Laura Gui1, Orane Lorton1, Max Scheffler2, and Rares Salomir1,2
1Image Guided Interventions Laboratory, Faculty of Medicine, University of Geneva, Geneva, Switzerland, 2Department of Radiology, University Hospitals of Geneva, Geneva, Switzerland
Synopsis
Magnetic
resonance thermometry via the classic time-referenced PRFS method is perturbed by
tissue motion and external magnetic field perturbations. We extend an existing 2-D
reference-free PRFS thermometry method to 3-D data, by performing near-harmonic
reconstruction of the background phase inside a sphere. Ex-vivo HIFU sonication
experiments and in-vivo “zero-measurement” of a human brain showed good method
accuracy and precision, fast computation times and similar results with the time-referenced
PRFS method for motion-free scenarios. This study demonstrates the method’s feasibility
for real time thermometry of moving tissues.
Introduction
To
date, the most widely used technique for monitoring thermal therapies is
magnetic resonance thermometry via the proton resonance frequency shift (PRFS)
method. In this method, temperature rise computation involves subtracting a reference phase map from the current phase map, which makes it
prone to errors stemming from tissue motion and external perturbations of the magnetic field. Salomir et al.1 and Kickhefel et al. 2 proposed a
reference-free PRFS thermometry technique for 2-D acquisitions, based on
near-harmonic 2-D reconstruction of the background phase from the measured
phase values on a thin, nearly-closed unheated border around the heated
region. This study extends the reference-free PRFS thermometry method1,2 to 3-D data, by performing near-harmonic reconstruction of the background
phase inside a sphere.Method
Experimental
data was acquired with a 3-D FLASH sequence (TE/TR=5/8ms,BW=600HZ/pixel,FA=12°, water-selective binomial, voxel size=2mm isotropic) on a 3T whole-body MRI scanner for two scenarios: ex-vivo high
intensity focused ultrasound (HIFU) sonication of a tissue-mimicking gel using
a single-loop coil (PF=6/8x6/8, acquisition time=25s), and in-vivo
acquisitions of a human voluntary brain (“zero-measurement”) using a 64-channel
coil and CAIPIRINHA=3x2 (PF=7/8x7/8, acquisition time=10s). HIFU was
generated by an MR-compatible phased array transducer (256 elements, operating
at 1MHz). Reference-free reconstruction of the background phase was performed inside
the volume of a sphere whose discretized surface (1 voxel thick) was assumed to
be unheated. The background phase map was modeled as a harmonic function (i.e. whose Laplacian is null) plus a second
order polynomial (non-harmonic component) in regions with homogenous or
linearly varying magnetic susceptibility1. We aimed to solve the inner
Dirichlet problem. After removing the non-harmonic component (i.e. constant
Laplacian) using the differential of the mean value on two concentric spheres,
the harmonic component of the background phase on the surface of a sphere of radius $$$R_0$$$ and polar coordinates $$$\theta\in[0,\pi),\varphi\in[0,2\pi)$$$ was approximated as a linear combination of real spherical harmonic
functions $$$C_l^m(\theta,\varphi),S_l^m(\theta,\varphi)$$$ with coefficients $$$a_{lm},b_{lm}$$$: $$\Phi_{bk}(r=R_0,\theta,\varphi)\approx\sum_{l=0}^L\sum_{m=0}^l\left[a_{lm}C_l^m(\theta,\varphi)+ b_{lm}S_l^m(\theta,\varphi)\right], \quad (1)$$ where $$C_l^m(\theta,\varphi)=N_{lm}\cdot P_l^m(\cos \theta)\cdot\cos(m\varphi)$$ $$S_l^m(\theta,\varphi)=N_{lm}\cdot P_l^m(\cos\theta)\cdot\sin(m \varphi)$$ $$$P_l^m$$$
– Legendre polynomials and $$$N_l^m=\sqrt{\frac{2l+1}{4\pi}\cdot\frac{(l-m)\,!}{(l+m)\,!}}.$$$ The background phase of each voxel inside the sphere
volume ($$$r<R_0$$$) was then estimated as: $$\Phi_{bk}(r,\theta,\varphi)\approx\sum_{l=0}^L\left(\frac{r}{R_0}\right)^l\sum_{m=0}^l\left[a_{lm}C_l^m(\theta,\varphi)+b_{lm}S_l^m(\theta,\varphi)\right]. \quad (2)$$
To solve for $$$a_{lm},b_{lm}$$$ from eq. (1), a uniform sample distribution on the sphere surface (Fig.1a,b) was obtained by Monte Carlo random sampling of $$$\varphi$$$ from a uniform distribution
