Maxim Terekhov1, Ibrahim A. Elabyad1, Carsten Kögler2, Theresa Reiter1,3, and Laura M. Schreiber1
1Chair of Cellular and Molecular Imaging, Comprehensive Heart Failure Center (CHFC), University Hospital Würzburg, Wuerzburg, Germany, 2Rapid Biomedical, Rimpar, Germany, 3Department of Internal Medicine l, University Hospital Wuerzburg, Wuerzburg, Germany
Synopsis
The purpose of this study was to simulate the
variation of SAR caused by shifting the position of a cardiac array prototype
when different B1-shimming phase vectors are employed. The variation
of 10g average SAR distribution due to displacing the anterior array position
on the thorax (central, shifted 5 cm to the right, left, head, foot and
diagonal directions) was computed and analyzed using two human voxel models.
The final goal is to determine maximal driving voltage of the array with
adequate safety margins considering new hot-spots originated from possible
array misplacement
Introduction
MRI using ultra-high B0 field (UHF) is an emerging
technique to improve the signal-to-noise ratio (SNR) compared to clinical scanners (B0=1.5-3T). However, the application of UHF in cardiac MRI
(cMRI) faces numerous hurdles. The shortened wavelength of
field at 7T (
λ
≈ 12 cm - in the order of a human thorax), results
in strong
field inhomogeneity. Different types of RF
coils including, multichannel Tx (mTX) loops1-4 and dipole antenna arrays5,6 are used for
mitigating these inhomogeneities performed by adjusting phases of the driving
voltage for the individual elements (B1+-shimming phase
vector). Motivation
For mTX arrays applied in UHF
cMRI the main problem remains achieving sufficiently high flip-angle (FA) at
the posterior wall of a myocardium. This plays a key role in using saturation
and inversion rf-pulses which is prerequisite in the functional cMRI
application (first-pass perfusion, late-gadolinium enhancement, T1-mapping).
Thus, dedicated B1-shimming and sufficient driving voltages are
essential to achieve maximal B1+ penetration depth. However,
at ultra-high-field the critical factor limiting mTX-array usage is the local energy deposition characterized by the specific absorption
rate (SAR). The peak SAR computed using human body models is used for
setting a safety limit for the driving voltage of the array via the so-called “k-factor”.
The individual human thorax anatomy varies in shape and dimensions influencing
the position of the array location. This may lead to an essential deviation of
the SAR from computed model values.
The purpose of this study was
to simulate the variation of SAR for an mTx array prototype for 7T cMRI caused
by a displacement of the coil position with respect to the thorax and different
B1-shimming phase vectors. The final goal is to determine the maximal
driving voltage of the array with adequate safety margins considering possible new
SAR “hot-spots” originated from possible array misplacement.Materials and Methods
Both the anterior and
posterior parts of the cardiac array prototype are composed of eight loop
elements (Fig 1a). For SAR safety and
field optimization, EM-simulations were
performed using CST-Microwave-Studio (CST-MWS). The simulations were performed
using Duke and Ella human voxel model
with 2×2×2 mm3 resolution. The final total number of mesh cells was
40.9 million. The local averaged 10g SAR values were evaluated in CST-MWS using
IEEE/IEC-62704-1 standard averaging method. For all cases, the 16 loop elements
of the array were excited with the same total input stimulated power (Pin=8W). In this work, we
analyzed the variation of the SAR which comes from the two sources:
1) Adjustment of the phases of individual elements
needed to provide homogeneous
profile. Four phase vectors {Φ}1-{Φ}4
preliminary pre-computed to provide optimized
profile in Duke and Ella models were
considered.
2) Variation of SAR due to different positioning of
the anterior part of the array on the human subject thorax. The anterior array
position was shifted from the central allocation on 50 mm to the right, left,
head, foot, all diagonal directions and tilted on 100 from the horizontal
allocation.
From the 3D SAR maps simulated for each of the phase
vectors the primary maximal intensity projection (MIP1) maps within
the slabs of 20 mm distance from the anterior part was computed (Figure
2b). Then the secondary MIP2 was computed using stack of the above
determined 4 MIP1 for each of 4 phase vectors. The relative
variation of SAR MIP2 with respect to the reference position was
analyzed in order to determine worst-case situation.Results
Fig. 1(a)
demonstrates a sketch of the 8Tx/16Rx mTX array prototype and its default
position on the human subject thorax. Fig. 1(b) shows the coronal slabs used to evaluate 10g SAR. Fig. 2(a) shows an example of the volume-rendered
SAR distribution for centered array and zero phase vector (prior to B1-shimming).
Figure 2b demonstrates slab for MIP1 computation. Fig.3 Shows the scheme
of computing anterior MIP2 SAR with 4 phase vectors. Fig 4 and 5 demonstrate
variation of MIP2 SAR due to change of the position of the array on
the thorax (four positions) of Duke and Ella models normalized to the maximal
SAR of the central array position.Discussion
Maximal variation of 10g SAR due to the changes of the
position of the anterior array part was observed for horizontal (left-right
shift). However, the maximal increase of SAR in the new hot-spot does not
exceed 25% from the initial position for the Duke model. For the smaller
thorax dimensions of the model Ella no increase of SAR is observed for the
shifted array positions. Usually, without such an analysis of conservative safety
margins for the worst-case SAR (and, therefore, k-factor) use to be applied
(typically factor 2 from the modeled value). The computed SAR variation originated
from the array displacement, however, should allow for setting more realistic safety
margins. Conclusion
The performed simulation demonstrated sufficient
stability of SAR calculation with respect to array position changes and will provide
more flexibility for the cMRI application demanding to the achievable B1+
magnitude.Acknowledgements
Financial support was obtained from the German Ministry of Education and Research (BMBF) under grants: 01EO1004 & 01EO1504.References
[1] Graessl, et al.,
(2014), Modular 32-channel transceiver coil array for cardiac MRI at 7.0
T, MRM, 72(1), 276-290.
[2] Dieringer, et al.,
(2011), Design and application of a four-channel transmit/receive surface coil
for functional cardiac imaging at 7T, JMRI, 33(3), 736-741.
[3] Gräßl, et al., (2013), Design, evaluation and application of an eight
channel transmit/receive coil array for cardiac MRI at 7.0
T. EJR, 82(5), 752-759.
[4] Thalhammer, et al.,
(2012), Two-dimensional sixteen channel transmit/receive coil array for cardiac
MRI at 7.0 T: design, evaluation, and application. JMRI, 36(4),
847-857.
[5] Oezerdem, et al.,
(2016), 16-channel bow tie antenna transceiver array for cardiac MR at 7.0
tesla, MRM, 75(6), 2553-2565. [6]
Raaijmakers, et al., (2016),
Dipole antennas for ultrahigh-field body imaging: a comparison with loop
coils. NMR in Biomedicine, 29(9), 1122-1130.