, Gorny KR, Woodrum DA.
MR-guided focused ultrasound for the treatment of uterine fibroids.
Cardiovascular and interventional radiology. 2013;36(1):5-13.
5.Tiegs-Heiden Christin A LVT, Gorny Krzysztof R. , Boon Andrea J. ,
Hesley Gina K. . A Case of Improved Treatment Response Following MR-guided
Focused Ultrasound for Lumbar Facet Joint Pain. Mayo Proceedings. 2019;In
press.
6.Ito H, Fukutake S, Yamamoto K,
Tanaka S, Yamaguchi T, Taira T, et al. Magnetic Resonance Imagingāguided
Focused Ultrasound Thalamotomy for Parkinson's Disease with Cardiac Pacemaker:
A Case Report. Movement disorders clinical practice. 2018;5(3):339.
7.Padmanabhan D, Kella DK, Deshmukh
AJ, Mulpuru SK, Mehta RA, Dalzell CM, et al. Safety of Thoracic Magnetic
Resonance Imaging for Patients with Pacemakers and Defibrillators. Heart rhythm.
2019.
8.Harnof S, Zibly Z, Shay L, Dogadkin
O, Hanannel A, Inbar Y, et al. Magnetic resonance-guided focused ultrasound
treatment of facet joint pain: summary of preclinical phase. Journal of
Therapeutic Ultrasound. 2014;2(1):9.
9.Krug R, Do L, Rieke V, Wilson MW,
Saeed M. Evaluation of MRI protocols for the assessment of lumbar facet joints
after MR-guided focused ultrasound treatment. Journal of Therapeutic
Ultrasound. 2016;4(1):14.
10.Grewal SS, Gorny KR, Favazza CP,
Watson RE, Kaufmann TJ, Van Gompel JJ. Safety of Laser Interstitial Thermal
Therapy in Patients With Pacemakers. Operative Neurosurgery.
2018;15(5):E69-E72.
11.Strom JB, Whelan JB, Shen C, Zheng SQ,
Mortele KJ, Kramer DB. Safety and utility of magnetic resonance imaging in
patients with cardiac implantable electronic devices. Heart rhythm.
2017;14(8):1138-44.
12.Sommer
T, Luechinger R, Barkhausen J, Gutberlet M, Quick H, Fischbach K, editors.
German Roentgen Society Statement on MR imaging of patients with cardiac
pacemakers. RöFo-Fortschritte auf dem Gebiet der Röntgenstrahlen und der
bildgebenden Verfahren; 2015: © Georg Thieme Verlag KG.