on $$$[0,2\pi)$$$, and of $$$\cos\theta$$$ from a uniform distribution on $$$[-1,1)$$$. This is because the
differential area element on a unit sphere is: $$$d\Omega=\sin\theta d\theta d\varphi = -d(\cos\theta)d\varphi$$$. The orthonormality of the sampled spherical harmonic functions can be verified by inspecting the matrix of their pairwise dot products (Fig.1c). The number of samples was empirically
chosen as proportional to the sphere surface: $$$N=\lfloor 4\pi(R_0)^2\rfloor$$$. Sample uncertainty
was considered by weighing with corresponding magnitude data (low magnitude =
high uncertainty). Equation (1) was written as $$$W\cdot\Phi_{bk}=W\cdot CS\cdot v,$$$ with $$$W$$$ – the $$$N\times N$$$ diagonal matrix of
weights given by sample magnitude values, $$$\Phi_{bk}$$$ – the $$$N\times 1$$$ vector of phase data, $$$CS$$$ – the $$$N\times (L+1)^2$$$ matrix of sampled spherical
functions, and $$$v$$$ – the $$$(L+1)^2\times 1$$$ vector concatenating coefficients $$$a_{lm},b_{lm}$$$. Vector $$$v$$$ was extracted as $$$v=(CS ^\top\cdot W\cdot CS)^{-1}\cdot(CS^\top\cdot W\cdot\Phi_{bk})$$$, allowing background phase estimation inside the sphere with eq.(2).
For comparison, time-referenced PRFS temperature elevation
maps were computed by subtraction of a reference phase map, and background phase
drift correction using an unheated volume. For both scenarios we used $$$L=4$$$ and sphere
radius = 10 voxels. For the in-vivo scenario, cerebro-spinal fluid (CSF) was
excluded from data sampling due to susceptibility contrast. In a real treatment
scenario, CSF would not be targeted by HIFU.Results
For the
ex-vivo sonication, endpoint temperature elevation maps obtained with the proposed
reference-free method and with the time-referenced PRFS method are in close
agreement (Fig.2a,b). “Zero-measurement” comparison (Fig.2d) revealed good method
precision (0.78°C) and accuracy (0.28°C). Measured and reconstructed phase values along
an equatorial circle on the sphere closely match (difference standard deviation
= 0.96°). “Zero-measurement”
reference-free temperature maps from in-vivo acquisitions (Fig.3) confirmed good
method accuracy (0.57°C) and precision (0.29°C). Average computation time was 75ms
per temperature map (radius = 10 voxels) on a 3.4GHz processor. No ill-conditioned operation occurred.Discussion
The good precision and accuracy of the proposed method, its fast computation times, and its results’ similarity with the time-referenced PRFS method for motion-free sonication, support its application to real-time temperature monitoring in tissues with homogenous or linearly varying magnetic susceptibility. Its main advantage compared to the time-referenced PRFS method is its robustness against inter-scan motion. The 3-D geometry is a natural mathematical framework for near-harmonic functions, contrary to the previous 2-D approach, involving analytical approximation on a restricted subspace1,2. The main limitation of the method is the need for a user-provided unheated border, and its adaptation to potential tissue motion, which could be tackled by tissue tracking. The approach can be coupled with a fast acquisition (TWIST-VIBE, compressed sensing) to improve temporal resolution and address abdominal targets.Conclusion
The
proposed reference-free PRFS thermometry method constitutes a promising
alternative to the time-referenced PRFS method, especially valuable during
sonication of moving tissues.
Acknowledgements
No acknowledgement found.References
[1]
Salomir R, Viallon M,
Kickhefel A, Roland J, Morel D, Petrusca L, Auboiroux V, Goget T, Terraz S, Becker
C, Gross P. Reference-free PRFS
MR-thermometry using near-harmonic 2D reconstruction of the background phase.
IEEE Trans Med Imag 2011;31:287–301.
[2]
Kickhefel A, Rosenberg C,
Roland J, Viallon M, Gross P, Schick F, Hosten N, Salomir R. A pilot study for
clinical feasibility of the near-harmonic 2D referenceless PRFS thermometry in
liver under free breathing using MR-guided LITT ablation data. Int J
Hyperthermia 2012;28(3):250-266.